The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
All Recalls
95,114 total recalls in our database
MultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038
Philips North America
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
Poly G Integris H5000, System code 72246
Philips North America
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
DS-0022 QUICK START DISTRACTION SCREW 12MM 5/BOX STERILE - Product Usage: These products are intended to be used by trained surgeons and are designed for use with TSI distractors and screwdrivers.
TeDan Surgical Innovations
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product to non sterile conditions. This situation presents a risk of infection to patients.
Rheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A. , UDI: 05051700018319
The Binding Site Group
It has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal results which may lead to unnecessary additional serological testing.
TRUMATCH Orthognathics - Genioplasty Surgical Kit, Add-on, Catalog Number: SD980.013
Materialise N.V.
Custom surgical kits contain a plate different than indicated by the package labeling.
Screws provided in the device package are 6 mm longer than intended.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product to non sterile conditions. This situation presents a risk of infection to patients.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
Medtronic CareLink Encore 29901 Programmer. Used to interrogate and program Medtronic and Vitatron implantable devices, such as pacemaker/ICD/CRT.
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.
Boston Scientific released a new Software Update Application that was downloaded to four (4) distributed Model 3300 programmers. The Software Update Application then directed these programmers to download software applications that have not been approved by the US Food and Drug Administration.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
Allura CV20, system code 722031
Philips North America
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
DS-0024 QUICK START DISTRACTION SCREW 14MM 5/BOX STERILE - Product Usage: These products are intended to be used by trained surgeons and are designed for use with TSI distractors and screwdrivers.
TeDan Surgical Innovations
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product to non sterile conditions. This situation presents a risk of infection to patients.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
The medical procedure trays contain one or more recalled surgical gowns. The gowns were manufactured at locations that did not maintain proper environmental conditions. There is no assurance that the surgical gowns are sterile. An inadequately sterilized surgical gown could compromise a sterile field and increase the risk of a surgical site infection. Due to the configuration of the trays, the recalled gowns may come into contact with other components within the tray posing a risk of cross contamination.
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
LEGION PS High Flex XLPE size 5-6 9mm Articular Insert
Smith & Nephew
LEGION PS High Flex XLPE Articular Insert was packaged with an incorrect part
Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.