All Recalls

95,075 total recalls in our database

SIGNA Architect AIR

GE Medical Systems

πŸ₯ Medical Devices Class I - Dangerous

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

May 23, 2025 Other Medical Devices Nationwide View Details β†’

The firm issued a field safety notice after becoming aware of three lots of products not having quarterly dose audits being completed on time in the first quarter of 2024. Routine lot by lot testing was completed properly however the quarterly dose audit (sampling/testing to monitor the gamma sterilization process and confirm the validated parameters remain effective to assure sterility and bioburden reduction per requirements) was not completed on time.

Jun 12, 2025 Other Medical Devices View Details β†’

Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. Issued discovered via a customer complaint.

Jun 12, 2025 Implants & Prosthetics View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

Some DIOX-602 units currently in use within the United States do not meet the minimum SSD requirement of 18 cm.

Jun 19, 2025 Other Medical Devices View Details β†’

SIGNA PET/MR

GE Medical Systems

πŸ₯ Medical Devices Class I - Dangerous

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

May 23, 2025 Other Medical Devices Nationwide View Details β†’

Discrepancies between sterilization instructions in the Japanese language product package inserts and those published in the manufacturer product Instructions For Use (IFU).

Jun 18, 2025 Surgical Instruments View Details β†’

Due to packaging discrepancy with the incorrect tibial inserts/implants being packaged. Issued discovered via a customer complaint.

Jun 12, 2025 Implants & Prosthetics View Details β†’

Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting

Jun 23, 2025 Diagnostic Equipment Nationwide View Details β†’

Discovery MR750w 3.0T

GE Medical Systems

πŸ₯ Medical Devices Class I - Dangerous

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

May 23, 2025 Other Medical Devices Nationwide View Details β†’