Complaints were received on the product, such as the re-breather bag would not fill/inflate or the oxygen line pops off after the oxygen is turned on due to a clogged port.
All Recalls
95,114 total recalls in our database
ECHELON FLEX Powered Plus ENDOPATH 60mm Stapler - 280mm shaft, Product Code PCEE60A
Ethicon Endo-Surgery
The staplers may contain an out of specification anvil component within the jaw of the device. This condition may lead to malformed staples, which can compromise staple line integrity. If the staple line is compromised, there is a potential risk of prolonged surgery, postoperative anastomotic leak, hemorrhage, hemorrhagic shock, additional surgical intervention, or death.
DEMO No Needle Corrugated Carton 20 U/D, Part Number 2260-03
Valeritas, Incorporated
There is a potential for "No Needle Demo Units" to contain a needle.
TherMax Blood Warmer Unit
Baxter Healthcare
TherMax Blood Warmers may not be in compliance with an electrical safety standard, which requires a protective earth connection that can sustain a current of 25 amps for 10 seconds.
Implant Direct Scan Adapter Assembly Replant platform for the following part numbers: 6035-09PT-M (3.5mmD)/UDI : 10841307123934,;6043-09PT-M (4.3mmD)/UDI : 10841307123941; and 6050-09PT-M (5.00mmD)/ UDI: 10841307123958
Implant Direct Sybron Manufacturing
The action is being initiated due to incorrect assembly of the scan adapter . The result of the incorrect assembly could cause the clinician not to be able to capture the position and orientation of a dental implant(s) or lab analog during the digital scanning process which could result in a delay for the patient.
A&E Medical Thorecon Plating System Kits Model No. 94-1400-04. For fracture fixation or reconstruction of the sternum.
Alto Development
Some Thorecon disposable tensioners supplied in different Plating System Kits exhibit a molding defect which occludes the through-channel for cable tensioning, rendering the tensioner inoperable.
PORT ACCESS INFUSION KIT - 1" NEEDLE Product Code: DYNDC2817
Centurion Medical Products
Supplier initiated recall of the GRIPPER Needles
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
Programmers & Remote Monitoring Software Apps: 29901 CareLink Encore Programmer
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is a potential for Medtronic programmer and remote monitoring software applications to display an inaccurate remaining longevity estimate for a subset of implanted cardiac device models. This issue does not impact device functionality.
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
Binding can occur between the attachment bolt and intramedullary nail, which may lead to delays in surgical procedures.
NIBP valve assembly board subject to this recall includes enhancements that are not included in the current marketing clearance:1. Anti-motion feature 2. NIBP hose blockage identification function
BD Vacutainer Eclipse Blood Collection Needles 22Gx1.25, cat. no. 368608
Becton Dickinson & Company
BD PAS received customer complaints indicating the absence of the bevel on the non-patient needle end of the product, causing blood leakage.
Maintenance Flush Kit/Blood Draw Product Code: DT19335
Centurion Medical Products
Supplier initiated recall of the GRIPPER Needles
DEMO No Needle Corrugated Carton 30 U/D, Part Number 2568-02
Valeritas, Incorporated
There is a potential for "No Needle Demo Units" to contain a needle.
DEMO No Needle Corrugated Carton 40 U/D, Part Number 2260-01
Valeritas, Incorporated
There is a potential for "No Needle Demo Units" to contain a needle.
Dario Blood Glucose Monitoring System, Model No. 1015-05, running Android software versions 4.3.0-4.3.2
LABSTYLE INNOVATIONS
The Dario-Blood Glucose Tracker Android App versions 4.3.0-4.3.2 may experience duplicate logging of a blood glucose level reading.
The product labels (pouch and carton) for products contain an incorrect value in millimeters (0.036mm) for the Guidewire Outer Diameter dimension. The inches dimension (0.014in) on the label is correct. The correct values are 0.014in (0.36mm). 0.036mm represents a typographical error.
The contract sterilizer did not have adequate packaging data to support the 2-year shelf life.