Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
There is a potential that hand control units may cause the tabletop to slide in the opposite direction from that selected by the user. This could result in a delay in procedure.
There is a potential for hydraulic leakage at the inclination function to occur resulting in an unintended Trendelenburg or anti-Trendelenburg movement of the table top depending on patient positioning which may lead to the patient having to be transferred to another OR table causing a delay in the procedure.
Arietta 850 software version 4.0.0., 4.0.1, and 4.0.2 has an error in the focus point and transducer aperture settings in the SWE function. When the affected software with SWE function is used in conjunction with the C252 probe, this error can result in out of specification MI/TI acoustic output. The MI/TI acoustic output is higher than regulatory limits.
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
There is a potential for the hemostatic valve to dislodge while introducing the dilator or device into the sheath. This could result in minor bleeding and/or embolism
Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.
Due to the Penetration Depth Switch used to select the desired penetration depth of the biopsy needle for soft tissue biopsies, detaches from the device during use.
Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.
Erroneously high FeNO levels which may contribute to premature discontinuation of effective asthma treatment and persistent underlying disease. Two issues affecting the FeNO readings: 1) Drift in the calibration gases and 2) Software error. Each of these issues can cause FeNO scores/results to be erroneously high.
bioMΒΏrieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEKΒΏ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEKΒΏ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).
Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.
Variability in the performance of the tubing in the Prismaflex Control Unit ARPS (Automatic Repositioning System) Pump Assembly, may lead to alarm situations during or after a system self-test.
Cardinal Health has determined that a manufacturing variance in a plastic film used to manufacture these products is causing the drapes and leggings to stick , which may prevent the products from being unfolded and applied properly.
The drive belt can break during operation and damage the inner clutch cover assembly and fuel line, which can result in a fuel leak, posing a fire hazard.