🏥 Medical Devices Class I - Dangerous

The firm has received customer reports of performance issues with the affected lot, including increased imprecision, out of range quality controls, and prolonged sample results. If quality controls are not performed or do not pass for each vial of reagent, there is a potential risk of reporting an erroneous result.

Oct 13, 2021 Diagnostic Equipment Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient rescans, and/or the need for additional contrast media may occur

Aug 30, 2021 Other Medical Devices Nationwide View Details →

Due to calibration curves producing a higher activity than routinely expected and that there may be a decrease in separation between calibrator point 5 and calibrator point 6, which can be exacerbated with calibrator 6 activity reaches the optical ceiling of the SPA+ analyzer.

Oct 15, 2021 Other Medical Devices View Details →

Incorrect expiration date on its package labeling (shipper box and inner carton label).

Sep 21, 2021 Surgical Instruments Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

Significant wear of the Monitor Bracket Assembly shaft may go unrecognized could lead to the Monitor Bracket Assembly detaching from the Suspension Arm, and resulting in injury to patient, operator, bystander, or service person located in the path of a falling part

May 1, 2021 Other Medical Devices Nationwide View Details →

Twelve software-related issues affecting software version 9.41 and earlier were identified. The issues include bypassing error codes, incorrect calibration, loss of module communication leading to overfilling wash buffer containers, missing QC flags, signal spikes, abnormal optics reads, and incorrect sample aspiration. The issues may lead to inaccurate results with an unknown bias direction and magnitude for assays measuring cardiac, metabolic (including diabetes), toxicology, prenatal, therapeutic drug monitoring, infectious disease, oncology, hepatic, pancreatic, hematological, endocrinologic, inflammatory analytes.

Sep 29, 2021 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

software syngo.CT VB20 in the installed base, with or without syngo.CT software versions VB20_SP1, SP2, SP3, or SP4, may result in workflow interruptions resulting in the delay in diagnosis, the necessity for patient rescans, and/or the need for additional contrast media may occur

Aug 30, 2021 Other Medical Devices Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

Inadequate/inappropriate Set-Up Instructions, which may result in delay in treatment and an unreasonable risk of harm to the patient

Oct 22, 2021 Other Medical Devices Nationwide View Details →

If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure

Oct 20, 2021 Diagnostic Equipment Nationwide View Details →

The processing of collected episode data may fail due to a software error.

Oct 12, 2021 Implants & Prosthetics Nationwide View Details →