When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Insulin pens may have been incorrectly assembled therefore users could experience the following: either the insulin cartridge will not fit into the cartridge holder; or the cartridge holder may be difficult to remove from the insulin pen. If the issue is present, and if the user does not have backup insulin they may experience temporary hyperglycemia.
The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.
Philips has become aware of the potential for unintended radiation exposure with Zenition 70 systems with Vascular Extension option and a Wireless Foot Switch.
The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
There have been at least 12 complaints from users as of March 3, 2025, reporting that they identified the ceramic C-Ring to be broken in half during use of the device. The ceramic C-Ring has a potential to break while in use. Maquet/Getinge determined that the issue was limited to the units of VH-4000 and VH-4001 manufactured with ceramic C-Ring, and the ceramic C-Ring was implemented in the Hemopro 2 devices from November 5, 2024, through February 14, 2025.
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.
Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.
Devices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.