95,123 total recalls in our database
AF 550x8ML/H 2ML 15LO 8.7" DUAL CATHS, 4, Product code MT8015XLYK10 CP, Arrow AutoFuser Disposable Pain Control Pump
Arrow International
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Intera 1.5T Omni/Stellar, Model Number 781104. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Ingenia 1.5T S, Model Number 781347. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
MARLIN 1.5T, Model Number 781474. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
0F 550 x 8ML/HR 2ML 60LO CPNB, Product code MT8060XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Arrow International
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
MR-OR 3.0T, Model Number 781438. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
NZM101 Ingenia 1.5T Linac, Model Number 781483. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Achieva 3.0T X, Model Number 781344. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Intera 1.5T Power/Pulsar, Model Number 781105. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications
Intera 1.5T CV, Model Number 781107. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
AF 550 x 8ML/HR 2ML 15LO CPNBMT8015XL CP, Product code MT8015XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Arrow International
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
Ingenia 3.0T CX, Model Number 781271. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
There is a potential for difficulty in sliding down the retention mechanism on the sets manufactured with a specific extension spring lot.
The product is contaminated with Pseudomonas.
There is a potential for difficulty in sliding down the retention mechanism on the sets manufactured with a specific extension spring lot.
Ingenia 1.5T CX, Model Number 781261. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
Multiva 1.5T, Model Number 781076. Nuclear Magnetic Resonance Imaging System
Philips Medical Systems Nederlands
There is a potential for the ceiling speaker(s) to fall from their position and be pulled to the MR system by the magnetic field.
The Vitality Shear-off Set Screw is not threading properly with the mating tulip head of the Vitality screws, hooks and connectors due to a manufacturing issue.These devices are indicated as an adjunct to fusion for various indications