Possibility of balloon rupture in intra-aortic occlusion devices. If the balloon bursts during cardiopulmonary bypass, the heart can fill and warm, the operative site may be obscured, and the device will need to be exchanged or operative strategy would need to change, including placement of an external cross-clamp, conversion to an open procedure, or performing the procedure under fibrillation.

May 14, 2019 Other Medical Devices Nationwide View Details →

Weld attaching the air heater assembly to the main sterilizer unit has shown signs of fatigue/failure

Apr 18, 2019 Other Medical Devices Nationwide View Details →

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

May 20, 2019 Surgical Instruments Nationwide View Details →

There has been a reported incident of a Distar CRT monitor that has fallen off the monitor wall mount related to product aging.

Mar 29, 2019 Infusion Pumps Nationwide View Details →

Possibility that inlet pump hose of the Washer Disinfector Getinge 86-Series (S-8666, S-8668) might collapse during use result in low water pressure within the circulation system and, in the worst case scenario, can result in the contents of the Washer Disinfector not being cleaned properly after finished process.

May 20, 2019 Surgical Instruments Nationwide View Details →

An intermittent failure which causes a blender initialization fault upon start-up or during an EPGS calibration attempt leads to the Central Control Monitor control sliders being disabled, such that the gas system can only be controlled by the local control knobs.

Oct 17, 2018 Other Medical Devices Nationwide View Details →

Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values

Apr 29, 2019 Diagnostic Equipment Nationwide View Details →

There is a potential for an unexpected increase in stimulation during InterStim programming with the A10 Clinician Application (on Medtronic's smart programmer).

May 15, 2019 Other Medical Devices Nationwide View Details →

Infusion Pump Data Storage Accuracy-Data from the BBraun Space LAN or Arcomed UniqueDoc infusion pumps transmitted via the HL7 output interface through the Patient Information Center iX using a LAN driver may be recorded in the patients chart or electronic medical record at exactly 100 times the actual bolus rate, infusion rate and total volume values

Apr 29, 2019 Diagnostic Equipment Nationwide View Details →