πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’

In certain scenarios, the Omnipod DASH PDM may suggest an inaccurate bolus amount based on a blood glucose value that is more than 10 minutes old when the user does not exit the bolus calculator as designed or when a system alarm interrupts a bolus calculation. If the user delivers the bolus, this may lead to hypoglycemia or hyperglycemia.

Feb 13, 2020 Infusion Pumps Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.

Aug 27, 2019 Other Medical Devices Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’

Due to an error in the programming of the Control CODE CHIP module for this lot, a failing result is inappropriately displayed for AMP and mAMP. This issue results in the Triage Meter displaying a failing control result for the AMP and mAMP tests when the control did not fail.

Apr 9, 2019 Other Medical Devices Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’

Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy

Feb 13, 2020 Diagnostic Equipment Nationwide View Details β†’

Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.

Oct 1, 2019 Implants & Prosthetics Nationwide View Details β†’

Firm has identified that the product may be at risk of incorrectly containing a tunneler with a barb tip meant to attach to a 6Fr catheter instead of the correct barb tip for a 9.6Fr catheter. This can result in a prolongation of operation.

Oct 1, 2019 Implants & Prosthetics Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

Samples of 114gm tubes of Lemon Prep, collected during a Food and Drug Administration inspection on October 15, 2019 were tested and found to be contaminated with Burkholderia cepacia. Mavidon is recalling all products manufactured at their facility.

Dec 23, 2019 Other Medical Devices Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’

FlipCutter III Drill

T.A.G. MEDICAL PRODUCTS CORPORATION

πŸ₯ Medical Devices Class I - Dangerous

There is potential for the blister tray to be cracked causing the sterile barrier to be compromised and potential for particulate within the sterile barrier.

Jan 20, 2020 Other Medical Devices View Details β†’

Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy

Feb 13, 2020 Diagnostic Equipment Nationwide View Details β†’

Due to physical limitations of the treatment machine, the maximum field opening for Siemens Virtual Wedge fields recommended by the LINAC vendor will be more restricted in the wedged direction than the maximum field opening without a wedge. Delivering a larger field opening with a virtual wedge will not result in a wedged field shape over the entire field. Instead, a high dose plateau region will be the result at the toe side of the field beyond the recommended field size.

Mar 4, 2020 Other Medical Devices Nationwide View Details β†’

Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy

Feb 13, 2020 Diagnostic Equipment Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

The action is being initiated due to potential cybersecurity vulnerabilities related to a series of insulin pumps that are designed to communicate using a wireless radio frequency (RF). An unauthorized person could potentially connect wirelessly to a nearby insulin pump to change settings and control insulin delivery.

Jun 27, 2019 Infusion Pumps Nationwide View Details β†’
πŸ₯ Medical Devices Class I - Dangerous

Instructions sent in a letter or in instructions for use for disinfecting blood glucose and/or ketone meters were distributed but not cleared by the U.S. FDA, and devices were not cleared for multiple-patient use. Instructions could lead to inappropriate meter disinfection or improper handling of contaminated meters, which may cause infection or exacerbation of an existing infection.

Nov 14, 2019 Diagnostic Equipment Nationwide View Details β†’