Coulter LH 750, Coulter LH 780, and Coulter LH 500 Analyzers Hematology Analyzers are quantitative, automated hematology analyzers and leukocyte differential counters For In Vitro Diagnostic used in clinical laboratories. They also provide automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The purpose of the LH 700 Series is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies of any of these parameters. These studies might include further measurements of cell size and platelet distribution, manual WBC differential or any other definitive test that helps diagnose the patients condition. The purpose of the LH 500 Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, or any other test that aids in diagnosing an abnormality

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Catalog Number(s): LH 750 System - 6605632, A85570, A68807; LH 780 System - 723585, A90728, A68808; LH 500 system - 6605525, 6605526, 6605527, A85566, A85564; All software versions
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Beckman Coulter Inc.
Reason for Recall:
Beckman Coulter has determined that additional clarification for the Blast Suspect messages is necessary. In rare situations, the LH 750, LH 780, LH 500 and HmX Analyzers may not flag or detect blasts in some blood samples.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Coulter LH 750, Coulter LH 780, and Coulter LH 500 Analyzers Hematology Analyzers are quantitative, automated hematology analyzers and leukocyte differential counters For In Vitro Diagnostic used in clinical laboratories. They also provide automated Reticulocyte analysis and enumeration of nucleated red blood cells (NRBCs) as well as an automated method for enumeration of RBCs and WBCs in body fluids. The purpose of the LH 700 Series is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies of any of these parameters. These studies might include further measurements of cell size and platelet distribution, manual WBC differential or any other definitive test that helps diagnose the patients condition. The purpose of the LH 500 Analyzer is to separate the normal patient, with all normal system-generated parameters, from the patient who needs additional studies. These studies include further measurements of cell size and cell distribution, or any other test that aids in diagnosing an abnormality

Product Codes/Lot Numbers:

Catalog Number(s): LH 750 System - 6605632, A85570, A68807; LH 780 System - 723585, A90728, A68808; LH 500 system - 6605525, 6605526, 6605527, A85566, A85564; All software versions

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-3113-2017

Related Recalls

It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.

Nov 7, 2025 Infusion Pumps Nationwide View Details →

Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.

Sep 22, 2025 Diagnostic Equipment Nationwide View Details →