The conduit length and branch length of a SG Pulmonary Hemi-Artery were switched on the product labeling. The graft labeling indicates the conduit length is 1.0cm and the branch length is 4.5cm.
Calibration setting was incorrectly entered during manufacturing causing the optical coherence tomography (OCT) to obtain the incorrect two-dimensional images of ocular structures of the anterior and posterior segments of the eye.
These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
The firm has identified performance anomalies with the device caused when more than 64 TD60 telemetry transmitters have been connected to a central station. The anomaly causes some ECG waveforms to appear as a flat line at the centralized viewing station and will manifest if more than 64 TD60 transmitters have been connected to a given central station server. Each central station supports only 32 transmitters at a time. For the issue to manifest, the BeneVision DMS needs to be configured to use flexible device assignment, have more than 64 TD60 transmitters, and the workflow must allow 64 TD60s to move around between multiple central stations in the same care area.
Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacksΒΏ and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
These catheters may have the following deficiencies: 1) barium sulfate omission, and 2) actual catheter tip length is one centimeter shorter than printed on the tip
Atellica Solution system software (SW) version 1.13 and lower, five (5) behavioral issues were identified: 1. Reagent Lock may not work as intended with Primary Reagent ReadyPacksΒΏ and Ancillary Packs when more than one (1) lot of specific Reagent kits are loaded on the analyzer at the same time. 2. Sample barcodes may be incorrectly read 3. The Atellica IM 1300 Analyzer and Atellica IM 1600 Analyzer may not run CEA dilutions at 50x and 100x. The Analyzer may post a Sample Integrity Error when 50x or 100x dilutions are ordered for Carcinoembryonic Antigen (CEA). Dilution factors lower than 50x and neat samples are not affected by this behavior. 4. The Atellica CH 930 Analyzer may continue to display the default reporting units after the reporting units have been changed. 5. The Atellica CH 930 Analyzer may not display results for sample if certain assays are configured to auto-dilute results above or below the assay measuring interval.
Reports of circulatory support system motor stopping. Motor stop can be attributed to a break of a wire or short circuit between wires within the electrical cable where the cable exits the motor cable bend protection.