Quality control results are not processed for Abbott Diagnostics Alinity ci series (abdicqii) instrument. This has been determined to be a driver malfunction. When an instrument transmits results from a quality control (QC) sample ID ("Q^Control" in SPM segment), and the option Hold Results Until Complete is enabled, the results do not parse into IM. The communication Trace shows the sample being transmitted, but no information is logged in the Specimen Event Log (SEL). There is a potential for harm if QC results are not processed, downstream auto verification rules may misfire based on missing QC results leading to a delay in, or incorrect treatment. QC results will not be available for evaluation by the lab. There have been no reported instances of patient harm from this malfunction.
All Recalls
95,064 total recalls in our database
CHS Custom Convenience Kit-GENERAL UTILITY TRAY Product Number:S1528-02
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
CHS Custom Convenience Kit- COBANΒΏ, LF (1"X5 YDS) Product Number:D5-14591
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
CyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS APPLIER, REF: 8065754001. For ophthalmic use.
Alcon Research
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
CHS Custom Convenience Kit-MOSQUITO FORCEPS 5 STR Product Number:B9-18914
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
STREAMLAB - REFRIGERATED STORAGE MODULE 15000
Siemens Healthcare Diagnostics
The modules may have been shipped without a warning label for potential risk of interference to pacemaker functionality.
Complaint monitoring noted a trend of packaging being breached for the MiniMAX Stem.
There is a potential risk associated with the PEEK tip of the J-Plasma Precise 360 Handpiece separating from the handpiece and retained in the tissue.
APTIO RACK OUTPUT MODULE
Siemens Healthcare Diagnostics
The modules may have been shipped without a warning label for potential risk of interference to pacemaker functionality.
TRINIAS Digital Angiographic System intended to be used for cardiac angiography, neurovascular angiography, abdominal angiography, peripheral angiography, rotational angiography, multi-purpose angiography and whole body radiographic/fluoroscopic procedures.
Shimadzu Medical Systems Usa Com
Due to a software issue the C-arm may unexpectedly moved when being moved to the park position and during specific operations. There is a possibility the C-arm will come into contact with a patient or healthcare personnel.
CHS Custom Convenience Kit-ROCHESTER PEAN, CURVED (8") Product Number:D36-16669
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
Reprocessing instructions for the SGS/SGA Surgical Handpieces have been updated with instructions of manual cleaning and specific parameters for autoclave sterilization. The previous operation manuals associated with the SGS/SGA products must be replaced by new operation manuals.
Software anomalies were reported to affect the following system functionalities: startup and shutdown, system and network communication, motion control, image acquisition and output, dose reporting, and logs generation. Software version 3.1.7 will be installed on all 2nd edition imaging systems.
CHS Custom Convenience Kit-SPECIALISTΒΏ CAST PADDING 3" X 4 YDS) Product Number:D4-324SCP
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility
The firm received complaints related to the locking mechanism on the steerable inserter not functioning as intended. The firm confirmed that the issue is related to lack of lubrication of the instrument. There are two potential hazards associated with the implant pivoting unexpectedly during implant insertion into the disc space. First, if the implant is initially inserted into the disc space via the impaction method and the implant pivots unexpectedly, the implant potentially may unintentionally contact sensitive anatomy. Additionally, the user may potentially experience difficulty positioning the implant within the disc space due to the lack of rigidity. A third potential hazard is associated with the inserter not unlocking at the articulating pivot head during implant positioning within the disc space. Once the implant is in the prepared disc space and contacts the ventral annulus, the surgeon unlocks the articulating pivot head and positions the implant perpendicular to the spinous process. The surgeon may then elect to lock/unlock the articulating pivot head for desired final positioning of the implant. If the inserter does not unlock during these final implant positioning steps, the user may potentially implant the cage in a suboptimal trajectory/placement.
NucliSENSΒΏ Magnetic Extraction Reagents
BioMerieux SA
Data loggers showed that one shipment to the United States experienced low temperatures that may impact performance of this product.
The pouches of certain lots of product may have wrinkles along the pouch seal that could result in a channel within the seal. Should there be a channel in the pouch seal it may result in a breach in the sterility of the product.
CHS Custom Convenience Kit-LP NEEDLE, 22G X 1 Product Number:NC401000
Custom Healthcare Systems
Convenience kits were sealed in immediate Tyvek pouches using incorrect specifications compromising product sterility