CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Product No. 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65; UDI: 10607567109053, 10607567111117, 10607567109008, 10607567111940, 10607567109107, 10607567114187, 10607567108421, 10607567108438, 10607567108391, 10607567108414, 10607567113432; Lot No. ALL LOTS.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Datascope Corp.
- Reason for Recall:
- Firm has developed a software correction (Revision D.01) for outside the US to mitigate product problems that pose a risk of hemodynamic instability and failure of transmission of clinical data in Cardiosave Hybrid and Cardiosave Rescue Intra-Aortic Balloon Pump devices.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
CARDIOSAVE HYBRID INTRA-AORTIC BALLOON PUMP (IABP). Electromechanical system used to inflate and deflate intra-aortic balloons.
Product Codes/Lot Numbers:
Product No. 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-00-0800-35, 0998-00-0800-36, 0998-00-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65; UDI: 10607567109053, 10607567111117, 10607567109008, 10607567111940, 10607567109107, 10607567114187, 10607567108421, 10607567108438, 10607567108391, 10607567108414, 10607567113432; Lot No. ALL LOTS.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-3136-2024
Related Recalls
The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
The IFU addendum revises the Preventative Maintenance schedule to align with the update introduced in the Service Manual as of June 2023.
There is discrepant labeling on the inner and outer packaging of the device.