All Recalls

95,018 total recalls in our database

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →
🏥 Medical Devices Class I - Dangerous

Software related alarm escalation defect occurs after approximately 25 days of continuous use. Technical alarm escalation for the following alarms, signal loss, all leads off, low battery, SpO2 sensor off, and signal interference, does not occur in the telemetry receiver as specified. Caregiver may not respond to the initial technical alarm and may not be aware of the continuing technical failure.

May 31, 2022 Infusion Pumps Nationwide View Details →
🏥 Medical Devices Class I - Dangerous

Mis-wired component-the improperly wired component can result in additional noise on the power supply output. Noise on the ultrasonic input circuit is not consistent with electromagnetic compatibility (EMC) standards and may decrease the overall reliability of the system and result in delay of treatment

Jul 7, 2022 Other Medical Devices Nationwide View Details →

As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.

Jun 2, 2022 Other Medical Devices Nationwide View Details →

cobas e801 Immunoassay Analyzer

Roche Diagnostics Operations

🏥 Medical Devices Class I - Dangerous

Software issue resulting in signals and sample test results of Pre-Wash tests being impacted. The following assays are considered impacted by the issue: Anti-HAV2, Anti-HBc IgM, IGF-1, Myogloin, Rubella IgG, and Toxo IgG.

Jul 14, 2022 Other Medical Devices Nationwide View Details →

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Jun 10, 2022 Surgical Instruments Nationwide View Details →

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →

Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user

Jun 10, 2022 Surgical Instruments Nationwide View Details →

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

Jun 22, 2022 Implants & Prosthetics View Details →

Footpads for Onewheel GT Electric Skateboards

Future Motion Inc., of Santa Cruz, California

🏠 Consumer Products Class II - Moderate

The footpad can fail to disengage after the rider has dismounted while the board is in motion and the skateboard can unexpectedly continue to operate, posing an injury hazard to bystanders.

Aug 11, 2022 Other Consumer Products Nationwide View Details →