Defective Container: Recall of these batches has been initiated due to complaints of difficult to open blister and tablet breaks while opening the blister .
Mckesson Medical-Surgical Inc. Corporate Office
CGMP Deviations: products were stored outside the drug label specifications.
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Medical Modeling
Scapula Marking Guide (anatomically Shaped) was incorrectly designed.
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Spacelabs Healthcare
Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.
One MetaFix size 7 collarless stem from batch 478179 which was incorrectly labelled as a MetaFix size 3 collared stem from batch 485630.
Aesculap Implant Systems
Application error problems: Rebalance scope message: 1.Rebalancing process 2.Usage of Robot Recovery: 3. Autofocus not available:
Olympus Corporation of the Americas
Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Roche Diagnostics Operations
Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.
Guangdong Haiou Medical Apparatus Co.
Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user
Haimen Shengbang Laboratory Equipment Co.
Distributed VTM outside of VTM Guidance and without clearance.
Haimen Shengbang Laboratory Equipment Co.
Distributed VTM outside of VTM Guidance and without clearance.
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).
Haimen Shengbang Laboratory Equipment Co.
Distributed VTM outside of VTM Guidance and without clearance.