HipLOC Hip Fracture Fixation System- Hiploc System 130¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-130-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Number: 3222008
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biomet UK Ltd.
Reason for Recall:
HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

HipLOC Hip Fracture Fixation System- Hiploc System 130¿ Plate 38 mm Barrel 4 Hole Plate Item Number: 2110-130-004 Product Usage: The HipLOC Compression Hip Screw may be used primarily for fixation of bone fractures. Specific indications include open reduction and internal fixation of a wide variety of fractures of the proximal femur: intracapsular fractures, intertrochanteric fractures and subtrochanteric fractures

Product Codes/Lot Numbers:

Lot Number: 3222008

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-3261-2018

Related Recalls

HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel

Aug 9, 2018 Implants & Prosthetics Nationwide View Details →

HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel

Aug 9, 2018 Implants & Prosthetics Nationwide View Details →

HipLOC System Plate Barrel due to an undersized bore on the HipLOC plate barrel, can result in the lag screw not fitting through the barrel

Aug 9, 2018 Implants & Prosthetics Nationwide View Details →