HVAD Battery Charger units manufactured with wrong inductors.
All Recalls
95,064 total recalls in our database
SonoSite Needle Guide Bracket, REF numbers P06013-06 and P06013-07, Reusable non-sterile bracket for use with SonoSite L25 series transducers.
CIVCO Medical Instruments Co.
Upon implant deployment, the Capsular Tab may not be delivered as the needle is retracted. In this failure mode, the needle is deployed into the prostate and retracted, leaving no implant behind. Failure to deliver a capsular tab may result in a delay in completing a treatment or an inability to complete a treatment for the patient. Use of a recalled device may lead to increased frequency of exiting known risks associated with the product, including bleeding and tissue trauma associated with delivering the needle.
Inspection process at the manufacturer found nonwoven material was used in specific lots of product that may break wile in use during the patient/resident transfer. Patient/resident might fall on the floor. IN consequence, the patient might sustain a major trauma, head injury - damage to tissue that is irreversible or may lead to death.
The most proximal unsegmented electrode of the Deep Brain Stimulation leads, may be constructed with MP35N instead of the required 90/10 platinum-iridium.
Patient monitors may restart due to network overload caused by network configuration.
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Xcela Power Injectable PICC with PASV (XCELA PASV 4F SL 55CM Catheter Kit CE PG), Cat no. 55-711
Angiodynamics Inc. (Navilyst Medical Inc.)
Specific lots of Xcela Power Injectable and BioFlo PICCs are labelled with incorrect expiration dates. The labelled expiration date extends the shelf life of the product beyond the date supported by validation testing. Although the product under recall has not yet reached its validated expiry date, removal of the affected product is warranted to prevent use beyond the validated expiry date.
Patient monitors may restart due to network overload caused by network configuration.
CIVCO Needle Guide Bracket, REF 674-057, Reusable non-sterile bracket for use with SonoSite L25 series tranducers, Rx.
CIVCO Medical Instruments Co.
CIVCO Needle Guide Starter Kit, REF 674-039, Reusable non-sterile bracket with sterile needle guides (1cm, 1.5cm, 2cm depths) with (10.2 x 147cm) CIV-Flex cover for use with SonoSite L25 series transducers, 22 GAUGE, Rx.
CIVCO Medical Instruments Co.
There is a potential for an undetected over temperature condition on the trans-esophageal (TEE) probe leading to a possible burn injury.
A percentage of the products shipped in Smartscope single-box packaging may exhibit a broken package seal, thus having compromised sterility. There have been two reports of compromised sterility; however, all products packaged in Smartscope single-box packaging are at risk of a compromised product seal. Broken seal pouches compromise sterility and could increase the risk of infection if used.
CIVCO Needle Guide Starter Kit, REF numbers P12922-01 and P12922-02, Reusable non-sterile bracket with sterile (1cm, 1.5cm, 2cm depths) needle guides (5), (10.2 x 147cm) CIV-Flex covers (5) for use with SonoSite L25 series transducers, 21 GAUGE, Rx.
CIVCO Medical Instruments Co.
SonoSite L25 Bracket Kit containing 1 Reusable non-sterile bracket, REF numbers P03122-01, P03122-02, P03122-03, P03122-04, and P03122-05, Mfg for: SonoSite, Inc., Bothell, WA.
CIVCO Medical Instruments Co.
The needle guide bracket may puncture the sterile cover applied over the transducer and bracket during assembly.
CIVCO Needle Guide Starter Kit, REF 674-037, Reusable non-sterile bracket with sterile needle guides (1.5cm, 2.5cm, 3.5cm, 4.5 depths) with (10.2 x 147cm) CIV-Flex cover for use with SonoSite L25 series transducers, 18 GAUGE, Rx.
CIVCO Medical Instruments Co.
Patient monitors may restart due to network overload caused by network configuration.
Due to some cards in this lot experiencing variations in wave form, resulting in some cards with shorter or longer times than expected
Tosoh Bioscience has become aware of a potential issue with the AIA-900 and AIA-2000 analyzers that could result in slippage of the incubator unit contained within the system. If the tension of the timing belt is not adjusted correctly, it is possible for a misalignment that will result in a mechanical error. Tosoh has taken the required actions to inspect the tension of the timing belt during preventative maintenance and correct the instruments where this issue has been observed. If incubator slippage occurs, the unit will cease to function, and a system error will be reported. If this occurs, there may be a delay in testing and reporting patient test results. There have been no reports of injury related to this issue.
Discovery NM/CT 670 Pro, model 5376204-70-54
GE Healthcare
When your system was installed, the Main disconnect switch, EPO button on the A1 panel, and Remote EPO button(s) (outside of the A1 panel) may not have been wired correctly which could result in these not shutting off all power to the entire system as intended, resulting in a potential electric hazard.