BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
There is a possibility that the liquid suction pump on some AIA-2000 analyzers may not perform as intended and cause an insufficient volume of wash solution to be removed during the washing step. Incomplete washing of the sample may result in falsely elevated test results for the sandwich assays and falsely decreased test results for the competitive binding assays. Insufficient washing is detectable through review of the calibration data and QC trending. For example, if incomplete washing occurs, the reported values for the zero- calibration rate will be above the acceptance range. The calibration value may shift and the instrument will not calibrate successfully. If this issue occurs, there may be a delay in testing and reporting patient test results. There have been no user or patient injuries reported related to this issue.
BD Vacutainer Urine Collection Cups may contribute to a slight elevation of Red Blood Cell (RBC) counts when used with the Sysmex UF Series Analyzers. BD recommends following Good Clinical Practice (GCP) by using microscopy to confirm an elevated RBC Count in a Urine sample as per CLSI GP16-3A. 29(4) and CAP Checklist on Urinalysis. Product will continue to ship to not impact critical patient care.
In vitro diagnostic devices, used greater than 6 months after date of manufacture, may have increased concentration times of greater than the specification of 3 hours and up to 24 hours, which may lead to delays in carrying out the subsequent analysis tests.
A manufacturing defect on the amplifier board of the Flow Cytometers my cause signal loss and/or signal drifting resulting in absence of data or a population shift in the data plots. This can impact patient results for any application.
A software issue may cause the liquid volume detection entries for 2nd and 3rd reagent bottles to be overwritten with the data from the 1st reagent bottle. Tosoh has released software version 2.51 (or later) to fix the problem. This failure is mitigated by the hardware design. The consequence would be a slight delay in testing patient sample or reporting patient result.
After running the calibration curve, the recovery of the IgM kit control results decrease demonstrating a reduction in accuracy within the same day; control results may be out of range one day after running the assay calibration curve. Samples have been shown to be similarly affected.
Zhejiang Focus-On Import & Export, of Binjiang District, China
π Consumer Products Class II - Moderate
Young children can fall into buckets and drown. The buckets do not have the required cautionary label warning of the drowning risk posed to young children.