IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Class I - DangerousWhat Should You Do?
- Check if you have this product: IntelliVue Patient Monitor MX400, 866060, UDI-DI: 00884838038752; IntelliVue Patient Monitor MX430, 866061, UDI-DI: 00884838057562; IntelliVue Patient Monitor MX450, 866062, UDI-DI: 00884838038769; IntelliVue Patient Monitor MX500, 866064, UDI-DI: 00884838038776; IntelliVue Patient Monitor MX550, 866066, UDI-DI: 00884838038783; Only MX400-550 devices shipped after 26-April-2024 are affected. Please refer to the manufacturing date on the back of your monitor.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America
- Reason for Recall:
- During a production process, Philips became aware of one IntelliVue power supply with a broken ground bolt upon disconnection of the ground cable from the equipotential ground connector. Loss of electrical grounding may negatively affect the devices electromagnetic immunity and emission.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
IntelliVue Patient Monitor MX400 (Model Number 866060), MX430 (Model Number 866061), MX450 (Model Number 866062), MX500 (Model Number 866064), and MX550 (Model Number 866066)
Product Codes/Lot Numbers:
IntelliVue Patient Monitor MX400, 866060, UDI-DI: 00884838038752; IntelliVue Patient Monitor MX430, 866061, UDI-DI: 00884838057562; IntelliVue Patient Monitor MX450, 866062, UDI-DI: 00884838038769; IntelliVue Patient Monitor MX500, 866064, UDI-DI: 00884838038776; IntelliVue Patient Monitor MX550, 866066, UDI-DI: 00884838038783; Only MX400-550 devices shipped after 26-April-2024 are affected. Please refer to the manufacturing date on the back of your monitor.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-3249-2024
Related Recalls
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Philips North America
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
IntelliVue MP90. Product Number: M8010A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.
IntelliVue MP30. Product Number: M8002A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.