SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model UDI-DI: 782113 (01)00884838098909(21); 781270 (01)00884838095083(21); 782129 (01)00884838105805(21). Serial Numbers: 17271 17389 17439 17449 17504 24039 24077 34055 34072 34129 34262 38117 38166 38222
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Philips North America Llc
Reason for Recall:
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

SmartPath to dStream for XR and 3.0T- For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: 1) 782113; 2) 781270; 3) 782129

Product Codes/Lot Numbers:

Model UDI-DI: 782113 (01)00884838098909(21); 781270 (01)00884838095083(21); 782129 (01)00884838105805(21). Serial Numbers: 17271 17389 17439 17449 17504 24039 24077 34055 34072 34129 34262 38117 38166 38222

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-3204-2024