Patient monitors may restart due to network overload caused by network configuration.
All Recalls
95,123 total recalls in our database
CIVCO Needle Guide Starter Kit, REF 674-038, Reusable non-sterile bracket with sterile needle guide (1cm, 1.5cm, 2cm depths) with (10.2 x 147cm) CIV-Flex cover for use with SonoSite L25 series transducers, 21 GAUGE, Rx. The firm name on the label is CIVCO, Kalona, IA..
CIVCO Medical Instruments Co.
Comprehensive Mini Stem, Item No. 113631:
Zimmer Biomet
Zimmer Biomet is conducting a lot specific medical device recall for two lots of the Orthopedic Salvage System (OSS) and the Comprehensive Shoulder System. The investigation determined that the Mini Humeral Stem is potentially labeled as the Bowed IM Stem lot. The associated risk of this product issue is a potential extension of surgery <30 minutes if a replacement is readily available.
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
CIVCO Needle Guide Bracket, REF P03122-06, Reusable non-sterile bracket for use with SonoSite L25/L25e transducers, Mfg by: CIVCO, Kalona, IA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
This voluntary recall is being conducted due to incorrect packaging. In one lot of Salem Sump Dual Lumen Stomach Tubes the 10Fr product was packaged in a 16Fr pouch. The use of a 10 Fr Salem Sump Tube when a 16Fr is needed could result in decreased fluid collection, with the potential to affect patient care, particularly in an emergency situation where the correct size is not readily available.
CIVCO Needle Guide Bracket, REF numbers M001457100, P06013-10, and P03122-08, reusable non-sterile single-angle bracket for use with SonoSite L25 series tranducers, RX. The firm name on the label is CIVCO, Kalona, IA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
CIVCO Needle Guide Starter Kit, REF numbers P12920-01 and P12920-02, Reusable non-sterile bracket with sterile (1.5cm, 2.5cm, 3.5cm, 4.5cm depths) needle guides (5), (10.2 x 147cm) CIV-Flex covers ((5) for use with SonoSite L25 series transducers, 18 GAUGE, Rx.
CIVCO Medical Instruments Co.
The firm has become aware of a potential issue that may affect the mixing motor drive harness and mixing home detection sensor relay harness on your AIA-900 Analyzer. It is possible that the harnesses could rub against the Plarail chain during operation and result in premature wear. If observed, this issue will cause the following system error to be reported D.lane Mix home overrun error [4093] . If observed, this issue will cause the AIA-900 Analyzer not to function and the D.lane Mix home overrun error [4093] will be reported by the system. Replacement of the motor drive harness and mixing home detection sensor relay harness will mitigate the potential for premature wear and effectively prevent this issue from causing the system error. If the sensor relay harness fails, the AIA-900 analyzer will cease to function and will be unavailable for use. If this issue occurs, there may be a delay in testing and reporting patient test results. Tosoh is aware of two complaints associated with the unexpected wearing of the harness. There have been no user or patient injuries reported that are related to this device malfunction.
Discovery NM/CT 670 ES, model 5376204-70-57
GE Healthcare
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.
CIVCO Needle Guide Starter Kit, REF numbers P12924-01 and P12924-02, Reusable non-sterile bracket with sterile (1cm, 1.5cm, 2cm depths) needle guides (5), (10.2 x 147cm) CIV-Flex covers (5) for use with SonoSite L25 series transducers, 22 GAUGE, Rx.
CIVCO Medical Instruments Co.
CIVCO Needle Guide Bracket, REF 674-057-10, Reusable non-sterile bracket for use with FUJIFILM SonoSite, Inc. L25 series tranducers, Rx.
CIVCO Medical Instruments Co.
HeartStart MRx, Rechargeable Li-Ion (lithium ion battery), REF: M3538A, which is a component of HeartStart MRx Monitor/Defibrillators.
Philips North America
Affected lithium-ion batteries may contain a defective component (Thermal Cut-off or TCO). Should the component fail, the battery will no longer charge or deliver power to the Monitor/Defibrillator, which may prevent operation of the device if it is not connected to AC or DC power. Should a second, unaffected battery be present in the Monitor/Defibrillator, the battery failure may go unnoticed.
Children's soft books
Unknown Manufacturer
A metal grommet used to attach a fabric flap to a page in the book can detach, posing a choking hazard to young children.
Women's Iron Complete Dietary Supplement (60 caplets)
Nutra Manufacturing, Inc., of Greenville, S.C. (a GNC Company)
The dietary supplement blister packaging is not child resistant, as required by federal law. If ingested by a child, these supplements could cause serious injury or death.
TerraTrike Rambler E.V.O adult tricycles
Forever Machine Industrial Co. Ltd., of Taiwan (frame)
The torque sensor in the wheel can activate unexpectedly and cause the trike to power forward, posing crash and injury hazards.
Infant snowsuits
Unknown Manufacturer
The metal snaps on the snowsuit can detach, posing a choking hazard to young children.
Lotus Foods rice ramen noodle soup cups
Unknown Manufacturer
The paper soup cup's labeling incorrectly instructs consumers to use the microwave for cooking, posing fire and burn hazards.
Children's robes
Unknown Manufacturer
The children's robes fail to meet the federal flammability standard for children's sleepwear, posing a risk of burn injuries to children.
Glass and ceramic drawer knobs
Unknown Manufacturer
The glass and ceramic drawer knobs can break or shatter when pulled, posing a laceration hazard