Artis zee floor MN, Model Number 10094142

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q ceiling, Model Number 10848281

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q zeego, Model Number 10848283

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis Q.zen floor, Model Number 10848353

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee biplane, Model Number 10094141

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Artis zee biplane MN, Model Number 10094143

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Software error may result in sporadic problems causing scanning workflow interruptions, unexpected user notifications and image artifacts. Sporadic software errors may also occur during interventional workflows. Results in potential patient issues: Possible rescan , Unexpected X-Ray dose and additional contrast media, Delay in diagnosis, scan abort,and patient rescan

Mar 7, 2022 Other Medical Devices Nationwide View Details →

Five potential software issues affecting Artis pheno and Artis icono systems in combination with a Siemens Healthineers table or Trumpf/MAQUET table (OEM). The potential software errors are related to the block movements function and manual rotation of the flat detector. Depending on the status of the intervention, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where clinical treatment may be delayed until an additional restart is performed

Feb 25, 2022 Diagnostic Equipment Nationwide View Details →

Gallant HF Implantable Cardioverter Defibrillator REF# CDHFA500Q

St. Jude Medical, Cardiac Rhythm Management Division

🏥 Medical Devices Class I - Dangerous

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Mar 10, 2022 Implants & Prosthetics Nationwide View Details →

In instances where custom reference ranges were configured on the device post installation, a possibility to inadvertently apply changes to reference range values was found on software versions 2.59.3 and all earlier versions, which can potentially lead to displaying and printing incorrectly-configured reference range values

Jan 20, 2022 Diagnostic Equipment Nationwide View Details →

During the de-installation of an MR system magnet, if all rigging and hardware that attach to the magnet for transportation are not properly installed and secured, the magnet may fall, leading to potential injury.

Mar 4, 2022 Diagnostic Equipment Nationwide View Details →

Artis Q floor, Model Number 10848280

Siemens Medical Solutions USA

🏥 Medical Devices Class I - Dangerous

If the coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message "TUBE HOT, have a break". Several minutes later the system will block X-ray to prevent further damage and display the message "NO XRAY: TUBE TOO HOT".

Feb 23, 2022 Other Medical Devices Nationwide View Details →

Gallant DR Implantable Cardioverter Defibrillator REF # CDDRA500Q

St. Jude Medical, Cardiac Rhythm Management Division

🏥 Medical Devices Class I - Dangerous

Bluetooth malfunction impacting Gallant devices post-implant. The failure causes the device to enter an inductive-only telemetry mode with a loss of Bluetooth communication. The Bluetooth malfunction results in the inability to remotely monitor these devices and resulted in reduced battery longevity.

Mar 10, 2022 Implants & Prosthetics Nationwide View Details →