Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
All Recalls
95,018 total recalls in our database
Cortoss Bone Augmentation Material 5cc Cartridge, Catalog Number 2101-0005, GTIN 00808232000962
Stryker Spine
The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Exposed to freezing temperatures due to refrigerated truck malfunctions
DJO Surgical EMPOWR 3D Knee Tibial Insert - e-plus, REF: 341-16-711 Left Sz. 11 16mm, CE 0086, Sterile H2O2, UDI: (01)00888912167185
Encore Medical
Packaging mixup; There is a potential that packaging identified as 12mm knee prosthesis contains 16mm knee prosthesis and 16mm knee prosthesis packaging contains 12 mm knee prosthesis.
Software showing wrong results reported, PDF not match HIS.
Medtronic Reveal LINQ LNQ11 / PA96000
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.
A caution statement is being added to remind the user to activate away from face and infant as reports of pack rupture or leaking during activation may resulting in the contents splashing into the user s eyes or onto the infant. Potential harms associated with contact of the liquid inside the pouch are rash, burn and inflammation.
Cell Marque Antibodies, MSH6 (44), Mouse Monoclonal Antibody, REF: 287M-10, IVD, CE, GTIN: 00841683104513
Cell Marque
Product distributed for use as an In-vitro Diagnostics without FDA clearance
Telepack may not power on when using the 3AA battery cradle.
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
Exposed to freezing temperatures due to refrigerated truck malfunctions
Risk of edge-loading and premature prosthesis wear is possible in a specific subset of patients with certain implant configurations and surgical implant positioning.
The firm is aware of a issue with ultrasound software that results in unintended video frames being included. This could lead to an incorrect automated ejection fraction (AutoEF) and image quality score (IQS) estimates.
026-20036A Liberty Drain Line, Luer Tap For patients with acute and chronic end stage renal disease going under PD in healthcare or home setting
Fresenius Medical Care Holdings
Exposed to freezing temperatures due to refrigerated truck malfunctions
Stimulen Collagen Moisturizing Lotion, ST9555B
Southwest Technologies
Dose audit failure after sterilization process
It has been identified that ventilators equipped with High Flow Therapy (Software Versions 3.00 and 3.10), that includes a designed safety mechanism that limits the amount of pressure that can be delivered to a patient, in situations where the system pressure reaches the maximum limit and sounds the "Cannot Reach Target Flow" alarm , the system will reduce pressure and simultaneously decreasing the flow rate to a level below what was set by the clinicians. If the reason for the increase in pressure is not resolved, the ventilator will continue to provide the lower flow rate, which result in patients experiencing oxygen desaturation which can be characterized as moderate or severe hypoxemia.
Cortoss Bone Augmentation Material 10cc Cartridge, Catalog Number 2101-0010, GTIN 00808232000979
Stryker Spine
The products experienced a 12-hour excursion from the required refrigerated temperature conditions at the distribution center prior to distribution.