The products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
All Recalls
95,123 total recalls in our database
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
ultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in vitro diagnostic
Ventana Medical Systems
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Syngo.via systems with SW VB20A
Siemens Medical Solutions USA
A software update was released on August 22, 2017. Performing this software installation and upgrading syngo.via system to the software version VB20A_HF02 will resolve various bugs and inconsistent diagnosis with the old system.
An incorrect ProLite Mesh configuration was packaged as product code 1010306-06, with lot number 412298. ProLite Mesh manufactured as product code 1010306-04 was inadvertently placed inside the pouch of packages labeled with product code 1010306-06 and lot number 412298.
Fracture of the bolt connecting the insertion guide to the intramedullary nail during insertion of the femoral nail.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
Better-Bladder cardiopulmonary bypass blood reservoir; Models: BB14 (individual, sterile, pouched) and BB14NS (sold bulk to kit manufacturers)
Circulatory Technology
The device is used as part of extracorporeal membrane oxygenation (ECMO) cardiopulmonary bypass circuits and may collapse during use which can increase resistance to flow in the venous line and cause a drop in circuit blood flow.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
MOOG Curlin Infusion Administration Set, Ref 340-4114, Non-DEHP Microbore Tubing with Non-Vented Bag Spike, packaged 20/box, for use with the MOOG Curlin Ambulatory Volumetric Infusion Pump, RX, Sterile. Product Usage: Administration set to be used with the Moog Curlin infusion Pump. The pump can be used for intravenous, intra-arterial, epidural, subcutaneous, or enteral infusions.
Zevex Incorporated (dba Moog Medical Medical Devices Group)
Particulate found on the Curlin spike.
On June 15, 2017, STERIS identified that the purple ink located on the outside label of Vis-U-All Pouches may migrate following steam processing causing cosmetic defects on the exterior of the pouch and in some instances, transference of the ink onto the device(s) contained within the pouch.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.
ST AIA-PACK¿ PROG II, Progesterone II Assay, Part Number: 025239
Tosoh Bioscience
Falsely elevated progesterone assay values will occur when using certain progesterone assay part numbers due to cross reactivity of DHEA-S, a metabolite of the DHEA (dehydroepiandrosterone) supplement.
The device could have a missing spanner nut and snap ring which could potentially result in the equipment falling.
Ultra Renew
Accord Media
Distribution of medical devices with unapproved green LED light used to treat hyperpigmentation
2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.
Fresenius Medical Care Renal Therapies Group
While reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
The recalled units are mislabeled with an incorrect expiration date. The expiration date printed on the label is 2202-06. The correct expiration date should be 2022-06.
Certain kits packed by the firm contain BD 22GA 1-1/2 inch SafetyGlide Non Sterile Bulk needles that may contain loose polypropylene foreign material above the release specification.