The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Jan 4, 2019 Implants & Prosthetics Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.

Oct 28, 2019 Diagnostic Equipment Nationwide View Details →

Incomplete seals on the sterile package may compromise the sterility

Dec 11, 2019 Infusion Pumps Nationwide View Details →

Incomplete seals on the sterile package may compromise the sterility

Dec 11, 2019 Other Medical Devices Nationwide View Details →

Following Customers complaints, investigations have been initiated on potentially false results obtained on VIDAS 3 with an expired calibration. Indeed, the calibrations of assays were valid in the calibration menu whereas in fact the calibrations were expired and no alarm displayed to warn the users about the expiration date of the calibrations. Software computes results of analysis for assays with expired calibrations. The anomaly is due to an incorrect update of the calibration status by the software.

Oct 22, 2019 Other Medical Devices View Details →