Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
All Recalls
95,102 total recalls in our database
Movable Core Wire Guide (Tapered Tefcor) (OUS ONLY), Guidewire Reference Part Number/GPN TCMTNA-35-145-3 G01678
Cook Incorporated
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Leica Surgical Operating Microscopes, Arveo, Part Number 10448999
Leica Microsystems
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
Short tibial bearing MK4 - Std, STM, Catalog No. SMMLTB02. Knee prosthesis component.
Howmedica Osteonics
A MK4 Tibial Bearing was supplied with the intention to mate with the Tibial Metal Casing in-situ. However, the device in-situ was a MK3 Tibial Metal Casing. The supplied MK4 tibial bearing was not exchangeable for the MK3 tibial bearing. The surgeon needed to complete the surgery by re-assembling and re-inserting the extracted MK3 tibial bearing.
MEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T
MEDLINE INDUSTRIES, LP - Northfield
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
Roadrunner PC Wire Guide (Nimble Floppy), Guidewire RFPC-35-145 G07914
Cook Incorporated
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.
a. DYNAREX Nice'n Fresh WIPE, BABY WET 6X6.75" (140/CN12CN/CS) Catalog # 1313. b. DYNAREX DynaCare WIPE, WET FLUSHABLE ADULT 9X13" (24/PK 24PK/CS) Catalog # 1322. c. DYNAREX dynacare WIPE, WET FLUSHABLE JR 5X8" (42/BX 12BX/CS) Catalog # 1324
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.
a. DYNAREX WeCare PETROLATUM JELLY, WHT 5GM (144/BX 6BX/CS) DYNREX Catalog #1140. b. DYNAREX Dynarex PETROLEUM JELLY, WHITE 4OZ TU(72/CS) DYNREX Catalog #1145
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.
a. Dynarex DRESSING, UNNA BOOT 3"X10' (12/CS), Catalog # 3453. b. Dynarex DRESSING, UNNA BOOT BANDAGE 4"X10' (12/CS) Catalog # 3454
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.
Dimension Magnesium (MG) Flex reagent cartridge Siemens Material Number (SMN)/REF (Catalog Number):10444963/DF57
Siemens Healthcare Diagnostics
Dimension Magnesium (MG) Flex reagent cartridge lots may exhibit imprecision for Quality Control (QC) and produce Abnormal Reaction flags. The issue is not always detected by QC and erroneous results may be produced in the absence of an Abnormal Reaction flag. Based on customer data, imprecision leading to erroneous, unflagged patient results with a bias of -15 to -59% may occur
a. INTEGRA LIFESCIENCE (WND CARE) Shur-Conform Oil Emulsion Oil Emulsion Impregnated Dressing Shur-Conform 3 X 8 Inch Knitted Cellulose Acetate Petrolatum Emulsion Sterile Model Number: 77042. b. INTEGRA LIFESCIENCE (WND CARE) ALGICELL Ag Silver Alginate Dressing Algicell Ag 4 X 5 Inch Rectangle Sterile Model Number: DUP88545
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.
Leica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950
Leica Microsystems
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
MEDLINE CLEARPRO TRACH DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12T b. TRACH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14T
MEDLINE INDUSTRIES, LP - Northfield
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
MERIT MEDICAL SYSTEMS StayFIX FIXATION DEVICE, STAYFIX F/PERCUTAN CATH 5FR-14FR Model Number: 884450023358. Used to secure anaesthesia catheter.
Mckesson Medical-Surgical Inc. Corporate Office
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.
Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Olympus Corporation of the Americas
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA10003 OSSUR Power Knee REF PKA01001L OSSUR Power Knee REF PKA01001T
Ossur Americas
Battery may dislodge from the Power Knee
The kits are being recalled due to a lack of sterility of a spare needle contained within the kits. There is a potential for skin infection or sepsis if the patient is exposed to a non-sterilized, unused needle.
Ambra PACS, Ambra ProViewer. Software for use as a primary diagnostic and analysis tool for diagnostic images.
DICOM Grid, Inc. d/b/a Ambra Health
A race condition between the storage system and services database has the potential to revert edits made to patient information upon first ingestion of a study.
Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21. This device with this specific storage temperature deviation is not likely to cause adverse health consequences.