Sterile Custom Packs to be used in surgical procedures.
American Contract Systems
The surgical gowns were manufactured in a facility that is not registered by the FDA.
95,114 total recalls in our database
American Contract Systems
The surgical gowns were manufactured in a facility that is not registered by the FDA.
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
Umano Medical
Potential Hazard of Unsupervised Patient being out of Bed and/or Potential Patient Fall.
American Contract Systems
The surgical gowns were manufactured in a facility that is not registered by the FDA.
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
Potential disconnection of tubing set.
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays: Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, and TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
Radiometer Medical ApS
The SafeCLINITUBES blood sampler may have insufficient aspiration capabilities ultimately resulting in lost samples.
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
The Pressure Monitoring Set and Tray may contain excess coating material on the outside and/or interior of the needle component which could cause occlusion.
The recalling firm characterized the potential interference effects of biotin (up to 1,200 ng/mL) on all biotin-mediated Access assays. The following six assays (Free T3-REF A13422, Total T3-REF 33830, Free T4-REF 33880, GI Mon-387687, Tg-REF 33860, TgAbII-REF A32898 displayed significant interference (defined as a result bias greater than +/-10%) at 1,200 ng/mL of biotin.
Medtronic
Internal testing results showed endotoxin-related test result did not meet specifications for implantable insulin pump. If pumps have higher than allowed endotoxin levels, patients may have an effect of fever, hypotension, anaphylactic shock, hyperglycemia, or diabetic ketoacidosis (DKA), potentially followed by secondary multi-organ failure (primarily renal and hepatic) and/or death.
Incorrect shelf-life for bone void filler kits; the product may not maintain its setting characteristics for the labeled shelf-life duration.
Nobel Biocare Usa
The top label for dental implant may incorrectly indicate the implant length. Specifically, the label may contain 11.5mm for the length, which is shorter than the actual length of the implant (15mm). An incorrect sized implant may cause patient discomfort, permanent nerve damage and may require an additional surgery.
Siemens Medical Solutions USA
Using Quantification Application (QVA/QCA) on DSA images may lead to a failure in vessel detection or incorrect quantification of vessel detection and result in an incorrect diagnosis and inappropriate treatment of the patient
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.
There are reports of inadequate locking of the extended tab screws. Complaints were reported where the closure top backed out of the screw post-operatively.