Administration sets assembled with inverted pinch clamp cause the fluid from the distal end of the administration set to be pumped to the proximal end of the tubing when secondary set is used with the primary set. Thus, instead of delivering fluid to the patient, blood may be drawn from the patient if there is space in the drug container to hold extra volume of fluid, under the condition that the reversed pinch clamp is not identified
All Recalls
95,050 total recalls in our database
Monoject Standard Hypodermic Needle, 18 G x 1-1/2" Item code: 8881200078
Cardinal Health 200
Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product
Terumo Sur-vet Needle, 25G x 5/8", Product Code 100220
Terumo Medical
There is a potential for a loss of package integrity that may compromise the sterility of the product.
U3-3 Solar Blue, Laptop integrated power supply used with the Solar Blue Urodynamics for assessing lower urinary tract dysfunctions
LABORIE MEDICAL TECHNOLOGIES
Power supply unit (PS-SBL) not correctly labeled, and insulation material used at the low voltage DC end does not meet all applicable flammability rating requirements. may lead to a potential fire hazard
AXIOM Artis BC, Model Number 5904649
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
RX Imola, Model Nos. RX4900
Randox Laboratories, Limited
There was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
AXIOM Artis FC, Model Number 5904433
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.
Monoject Standard Hypodermic Needle, 16 G x 1-1/2" Item code: 8881200037
Cardinal Health 200
Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product
AXIOM Artis dBC, Model Number 5917054
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
AXIOM Artis MP, Model Number 5904466
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Potential for tube leakage
AXIOM Artis TC, Model Number 7728350
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
IGuide System, software versions: 2.2.0, 2.2.1, 2.2.2 Product Usage: Linear Accelerator. The intended use of the device is the control of accurate patient positioning with assistance of a 30 Tracking System in a radiotherapy environment.
Medical Intelligence Medizintechnik Gmbh
Sometimes, although the HexaPOD has reached the target position, the Constant Correlation Check (CCC) can be too sensitive and iGUIDE may incorrectly demand a verification scan.
An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.
Monoject" 201 Vet Pak Hypodermic Needle, 16 ga x 1-1/2 Item code: 8881201654
Cardinal Health 200
Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product
An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.
AXIOM Artis dFC, Model Number 7412807
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Artis zee biplane MN, Model Number 10094143
Siemens Medical Solutions USA
A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.
Monoject Blunt Cannula, 19 G x 1-1/2" (1.067 mm x 3.8 cm) Item code: 8881202355
Cardinal Health 200
Manufacturing defect was found for the cartridge component, which could compromise the sterility barrier of the product