Due to a calculation error in the measurement when using 2D trace (manual trace) tool. The trace circumference value is overestimated and may potentially result in misdiagnosis of a patient's condition.
The European version of the DiaTrust COVID-19 Ag Rapid Test is not approved for marketing in the U.S. but was illegally distributed in the U.S. The packaging and labeling of the European test kit is different from the US test kit and is NOT authorized for distribution in the US.
COVID test kits were offered for sale and distributed to consumers in the United States without marketing approval, clearance, or authorization from FDA.
The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.
Impacted lot may leak inside the ARIES instrument. Testing confirmed a leak rate of 0.9%. Leaking cassettes could potentially lead to incorrect results or a delay in reporting results due to an Invalid.
COVID test kits were offered for sale and distributed to consumers in the United States without marketing approval, clearance, or authorization from FDA.
The firm has received complaints of the OXY-1 System Console experiencing a power interruption while in use, which disrupts therapy delivered to a patient on support. Disruption of therapy could lead to prolonged hypoxia.
There is the potential for reduced or non-delivery of medication, in some cases without alerting the user via pump alarm. This may occur as a result of incorrect administration set setup and/or incomplete resolution of upstream occlusion alarms when using Spectrum V8 and Spectrum IQ infusion pumps.