The firm was informed that the product did not contain vitamin B12 (Methylcobalamyn) as listed on the label.
All Recalls
95,075 total recalls in our database
McEvoy Ranch 100% Certified Extra VirginOlive Oil; Made in California; 1 L, 33,8 FL OZ UPC 6 34832 00291
McEvoy of Marin
Reports of bottles cracking and breaking due to inconsistent filling.
Subpotent Drug: final drug concentration is subpotent as compared to the concentration indicated on the label.
Brilliance Big Bore Radiology 728244
Philips Medical Systems (Cleveland)
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
CS-7, Model No. A4C1
Konica Minolta Medical Imaging USA
Panasonic has issued a Mandatory Product Safety Notice, indicating a possibility of battery ignition (fire) in CF-C2 "Toughbook" laptops. Since Konica Minolta uses this laptop in some of its product configurations, it notified its customers to utilize a BIOS Utility for the laptops to minimize the risk of an incident while Panasonic develops a final solution.
Merlin PCS programmer (Model # 3650), Software (Model # 3330)
St Jude Medical
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
ABX Pentra Micro ALBUMIN 2 CP, Serum/Plasma, REF 1300032563 The device is intended for use in conjunction with certain materials to measure a variety of analytes.
Horiba Instruments, Inc
When the ABX Pentra Micro ALBUMIN 2 CP (ALBT2) Reagent 2 is sampled and dispensed immediately after an ABX Pentra Creatinine 120 CP (CREA3) Reagent 2 dispense, the final result of the microalbumin may be elevated more than 10%. This increase may affect serum, plasma and urine sample types.
Invalid calibration with low calibrator S1 while using the product.
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Artis Q.zen floor, Material no. 10848353, for angiography and whole body radiographic/fluoroscopic procedures.
Siemens Medical Solutions USA
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Artis Q.zeego, Material no. 10848283, for angiography and whole body radiographic/fluoroscopic procedures.
Siemens Medical Solutions USA
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
Philips Ingenuity Core 728321
Philips Medical Systems (Cleveland)
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Randox Liquid Assayed Specific Protein Control - Level 2 (SP Control 2), Cat. no. PS2683
Randox Laboratories
The recalling firm has confirmed that Free Kappa light chains in Liquid Assayed Specific Protein Controls increase over the shelf life of the product. This may lead to a delay in reporting Free Kappa Light Chains results, which are used as an aid in diagnosis of monoclonal gammopathy of undetermined significance and for Multiple Myeloma through review of the ratio with free lambda chains.
Ingenuity Core128 728323
Philips Medical Systems (Cleveland)
A problem has been detected in the X-Ray Tube. The affected tubes have a production defect that allows oil leakage from the tube housing, if it were to re-occur, could pose a risk for patients or users.
Artis Q biplane, Material no. 10848282, for angiography and whole body radiographic/fluoroscopic procedures.
Siemens Medical Solutions USA
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
There is a potential for a different set to be mixed within the lot, where the incorrect set does not contain an air eliminating filter.
Artis Q ceiling, Material no. 10848281, for angiography and whole body radiographic/fluoroscopic procedures.
Siemens Medical Solutions USA
After the Large Display returns from power save mode, it may not show an image, and stay black without showing an error message although X-ray is still possible. The problem does not occur during an ongoing procedure. If the problem occurs, the system cannot be operated normally. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
The firm is releasing new firmware for high voltage implantable cardiac devices (ICD) and cardiac resynchronization therapy defibrillators (CRT-D) intended to strengthen security by providing an additional layer of protection against unauthorized device access. For older generation devices not capable of accepting the firmware, the firm is advising customers of the option to permanently disable the RF communication capability in these devices.
Teleflex MEDICAL Pilling(R) UNIVERSAL SHEARS 7 1/2", REF 142542 Product Usage: general surgical scissors
Teleflex Medical Europe
The incorrect plastic was used to manufacture the handles of these shears, which causes the handles to melt if sterilized.