Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
All Recalls
95,018 total recalls in our database
Reports have been received that after normal or routine lateral forces are applied to the head fixation device, the split block clamp may deflect (rotationally around the axis) and not return to within acceptable distance from its prior position after the force is removed. In order to reduce deflection at the table mount location, users have tightened the hand knob screw to a degree that it is then difficult to untighten it later in the clinical workflow. Rotational movement at the table attachment point could result in HFD movement going unnoticed, which could result in a situation where surgical or medical intervention is necessary, a situation where navigation needs to be reestablished, or a situation resulting in extended time under anesthesia for the patient. It may also be difficult to loosen the mount clamp component after use, potentially leading to ergonomic injury or a workaround during equipment teardown.
Labeling change to due to the risk of severe anaphylactoid reactions including shock in patients treated with the Liposorber LA-15 LDL Adsorption Column under concomitant ACE inhibitor medication treatment.
Firm learned of the potential for cracks to form in the conductivity sensors during use, which may lead to a leak in the dialysate circuit. This could result in excessive ultrafiltration or insufficient ultrafiltration depending on whether the dialysis machine is equipped with a Dialysis Fluid filter.
Due to an increase in complaint trend for leaks and intermittent/loose connections.
The IFU for the Autoclavable Temperature Probes: Part # 453564635891, providing in the packaging for the following product model #s: 21075A, 21076A, 21078A, 989803162621, 989803162631, 989803162641.
Philips North America
Update to instructions for use regarding the cleaning and disinfection process for the reusable probes.
ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Floppy
Boston Scientific
Pouch seals may be open, compromising sterility.
INDEFLATOR Plus 30 - REF 1000183 WPL2122277-01 (2019-03-01)
Abbott Vascular
Due to an increase in complaint trend for leaks and intermittent/loose connections.
Two of three complaints were initiated due to the customers repeatedly obtaining No Call results on the Processor SPs in question.
iTotal Identity Posterior Stabilised (PS) Knee Replacement System Model Number: TPS-301-1113-020101
Conformis
Incorrect tibial base plate implant packaged in kit
ROTAWIRE Drive and wireClip Torquer, ROTAWIRE Drive Extra Support
Boston Scientific
Pouch seals may be open, compromising sterility.
Cardinal Health PROTEXIS PI and PI Classic, Surgical Gloves
Cardinal Health 200
Product was distributed without being sterilized.
Nextra Hammertoe Correction System 3.5 Middle and 3.2 Proximal Saw Cut SKU: NX-3532K-SC Orthopedic surgical instrument
Nextremity Solutions
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Atellica CH Fructosamine (Fruc), Siemens Material Number 11097637
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
Nextra Hammertoe Correction System Driver SKU: NX-DR Orthopedic surgical instrument
Nextremity Solutions
Proximal end of the driver is oversized in varying degrees, making it difficult to fully connect with and remove from the proximal Nextra implant, causing delay in surgery
Atellica CH Lithium_2 (LITH_2), Siemens Material Number 11097637
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics Inc. has confirmed the potential for reagent carryover impacting quality control (QC), patient samples, and calibrator results.
Medline Standard PVC Laryngeal Masks, 1 device per pouch, 5 each pouches per box
MEDLINE INDUSTRIES, LP - Northfield
The mask cuff may disconnect from the device's breathing tube.
Cub 2 Enclosed Bed Canopy System
Sensory Medical
Due potential misuse of the product, firm is updating warnings and precautions booklet for products distributed before December 16, 2021.
Breas Ventilator Trolley with the Vivo 50/60/65 and Vivo 45LS ventilator models Article no 007384
Breas Medical
Ventilator mounting brackets for the Vivo 50/60/65 and Vivo 45LS ventilator models can become detached from the Breas Ventilator Trolley, this may cause the ventilator to fall from the trolley and cause injury to a person who is in the immediate proximity
Stanbio TDM/Beta-Hydroxybutyrate Linearity Standards, packaged in boxes containing Levels 1-6, 1 x 4 mL each, REF 2450-604, Medline catalog number S-O2450604, IVD.
MEDLINE INDUSTRIES, LP - Northfield
The product was stored incorrectly due to improper storage controls which may lead to delayed results.