ECT Internal Fracture Fixation Quick-Connect Drill 4.0 mm Diameter 20cm Length, Item Number/EDI 00241004000, Nonsterile. orthopedic surgical instrument.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Number: 63518082
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Zimmer Biomet, Inc.
Reason for Recall:
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

ECT Internal Fracture Fixation Quick-Connect Drill 4.0 mm Diameter 20cm Length, Item Number/EDI 00241004000, Nonsterile. orthopedic surgical instrument.

Product Codes/Lot Numbers:

Lot Number: 63518082

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-3188-2017

Related Recalls

Zimmer Biomet is conducting a medical device correction to update the Instructions for Use (IFU) and the Surgical Techniques for the Polaris Spinal System Translation Screw and Cypher MIS Screw System. Users should be aware that changes were made to the contraindications section indicating that the Translation Screws should not be used with a direct current stimulation, as a high probability of construct failure exists when using Translation Screws containing CoCr with direct current simulation system.

Nov 17, 2020 Other Medical Devices Nationwide View Details →