Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
All Recalls
95,114 total recalls in our database
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
Fentanyl Reagent Kit, Product N. CC444, Contains: FENT R1 Reagent and FENT R2 Reagent
Carolina Liquid Chemistries
Labeled for diagnostic use and sold to clinical labs without a 510(k). Firm corrected labeling to remove clinical diagnostic language and notified customers that products are labeled for "Forensic Use Only" and are not FDA Cleared and not to be used for clinical purposes
ABL90 FLEX PLUS Analyzer, Model Number 393-092 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Radiometer Medical ApS
The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
Sensor Nitinol Wire with Hydrophilic Tip .035 3CM FLEX /150CM UPN: M0066703061 box 5 UPN: M0066703060 single unit
Boston Scientific
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
There is a potential for the occurrence of false positive OXSF results for some Staphylococcus aureus strains in conjunction with the VITEK 2 AST-P652 test card (ref. 421857, lot 8021232403). .
ABL90 FLEX Analyzer, Model Numbers 393-090, with software versions below 3.4MR2 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.
Radiometer Medical ApS
The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.
There were unknown bioburden levels prior to sterilization, which may affect sterility of the product.
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device
Space between the device and the packaging permits excessive movement which could potentially lead to the displacement of the anchor s tip protector allowing the tip of the anchor to create a sterile breach.
Temperature Sensing 100% Silicone Foley Catheter, Part Number 102201101063MD - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.
Degania Medical Devices Pvt.
Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.
Reports have been received about the transcutaneous monitoring system display flickering or monitor becoming non responsive and beeping. This may result in delayed patient monitoring as well as potentially compromise the product performance.
A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine
Fresenius 2008T BlueStar Hemodialysis Machine Part Number Description: 191124 2008T HD SYS. CDX BLUEST AR; 191126 2008T HD SYS. CDX W/BIBAG BLUEST AR; 191128 2008T HD SYS. W/O CDX BLUEST AR; 191130 2008T HD SYS. W/O CDX W/BIBAG BLUEST AR
Fresenius Medical Care Holdings
A "Remove USB Device 2" false alarm may be displayed when no USB device or a non-powered USB device is connected to the USB port on the rear of the machine
Rhythm Xience Guider Catheter Introducer with Lancer Intergrated Dilator/Transseptal Needle, Ref 112852, Sterile EO, RxOnly
Acutus Medical
It has been determined that the manufacturing process may have left foreign material particulates on the finished device. This could result in a range of event from procedure delay to pulmonary embolism.
ARCHITECT HAVAB G Reagent Kit. List Number 6L27 25.
Abbott Gmbh & Co. KG
An erroneous concentration value for hepatitis A virus was utilized during manufacture, leading to a decreased amount of antigen used for the impacted lots.
Medivators is recalling Endo SmartCap Tubing due to a manufacturing anomaly that may present during procedures as inadequate insufflation, presence of air bubbles inside the endo smartcap tubing, insufficient water to rinse the endoscope lens, or water leaking from the air/water valve.
ACUVUE OASYS for ASTIGMATISM Brand Contact Lenses. Lot no. B00NKM0G. Base Curve 8.6, Refractive Power -3.00D-1.25/90.
Johnson & Johnson Vision Care
Package integrity failure observed during routine shift package integrity testing at an external re-packager.
Cross-reactivity testing may not meet the claims indicated in the Opiate 2000 ng (OP2) Instructions For Use. The presence of certain compounds in urine may result in false positive or false negative patient test results.
Sensor Nitinol Wire with Hydrophilic Tip .038 3CM FLEX STR/150CM UPN: M006670309 1box 5 UPN: M0066703090 single unit
Boston Scientific
An increase in the rate of complaints for difficulty or inability to track over the guidewire, may result in a procedural delay due to the need to exchange the affected device