DBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S, The lead kit contains one DBS lead in addition to multiple accessories used in DBS lead implant procedures. Among these accessories is the DBS lead holder, also referred to as the lead depth stop. The lead holder affixes to the DBS lead during the lead implant procedure, marking the distance to the lead tip and providing control of DBS lead depth during implant by interfacing with the stereotactic system

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 3387, Lot Serial #: 0206880294, 0206882647, 0206882779, 0206882780, 0206882781, 0206882782, 0206882783, 0206886184, 0206886323, 0206886328, 0206886330, 0206886331, 0206886332, 0206886333, 0206886603, 0206887813, 0206890443, 0206890444, 0206903488, 0206903489, 0206906792, 0206906793, 0206907159, 0206916391, 0206916392, 0206916393, 0206916394, 0206916395, 0206916396, 0206916397, 0206916398, 0206916399, 0206916797, 0206917120, 0206917121, 0206917122, 0206917123, 0206917124, 0206917125, 0206917126, 0206917127, 0206917128, 0206917129, 0206917130, 0206917131, 0206917132, 0206917302, 0206917303, 0206917304, 0206917306, 0206917307, 0206917308, 0206917310, 0206917312, 0206917560, 0206917561, 0206917563, 0206917564, 0206917565, 0206917566, 0206917568, 0206917569, 0206917570, 0206917572, 0206917573, 0206917574, 0206917575, 0206917576, 0206917577, 0206917578, 0206917579, 0206917620, 0206917621, 0206917622, 0206917623, 0206917624, 0206917625, 0206917626, 0206917627, 0206917628, 0206917629, 0206917630, 0206917631, 0206917632, 0206917633, 0206917634, 0206917635, 0206917636, 0206917637, 0206917638, 0206917639, 0206917660, 0206917661, 0206917662, 0206917663, 0206917664, 0206917665, 0206917666, 0206917667, 0206917668, 0206917669, 0206917670, 0206917671, 0206920443, 0206920444, 0206920445, 0206920447, 0206920448, 0206920449, 0206920450, 0206920451, 0206920452, 0206920453, 0206920454, 0206920455, 0206920456, 0206920457, 0206920458, 0206920459, 0206920460, 0206920461, 0206920462, 0206920463, 0206920663, 0206920664, 0206920665, 0206920666, 0206920667, 0206920668, 0206920669, 0206920670, 0206920671, 0206920672, 0206920673, 0206924158, 0206924492, 0206924493, 0206924494, 0206924604, 0206924605, 0206924606, 0206924607, 0206924608, 0206924609, 0206924610, 0206924611, 0206924612, 0206924613, 0206924614, 0206924615, 0206924616, 0206924617, 0206924618, 0206924619, 0206925140, 0206925141, 0206925142, 0206925143, 0206925245, 0206925290, 0206925291, 0206925292, 0206926111, 0206926112, 0206926414, 0206929827, 0206929828, 0206929829, 0206929830, 0206929849, 0206930146, 0206930736, 0206930737, 0206930941, 0206930942, 0206930943, 0206931910, 0206931911, 0206931927, 0206931928, 0206933408, 0206934345, 0206934346, 0206934347, 0206937919, 0206937920, 0206942506, 0206946796, 0206946798, 0206949034, 0206949035, 0206950164, 0206955498, 0206955499, 0206956649, 0206956650, 0206957515, 0206957516, 0206957517, 0206957518, 0206957519, 0206957801, 0206957802, 0206957803, 0206957804, 0206957805, 0206957806, 0206957807, 0206959205, 0206959206, 0206959207, 0206959208, 0206959209, 0206959210, 0206959211, 0206959212, 0206959213, 0206959214, 0206959215, 0206959216, 0206959217, 0206959218, 0206959220, 0206959221, 0206959222, 0206959223, 0206959224, 0206959501, 0206959502, 0206959503, 0206959682, 0206959684, 0206959686, 0206959687, 0206960104, 0206960105, 0206960106, 0206960107, 0206960108, 0206960109, 0206960147, 0206960148, 0206960149, 0206960655, 0206960658, 0206960666, 0206960667, 0206960668, 0206960669, 0206960670, 0206960671, 0206960676, 0206960677, 0206960678, 0206960679, 0206960680, 0206960681, 0206960694, 0206960695, 0206960696, 0206969559, 0206969560, 0206969561, 0206969562, 0206969563, 0206969564, 0206969565, 0206969566, 0206969567, 0206969568, 0206969569, 0206969570, 0206969571, 0206969572, 0206969573, 0206969574, 0206969575, 0206969576, 0206969577, 0206969578, 0206969691, 0206969692, 0206970927, 0206973200, 0206973201, 0206973202, 0206973203, 0206973204, 0206973205, 0206973206, 0206973800, 0206973801, 0206978789, 0206980541, 0206980542, 0206980543, 0206980970, 0206980971, 0206981705, 0206981706, 0206981707, 0206981714, 0206981715, 0206981754, 0206981771, 0206981772, 0206984645, 0206984646, 0206984650, 0206984675, 0206991945, 0206991946, 0206991947, 0206991948, 0206991949, 0206991950, 0206991951, 0206991952, 0206991953, 0206991954, 0206991955, 0206991956, 0206991957, 0206991958, 0206991959, 0206991960, 0206991961, 0206991962, 0206991963, 0206991964, 0206991965, 0206991966, 0206991967, 0206991968, 0206991969, 0206991970, 0206991971, 0206991972, 0206991973, 0206991974, 0206991975, 0206991976, 0206991977, 0206991978, 0206991979, 0206991984, 0206991985, 0206991986, 0206991987, 0206991988, 0206991989, 0206991990, 0206991991, 0206991992, 0206991993, 0206991994, 0206991995, 0206991996, 0206991997, 0206991998, 0206991999, 0206992000, 0206992001, 0206992002, 0206992003, 0206992004, 0206992005, 0206992006, 0206992007, 0206992008, 0206992009, 0206992010, 0206992011, 0206992012, 0206992013, 0206992014, 0206992015, 0206992016, 0206992017, 0206992018, 0206992019, 0206992020, 0206992021, 0206992022, 0206992023, 0206992024, 0206992025, 0206992026, 0206992027, 0206992028, 0206992029, 0206992030, 0206992031, 0206992032, 0206992033, 0206992034, 0206992035, 0206992036, 0206992037, 0206992038, 0206992039, 0206992040, 0206996540, 0206996541, 0206996542, 0206996543, 0206996544, 0206996545, 0206996546, 0206996547, 0206996548, 0206996549, 0206996550, 0206996551, 0206996552, 0206996553, 0206996554, 0206996555, 0206996556, 0206996557, 0206996558, 0206996559, 0206996560, 0206996561, 0206996562, 0206996563, 0207000225, 0207006520, 0207006777, 0207006778, 0207006779, 0207006780, 0207007919, 0207008220, 0207008221, 0207008222, 0207011212, 0207011213, 0207015958, 0207016168, 0207016796, 0207018882, 0207018883, 0207018884, 0207021792, 0207021793, 0207021794, 0207021795, 0207021796, 0207021797, 0207021798, 0207021799, 0207021800, 0207021801, 0207021802, 0207021803, 0207022120, 0207022121, 0207022122, 0207027753, 0207030645, 0207030646, 0207030647, 0207030648, 0207030649, 0207030650, 0207030651, 0207030652, 0207030653, 0207030654, 0207030655, 0207030656, 0207030657, 0207030658, 0207030659, 0207030780, 0207030781, 0207030980, 0207030981, 0207030982, 0207030983, 0207030985, 0207038190, 0207038192, 0207038195, 0207038423, 0207038424, 0207038425, 0207038426, 0207038427, 0207038434, 0207038435, 0207038436, 0207038437, 0207038438, 0207038439, 0207038440, 0207038441, 0207038442, 0207038443, 0207038457, 0207038458, 0207038459, 0207038460, 0207038461, 0207038462, 0207040184, 0207040185, 0207040187, 0207040188, 0207040189, 0207040190, 0207040191, 0207040239, 0207040240, 0207040241, 0207040242, 0207040243, 0207040244, 0207040573, 0207040574, 0207040575, 0207040576, 0207040712, 0207040713, 0207040714, 0207040715, 0207040718, 0207040719, 0207041016, 0207041811, 0207043628, 0207043629, 0207043630, 0207043631, 0207043632, 0207043658, 0207043659, 0207043660, 0207043871, 0207043872, 0207043874, 0207043875, 0207047746, 0207052711, 0207054947, 0207056625, 0207056626, 0207056628, 0207056629, 0207056630, 0207056631, 0207056807, 0207057256, 0207057257, 0207057427, 0207057428, 0207057429, 0207057430, 0207057431, 0207057432, 0207057433, 0207057434, 0207057435, 0207057436, 0207057437, 0207057438, 0207057439, 0207057440, 0207057441, 0207057442, 0207059495, 0207059565, 0207059566, 0207059567, 0207059956, 0207059959, 0207059963, 0207059985, 0207059986, 0207059987, 0207062240, 0207063038, 0207063039, 0207063040, 0207063133, 0207063134, 0207063135, 0207063136, 0207063137, 0207063138, 0207063139, 0207063140, 0207063141, 0207063142, 0207067134, 0207067135, 0207067136, 0207067137, 0207067138, 0207067139, 0207067140, 0207067141, 0207067142, 0207067143, 0207067146, 0207067147, 0207067148, 0207067847, 0207067848, 0207068325, 0207068326, 0207068327, 0207068449, 0207070300, 0207070321, 0207071704, 0207071705, 0207072960, 0207072961, 0207072962, 0207072963, 0207072964, 0207072965, 0207072966, 0207072967, 0207072968, 0207072969, 0207072970, 0207072971, 0207072972, 0207072973, 0207072974, 0207079003, 0207079004, 0207079339, 0207079340, 0207079341, 0207079342, 0207079343, 0207079414, 0207079415, 0207079416, 0207079417, 0207079418, 0207079419, 0207079435, 0207079436, 0207079437, 0207079438, 0207079439, 0207079440, 0207079441, 0207079467, 0207079468, 0207079469, 0207079470, 0207079471, 0207079472, 0207079473, 0207079474, 0207079475, 0207079476, 0207079477, 0207079478, 0207079479, 0207079480, 0207079481, 0207079482, 0207079483, 0207079484, 0207079485, 0207079486, 0207079487, 0207079488, 0207079489, 0207079490, 0207079491, 0207079492, 0207079493, 0207079494, 0207079495, 0207079496, 0207079497, 0207079498, 0207079641, 0207079642, 0207079643, 0207079645, 0207079680, 0207079681, 0207079701, 0207083910, 0207083911, 0207083912, 0207084903, 0207084904, 0207087901, 0207087902, 0207087903, 0207087904, 0207087905, 0207087906, 0207087907, 0207087908, 0207087909, 0207087910, 0207087951, 0207087952, 0207088003, 0207088004, 0207088005, 0207088431, 0207094140, 0207094142, 0207097722, 0207097723, 0207098161, 0207098162, 0207098163, 0207098164, 0207100512, 0207100514, 0207100515, 0207100516, 0207100517, 0207100810, 0207100848, 0207101008, 0207101009, 0207101010, 0207101023, 0207101031, 0207101053, 0207101120, 0207105162, 0207105163, 0207105164, 0207105165, 0207105166, 0207105167, 0207105168, 0207105169, 0207105170, 0207105171, 0207105172, 0207105176, 0207105321, 0207105322, 0207105323, 0207105324, 0207105325, 0207105326, 0207105327, 0207105328, 0207105329, 0207105330, 0207105331, 0207105332, 0207105333, 0207105334, 0207105335, 0207105336, 0207105352, 0207105356, 0207108624, 0207108625, 0207108635, 0207108636, 0207109103, 0207109704, 0207109705, 0207109706, 0207109708, 0207109709, 0207109710, 0207109711, 0207109712, 0207109713, 0207109714, 0207109715, 0207109716, 0207109717, 0207109718, 0207109719, 0207109720, 0207109725, 0207109726, 0207109728, 0207109729, 0207109733, 0207109734, 0207109737, 0207109815, 0207110541, 0207110542, 0207110543, 0207112954, 0207113814, 0207113815, 0207113816, 0207113817, 0207113818, 0207114006, 0207114007, 0207114746, 0207114852, 0207114853, 0207114854, 0207114855, 0207117588, 0207117589, 0207117590, 0207117741, 0207118073, 0207119315, 0207119316, 0207119317, 0207119318, 0207119319, 0207120110, 0207121868, 0207122061, 0207122062, 0207122063, 0207122064, 0207122065, 0207122066, 0207122067, 0207122068, 0207122069, 0207122070, 0207122071, 0207122072, 0207122099, 0207122100, 0207122101, 0207122127, 0207122128, 0207122129, 0207122183, 0207125639, 0207125641, 0207125642, 0207125643, 0207125754, 0207125851, 0207125876, 0207129909, 0207129910, 0207129912, 0207129913, 0207129917, 0207129920, 0207129921, 0207129922, 0207129923, 0207130018, 0207130019, 0207130020, 0207130021, 0207130022, 0207130023, 0207130024, 0207130025, 0207130026, 0207130027, 0207130028, 0207130029, 0207130081, 0207130299, 0207130300, 0207130301, 0207130302, 0207130303, 0207130304, 0207130305, 0207130306, 0207130307, 0207130308, 0207130724, 0207130786, 0207130787, 0207130788, 0207130789, 0207130790, 0207130845, 0207130846, 0207130847, 0207130848, 0207130849, 0207130850, 0207130851, 0207131732, 0207131734, 0207131757, 0207131758, 0207131759, 0207131780, 0207132032, 0207132033, 0207132098, 0207132677, 0207135004, 0207136742, 0207136743, 0207139918, 0207139919, 0207139920, 0207139921, 0207139922, 0207139923, 0207139924, 0207139925, 0207139926, 0207139927, 0207140714, 0207140715, 0207140716, 0207140717, 0207140718, 0207140719, 0207140720, 0207140724, 0207140725, 0207141032, 0207141033, 0207141034, 0207141035, 0207141036, 0207143663, 0207143678, 0207143679, 0207143680, 0207143681, 0207143682, 0207143683, 0207143697, 0207143698, 0207143700, 0207143701, 0207143707, 0207143708, 0207143712, 0207143713, 0207143714, 0207143715, 0207143716, 0207143717, 0207143718, 0207143719, 0207144809, 0207144817, 0207144818, 0207144819, 0207144840, 0207144841, 0207144842, 0207144843, 0207144844, 0207144846, 0207144847, 0207144848, 0207144883, 0207144884, 0207144885, 0207144886, 0207144887, 0207144888, 0207144889, 0207144890, 0207145866, 0207145867, 0207148304, 0207148305, 0207148306, 0207159258, 0207159259, 0207159260, 0207160611, 0207160612, 0207160614, 0207160615, 0207160678, 0207160679, 0207160680, 0207160708, 0207160709, 0207160710, 0207161115, 0207163879, 0207163881, 0207163882, 0207163883, 0207163884, 0207163887, 0207163889, 0207164673, 0207164674, 0207164675, 0207164676, 0207164677, 0207164678, 0207164870, 0207164871, 0207164872, 0207168561, 0207168562, 0207168563, 0207168564, 0207168565, 0207168566, 0207168567, 0207168568, 0207168569, 0207168570, 0207168571, 0207168572, 0207168573, 0207168574, 0207168575, 0207168576, 0207168577, 0207168595, 0207168596, 0207168597, 0207168598, 0207168599, 0207168600, 0207168929, 0207168930, 0207168931, 0207168932, 0207168954, 0207168955, 0207168956, 0207168957, 0207168959, 0207168961, 0207168962, 0207168963, 0207168965, 0207168966, 0207168967, 0207168968, 0207168969, 0207168970, 0207168972, 0207168979, 0207168980, 0207168985, 0207168986, 0207168987, 0207169010, 0207174625, 0207174627, 0207174628, 0207174629, 0207174630, 0207174654, 0207175291, 0207175292, 0207175293, 0207175294, 0207175295, 0207175296, 0207176745, 0207177151, 0207177152, 0207177153, 0207177154, 0207180633, 0207180634, 0207180635, 0207180636, 0207180637, 0207180638, 0207180639, 0207180640, 0207180641, 0207180642, 0207180643, 0207180644, 0207180645, 0207180646, 0207180647, 0207180648, 0207180649, 0207180650, 0207180651, 0207181930, 0207181931, 0207181932, 0207181933, 0207181934, 0207181987, 0207181988, 0207181989, 0207181990, 0207181991, 0207181992, 0207181993, 0207181994, 0207181995, 0207181996, 0207181997, 0207182077, 0207182782, 0207182783, 0207182785, 0207182786, 0207182787, 0207182788, 0207182790, 0207182791, 0207182793, 0207182794, 0207182795, 0207182796, 0207182797, 0207182798, 0207182799, 0207182840, 0207182841, 0207182842, 0207182843, 0207182844, 0207182846, 0207185517, 0207185518, 0207185519, 0207185520, 0207185521, 0207185522, 0207185523, 0207185524, 0207185525, 0207185636, 0207185651, 0207192753, 0207192754, 0207192790, 0207192791, 0207192815, 0207192816, 0207192817, 0207192818, 0207192819, 0207192820, 0207192821, 0207192822, 0207192823, 0207192824, 0207192825, 0207192826, 0207192827, 0207192828, 0207192829, 0207192910, 0207192911, 0207202710, 0207202711, 0207202712, 0207202717, 0207202718, 0207202719, 0207202740, 0207202741, 0207202742, 0207202743, 0207202744, 0207202745, 0207202746, 0207202747, 0207202748, 0207202750, 0207202754, 0207202755, 0207202756, 0207202757, 0207202758, 0207202764, 0207202765, 0207202766, 0207202801, 0207202805, 0207202932, 0207202936, 0207202943, 0207202944, 0207202961, 0207203570, 0207203571, 0207208667, 0207208668, 0207208714, 0207208715, 0207208716, 0207208723, 0207220085, 0207220740, 0207220741, 0207220742, 0207220743, 0207220745, 0207220746, 0207225509, 0207225523, 0207225564, 0207225565, 0207225566, 0207225567, 0207225568, 0207225569, 0207225570, 0207225571, 0207226531, 0207226532, 0207226533, 0207226534, 0207226535, 0207226626, 0207226627, 0207226628, 0207226629, 0207226630, 0207226631, 0207226632, 0207226633, 0207226634, 0207226635, 0207226636, 0207226637, 0207230154, 0207232091, 0207232126, 0207232266, 0207234900, 0207234901, 0207234902, 0207240828, 0207240832, 0207240833, 0207240834, 0207240841, 0207240842, 0207240843, 0207240844, 0207240845, 0207240856, 0207240857, 0207240858, 0207241071, 0207241072, 0207241073, 0207241444, 0207241445, 0207241455, 0207241456, 0207241457, 0207241458, 0207241459, 0207241460, 0207241461, 0207241463, 0207246705, 0207247848, 0207247849, 0207247850, 0207250567, 0207250707, 0207250708, 0207250710, 0207250711, 0207250712, 0207250713, 0207250714, 0207250715, 0207250716, 0207250717, 0207250718, 0207250719, 0207250720, 0207250721, 0207250722, 0207250723, 0207250724, 0207250725, 0207250726, 0207250727, 0207250728, 0207250729, 0207250730, 0207250731, 0207250809, 0207250810, 0207250812, 0207250813, 0207250814, 0207250815, 0207250816, 0207250817, 0207250818, 0207250829, 0207254564, 0207254565, 0207254742, 0207255595, 0207257987, 0207257988, 0207257989, 0207257990, 0207257991, 0207258020, 0207258045, 0207258259, 0207258280, 0207258638, 0207258639, 0207258640, 0207258641, 0207259128, 0207259174, 0207259175, 0207260433, 0207260434, 0207261780, 0207261788, 0207261789, 0207261790, 0207266491, 0207266492, 0207266530, 0207266531, 0207266532, 0207268177, 0207268178, 0207268179, 0207268231, 0207268232, 0207270938, 0207270939, 0207270972, 0207270973, 0207275052, 0207275053, 0207275054, 0207275058, 0207275218, 0207275219, 0207278567, 0207278568, 0207278569, 0207278570, 0207278572, 0207278573, 0207278574, 0207278575, 0207278576, 0207278577, 0207278578, 0207278579, 0207278639, 0207279750, 0207280151, 0207280180, 0207280181, 0207280200, 0207280201, 0207280202, 0207280203, 0207280204, 0207280205, 0207280206, 0207280207, 0207280208, 0207280209, 0207280210, 0207280211, 0207280217, 0207280218, 0207280219, 0207280220, 0207280221, 0207280222, 0207280223, 0207280224, 0207280225, 0207280226, 0207280227, 0207280228, 0207280229, 0207280230, 0207280231, 0207280232, 0207280233, 0207280238, 0207280239, 0207283309, 0207283310, 0207283614, 0207283627, 0207283628, 0207283629, 0207283630, 0207283631, 0207283632, 0207283638, 0207283639, 0207283841, 0207283883, 0207283884, 0207283885, 0207283886, 0207283903, 0207283944, 0207283945, 0207283946, 0207283947, 0207283948, 0207283949, 0207283950, 0207283951, 0207283952, 0207283953, 0207283954, 0207283955, 0207283956, 0207283957, 0207288214, 0207288219, 0207289211, 0207289212, 0207289214, 0207289215, 0207289216, 0207291547, 0207291548, 0207291549, 0207291686, 0207291687, 0207291688, 0207294519, 0207298853, 0207298854, 0207298855, 0207298856, 0207298857, 0207298858, 0207298859, 0207298996, 0207299404, 0207299920, 0207299921, 0207299922, 0207299923, 0207299924, 0207299925, 0207299926, 0207299927, 0207299928, 0207300520, 0207300521, 0207300522, 0207300523, 0207300699, 0207301220, 0207301221, 0207303119, 0207303499, 0207303641, 0207303656, 0207303885, 0207303886, 0207304245, 0207304800, 0207304801, 0207305118, 0207305119, 0207306460, 0207308323, 0207308324, 0207308325, 0207308326, 0207308327, 0207308328, 0207309474, 0207309475, 0207309476, 0207309477, 0207309478, 0207309479, 0207309480, 0207309481, 0207309482, 0207309483, 0207309484, 0207309485, 0207309486, 0207309487, 0207309488, 0207309489, 0207309490, 0207309491, 0207309492, 0207309493, 0207309494, 0207309495, 0207309496, 0207309497, 0207309498, 0207309499, 0207309500, 0207309501, 0207309502, 0207309505, 0207309506, 0207309507, 0207309508, 0207309516, 0207309517, 0207309518, 0207309519, 0207309520, 0207309521, 0207309522, 0207309523, 0207309524, 0207309525, 0207309526, 0207309527, 0207309530, 0207309531, 0207309532, 0207309533, 0207309534, 0207309535, 0207309536, 0207319594, 0207324654, 0207324655, 0207324656, 0207324657, 0207324658, 0207324659, 0207324680, 0207324681, 0207324682, 0207324683, 0207324684, 0207324685, 0207324686, 0207324687, 0207324688, 0207324689, 0207324690, 0207324691, 0207324692, 0207324693, 0207324694, 0207324695, 0207324696, 0207324697, 0207324698, 0207324699, 0207324700, 0207326089, 0207326090, 0207326091, 0207326131, 0207326132, 0207327178, 0207337977, 0207342360, 0207348166, 0207348167, 0207348168, 0207348211, 0207348212, 0207348214, 0207348215, 0207348216, 0207348217, 0207348218, 0207348583, 0207350341, 0207353632, 0207353633, 0207353634, 0207353751, 0207353898, 0207359405, 0207359406, 0207359407, 0207359408, 0207359409, 0207359410, 0207359411, 0207359412, 0207359413, 0207359414, 0207359415, 0207359416, 0207359417, 0207359418, 0207359419, 0207360423, 0207360424, 0207360425, 0207360426, 0207360427, 0207360428, 0207360429, 0207360430, 0207360431, 0207360432, 0207360433, 0207360434, 0207360435, 0207360436, 0207360437, 0207360438, 0207360439, 0207361120, 0207361121, 0207361122, 0207361123, 0207361280, 0207361281, 0207361282, 0207361283, 0207361284, 0207361285, 0207361286, 0207361287, 0207361348, 0207368928, 0207368929, 0207368930, 0207368931, 0207368932, 0207368933, 0207368934, 0207368935, 0207372779, 0207372793, 0207372892, 0207372893, 0207372895, 0207372896, 0207379102, 0207379355, 0207379356, 0207379405, 0207379406, 0207380120, 0207380121, 0207380122, 0207380123, 0207385226, 0207396642, 0207396707, 0207396708, 0207397023, 0207397864, 0207397880, 0207397887, 0207397888, 0207397890, 0207397892, 0207397896, 0207397897, 0207397898, 0207397899, 0207397900, 0207397901, 0207397902, 0207397904, 0207397905, 0207397906, 0207397907, 0207397956, 0207401852, 0207401853, 0207404964, 0207407581, 0207409006, 0207409007, 0207409008, 0207409009, 0207413918, 0207413919, 0207414060, 0207414061, 0207414062, 0207414063, 0207414064, 0207414065, 0207414066, 0207414067, 0207419438, 0207422718, 0207424300, 0207424302, 0207424303, 0207424304, 0207424305, 0207424756, 0207428968, 0207428969, 0207428970, 0207428971, 0207428972, 0207428973, 0207430334, 0207430399, 0207430403, 0207430404, 0207430405, 0207432103, 0207432104, 0207432105, 0207432666, 0207437724, 0207441876, 0207444058, 0207444059, 0207444300, 0207446984, 0207448988, 0207448989, 0207448990, 0207448991, 0207448992, 0207448996, 0207449981, 0207449982, 0207449993, 0207449994, 0207449995, 0207449996, 0207449997, 0207449998, 0207454223, 0207454224, 0207454225, 0207454226, 0207454227, 0207454228, 0207454229, 0207454401, 0207454402, 0207454403, 0207454404, 0207454405, 0207454406, 0207454407, 0207454408, 0207454409, 0207456279, 0207457340, 0207457341, 0207457342, 0207457343, 0207457344, 0207457345, 0207457346, 0207457347, 0207457348, 0207457349, 0207457350, 0207457351, 0207457352, 0207457353, 0207457354, 0207457355, 0207457356, 0207457357, 0207457358, 0207457359, 0207457360, 0207457361, 0207457362, 0207457363, 0207457364, 0207457365, 0207457366, 0207457367, 0207457368, 0207457369, 0207457370, 0207457371, 0207457372, 0207457373, 0207457374, 0207457375, 0207457376, 0207457377, 0207457378, 0207457379, 0207457380, 0207457381, 0207457382, 0207457383, 0207457384, 0207457385, 0207457386, 0207457387, 0207457388, 0207457389, 0207457390, 0207457391, 0207457392, 0207457393, 0207457394, 0207457395, 0207457396, 0207457397, 0207457398, 0207457399, 0207457400, 0207457401, 0207457402, 0207457403, 0207457404, 0207457405, 0207457406, 0207457407, 0207457408, 0207457409, 0207457410, 0207457411, 0207457412, 0207457413, 0207457414, 0207457415, 0207457416, 0207457417, 0207457418, 0207457419, 0207457420, 0207457421, 0207457422, 0207460357, 0207460358, 0207460359, 0207460360, 0207460361, 0207460871, 0207460872, 0207466531, 0207466532, 0207466591, 0207466592, 0207466593, 0207466594, 0207466595, 0207466596, 0207466597, 0207466598, 0207466599, 0207466600, 0207466601, 0207466602, 0207466603, 0207466604, 0207466605, 0207466606, 0207466607, 0207466608, 0207466674, 0207469251, 0207469252, 0207471095, 0207471096, 0207474219, 0207498243, 0207498244, 0207498245, 0207498246, 0207498247, 0207498250, 0207498251, 0207507151, 0207507152, 0207507153, 0207507154, 0207507447, 0207507448, 0207507449, 0207520659, 0207523074, 0207526687, 0207531718, 0207531719, 0207531720, 0207534092, 0207534093, 0207534573, 0207534575, 0207536498, 0207536499, 0207536500, 0207536501, 0207536503, 0207536504, 0207536505, 0207536506, 0207538329, 0207538744, 0207538745, 0207541018, 0207546649, 0207546650, 0207546651, 0207546652, 0207546653, 0207546654, 0207546655, 0207546656, 0207546657, 0207546658, 0207547088, 0207547089, 0207547090, 0207547091, 0207547092, 0207547093, 0207547094, 0207547095, 0207547096, 0207547097, 0207547098, 0207547099, 0207547100, 0207547101, 0207547102, 0207547111, 0207547112, 0207547113, 0207547114, 0207547115, 0207547116, 0207547117, 0207547118, 0207547119, 0207547120, 0207547121, 0207547122, 0207547123, 0207547311, 0207548014, 0207548015, 0207548639, 0207548883, 0207548884, 0207548885, 0207548886, 0207549280, 0207550658, 0207554148, 0207555823, 0207561059, 0207561060, 0207567863, 0207574365, 0207585359, 0207588282, 0207588640, 0207588643, 0207588644, 0207588645, 0207588672, 0207588674, 0207588808, 0207588809, 0207588810, 0207588811, 0207588860, 0207588861, 0207588865, 0207588866, 0207592741, 0207595000, 0207595001, 0207595002, 0207595003, 0207595017, 0207598167, 0207598168, 0207598169, 0207600533, 0207600534, 0207600535, 0207601075, 0207602702, 0207604544, 0207605819, 0207606004, 0207609972, 0207609973, 0207609974, 0207609975, 0207609976, 0207609977, 0207609978, 0207611974, 0207611975, 0207612377, 0207615806, 0207615807, 0207615808, 0207615809, 0207615810, 0207615811, 0207615812, 0207615813, 0207615814, 0207615815, 0207615816, 0207615817, 0207615818, 0207615819, 0207615835, 0207615836, 0207620203, 0207620204, 0207620205, 0207620206, 0207620207, 0207620364, 0207620790, 0207620791, 0207620793, 0207620794, 0207620795, 0207620796, 0207621611, 0207621612, 0207621613, 0207621614, 0207621615, 0207621616, 0207621617, 0207621618, 0207621619, 0207621620, 0207621621, 0207621622, 0207621623, 0207621624, 0207621625, 0207621626, 0207621627, 0207621628, 0207621629, 0207621630, 0207621631, 0207621632, 0207621633, 0207621634, 0207621635, 0207621636, 0207621637, 0207621638, 0207621639, 0207621640, 0207621641, 0207621642, 0207621643, 0207621644, 0207621645, 0207621646, 0207621647, 0207621648, 0207621649, 0207621650, 0207621651, 0207621652, 0207621653, 0207621654, 0207621830, 0207621831, 0207622044, 0207622045, 0207622169, 0207622176, 0207622177, 0207622178, 0207622179, 0207623055, 0207625459, 0207627667, 0207627668, 0207627669, 0207627690, 0207627691, 0207627692, 0207627693, 0207627694, 0207627695, 0207627696, 0207627949, 0207628172, 0207631094, 0207632116, 0207634225, 0207634233, 0207637637, 0207642804, 0207642807, 0207642813, 0207647090, 0207647249, 0207651194, 0207651195, 0207651196, 0207651197, 0207651352, 0207652932, 0207652933, 0207653543, 0207660129, 0207660130, 0207660131, 0207660132, 0207660133, 0207660134, 0207660135, 0207660136, 0207660171, 0207660878, 0207664645, 0207664646, 0207664647, 0207670273, 0207670274, 0207670278, 0207670279, 0207674827, 0207674828, 0207674829, 0207681428, 0207682857, 0207682886, 0207682888, 0207682890, 0207682892, 0207682903, 0207682904, 0207682906, 0207682908, 0207682909, 0207682913, 0207682917, 0207682918, 0207682940, 0207682942, 0207682944, 0207682946, 0207682948, 0207682950, 0207682951, 0207682953, 0207682955, 0207682957, 0207682959, 0207682960, 0207682962, 0207682964, 0207683075, 0207683076, 0207683077, 0207683078, 0207683079, 0207683100, 0207683101, 0207683103, 0207683104, 0207683105, 0207683106, 0207683108, 0207683109, 0207683110, 0207683111, 0207683112, 0207683113, 0207683114, 0207683115, 0207683116, 0207683117, 0207684419, 0207684662, 0207684663, 0207685023, 0207685138, 0207685139, 0207685454, 0207687157, 0207687571, 0207687572, 0207692231, 0207692232, 0207694357, 0207699502, 0207699503, 0207699504, 0207699507, 0207700963, 0207704751, 0207704753, 0207704959, 0207704960, 0207705029, 0207705030, 0207705031, 0207705032, 0207705033, 0207705034, 0207705035, 0207705036, 0207705041, 0207705042, 0207705075, 0207708402, 0207710979, 0207711007, 0207720013, 0207720602, 0207721774, 0207722086, 0207729498, 0207729499, 0207730540, 0207730541, 0207730542, 0207730543, 0207730544, 0207730545, 0207730546, 0207730547, 0207730548, 0207730549, 0207730550, 0207731594, 0207731595, 0207733668, 0207733678, 0207733695, 0207733696, 0207733697, 0207733698, 0207733699, 0207733819, 0207734053, 0207734799, 0207735690, 0207737625, 0207737626, 0207737627, 0207737628, 0207737629, 0207737630, 0207737838, 0207737839, 0207742414, 0207742415, 0207742416, 0207742417, 0207742418, 0207742419, 0207742420, 0207742421, 0207742422, 0207742423, 0207742452, 0207742453, 0207742454, 0207742455, 0207742456, 0207742457, 0207742458, 0207742459, 0207743128, 0207747304, 0207747305, 0207747306, 0207747307, 0207747308, 0207747309, 0207747417, 0207747418, 0207747419, 0207750080, 0207752765, 0207752766, 0207752767, 0207752768, 0207752769, 0207752770, 0207752771, 0207752772, 0207756643, 0207756644, 0207756645, 0207756646, 0207756648, 0207756716, 0207756717, 0207756718, 0207756719, 0207756720, 0207756721, 0207756722, 0207756757, 0207757244, 0207757245, 0207757246, 0207757247, 0207757352, 0207757353, 0207757354, 0207757355, 0207757356, 0207757357, 0207757358, 0207757359, 0207757361, 0207757362, 0207757451, 0207757452, 0207757453, 0207761819, 0207761900, 0207762185, 0207762186, 0207762187, 0207762188, 0207762189, 0207762190, 0207762191, 0207762192, 0207762193, 0207762194, 0207762195, 0207762196, 0207762197, 0207762198, 0207762199, 0207762200, 0207762201, 0207762202, 0207762203, 0207762204, 0207762205, 0207762206, 0207762207, 0207762208, 0207762209, 0207762210, 0207762211, 0207762212, 0207762213, 0207762214, 0207762215, 0207762216, 0207762217, 0207762218, 0207762681, 0207762682, 0207762683, 0207762684, 0207762722, 0207762723, 0207764452, 0207767622, 0207767623, 0207767624, 0207767625, 0207767626, 0207767627, 0207767628, 0207767629, 0207767928, 0207767929, 0207768006, 0207768007, 0207768008, 0207768009, 0207768010, 0207768011, 0207768012, 0207768013, 0207769247, 0207769248, 0207769249, 0207769250, 0207769251, 0207769252, 0207769253, 0207770721, 0207770722, 0207770723, 0207773147, 0207773148, 0207773149, 0207773150, 0207773151, 0207777434, 0207777435, 0207777436, 0207777437, 0207777438, 0207777439, 0207777440, 0207777441, 0207777442, 0207777443, 0207777444, 0207777445, 0207777446, 0207777447, 0207777878, 0207778540, 0207778541, 0207780669, 0207780670, 0207780671, 0207780672, 0207780673, 0207780761, 0207780762, 0207785216, 0207787723, 0207787724, 0207787725, 0207787726, 0207787727, 0207787728, 0207787729, 0207789155, 0207795322, 0207795323, 0207795324, 0207800432, 0207800433, 0207800481, 0207800482, 0207800483, 0207800484, 0207800485, 0207800486, 0207800487, 0207800494, 0207800495, 0207800496, 0207800497, 0207800498, 0207800500, 0207800509, 0207800510, 0207800511, 0207800512, 0207800655, 0207806145, 0207810062, 0207815195, 0207815196, 0207815197, 0207815198, 0207815211, 0207815225, 0207817620, 0207820647, 0207820648, 0207820726, 0207828400, 0207828401, 0207836043, 0207836044, 0207836119, 0207836120, 0207836121, 0207836122, 0207836123, 0207836201, 0207836212, 0207836213, 0207836214, 0207836215, 0207836247, 0207836248, 0207836249, 0207836250, 0207836251, 0207840219, 0207845496, 0207845535, 0207845537, 0207845549, 0207845550, 0207845689, 0207845691, 0207845692, 0207845706, 0207845707, 0207845708, 0207845709, 0207845710, 0207845711, 0207845712, 0207845713, 0207845726, 0207845748, 0207845749, 0207845750, 0207845882, 0207845883, 0207852426, 0207852427, 0207852428, 0207852429, 0207852430, 0207852431, 0207852432, 0207852433, 0207852434, 0207852435, 0207852436, 0207852437, 0207852868, 0207856675, 0207856676, 0207856677, 0207856678, 0207858516, 0207858517, 0207858538, 0207861043, 0207861044, 0207861705, 0207864175, 0207864176, 0207864177, 0207864361, 0207864362, 0207864363, 0207864364, 0207864365, 0207864366, 0207864367, 0207864368, 0207864369, 0207864370, 0207864371, 0207864372, 0207864383, 0207864384, 0207864385, 0207864398, 0207864828, 0207864829, 0207866171, 0207866172, 0207866173, 0207866174, 0207866326, 0207868463, 0207868464, 0207868465, 0207868466, 0207868467, 0207868468, 0207868469, 0207868470, 0207868471, 0207868472, 0207868473, 0207868474, 0207869975, 0207869976, 0207869977, 0207874443, 0207874444, 0207874863, 0207874864, 0207874865, 0207874866, 0207874867, 0207874868, 0207874869, 0207874870, 0207874871, 0207874872, 0207874873, 0207874874, 0207874875, 0207874876, 0207875002, 0207877664, 0207893315, 0207893316, 0207893619, 0207893620, 0207893621, 0207893622, 0207893623, 0207894148, 0207898073, 0207898074, 0207908591, 0207908592, 0207908593, 0207908669, 0207908670, 0207908671, 0207911576, 0207911942, 0207911943, 0207914091, 0207914191, 0207914192, 0207914193, 0207915278, 0207915279, 0207915400, 0207915401, 0207915402, 0207915403, 0207915404, 0207915405, 0207915406, 0207915407, 0207915408, 0207915409, 0207915410, 0207915411, 0207915412, 0207915413, 0207915414, 0207915415, 0207915416, 0207915417, 0207915418, 0207915419, 0207915420, 0207915421, 0207915422, 0207915423, 0207915424, 0207915425, 0207915426, 0207915427, 0207915428, 0207915429, 0207915430, 0207915431, 0207915432, 0207915433, 0207915434, 0207915718, 0207915719, 0207915751, 0207915752, 0207915753, 0207915754, 0207917203, 0207917204, 0207920752, 0207920849, 0207928098, 0207928099, 0207930499, 0207931724, 0207931725, 0207931726, 0207931727, 0207931728, 0207931729, 0207931730, 0207931731, 0207931732, 0207932178, 0207932179, 0207932420, 0207932422, 0207932423, 0207932425, 0207932922, 0207932931, 0207932932, 0207932933, 0207932936, 0207933359, 0207933620, 0207933621, 0207933622, 0207933623, 0207936872, 0207936873, 0207939381, 0207939382, 0207946682, 0207946683, 0207946685, 0207946687, 0207953913, 0207954846, 0207954854, 0207954855, 0207954856, 0207954857, 0207954858, 0207954859, 02079
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Neuromodulation
Reason for Recall:
Medtronic received reports that the DBS depth stop did not adequately secure to the lead, which can result in DBS lead placement beyond the intended target.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

DBS Lead Depth Stop contained in Medtronic Deep Brain Stimulation (DBS) lead kits, models 3387/3387S, 3389/3389S, and 3391/3391S, The lead kit contains one DBS lead in addition to multiple accessories used in DBS lead implant procedures. Among these accessories is the DBS lead holder, also referred to as the lead depth stop. The lead holder affixes to the DBS lead during the lead implant procedure, marking the distance to the lead tip and providing control of DBS lead depth during implant by interfacing with the stereotactic system

Product Codes/Lot Numbers:

Model 3387, Lot Serial #: 0206880294, 0206882647, 0206882779, 0206882780, 0206882781, 0206882782, 0206882783, 0206886184, 0206886323, 0206886328, 0206886330, 0206886331, 0206886332, 0206886333, 0206886603, 0206887813, 0206890443, 0206890444, 0206903488, 0206903489, 0206906792, 0206906793, 0206907159, 0206916391, 0206916392, 0206916393, 0206916394, 0206916395, 0206916396, 0206916397, 0206916398, 0206916399, 0206916797, 0206917120, 0206917121, 0206917122, 0206917123, 0206917124, 0206917125, 0206917126, 0206917127, 0206917128, 0206917129, 0206917130, 0206917131, 0206917132, 0206917302, 0206917303, 0206917304, 0206917306, 0206917307, 0206917308, 0206917310, 0206917312, 0206917560, 0206917561, 0206917563, 0206917564, 0206917565, 0206917566, 0206917568, 0206917569, 0206917570, 0206917572, 0206917573, 0206917574, 0206917575, 0206917576, 0206917577, 0206917578, 0206917579, 0206917620, 0206917621, 0206917622, 0206917623, 0206917624, 0206917625, 0206917626, 0206917627, 0206917628, 0206917629, 0206917630, 0206917631, 0206917632, 0206917633, 0206917634, 0206917635, 0206917636, 0206917637, 0206917638, 0206917639, 0206917660, 0206917661, 0206917662, 0206917663, 0206917664, 0206917665, 0206917666, 0206917667, 0206917668, 0206917669, 0206917670, 0206917671, 0206920443, 0206920444, 0206920445, 0206920447, 0206920448, 0206920449, 0206920450, 0206920451, 0206920452, 0206920453, 0206920454, 0206920455, 0206920456, 0206920457, 0206920458, 0206920459, 0206920460, 0206920461, 0206920462, 0206920463, 0206920663, 0206920664, 0206920665, 0206920666, 0206920667, 0206920668, 0206920669, 0206920670, 0206920671, 0206920672, 0206920673, 0206924158, 0206924492, 0206924493, 0206924494, 0206924604, 0206924605, 0206924606, 0206924607, 0206924608, 0206924609, 0206924610, 0206924611, 0206924612, 0206924613, 0206924614, 0206924615, 0206924616, 0206924617, 0206924618, 0206924619, 0206925140, 0206925141, 0206925142, 0206925143, 0206925245, 0206925290, 0206925291, 0206925292, 0206926111, 0206926112, 0206926414, 0206929827, 0206929828, 0206929829, 0206929830, 0206929849, 0206930146, 0206930736, 0206930737, 0206930941, 0206930942, 0206930943, 0206931910, 0206931911, 0206931927, 0206931928, 0206933408, 0206934345, 0206934346, 0206934347, 0206937919, 0206937920, 0206942506, 0206946796, 0206946798, 0206949034, 0206949035, 0206950164, 0206955498, 0206955499, 0206956649, 0206956650, 0206957515, 0206957516, 0206957517, 0206957518, 0206957519, 0206957801, 0206957802, 0206957803, 0206957804, 0206957805, 0206957806, 0206957807, 0206959205, 0206959206, 0206959207, 0206959208, 0206959209, 0206959210, 0206959211, 0206959212, 0206959213, 0206959214, 0206959215, 0206959216, 0206959217, 0206959218, 0206959220, 0206959221, 0206959222, 0206959223, 0206959224, 0206959501, 0206959502, 0206959503, 0206959682, 0206959684, 0206959686, 0206959687, 0206960104, 0206960105, 0206960106, 0206960107, 0206960108, 0206960109, 0206960147, 0206960148, 0206960149, 0206960655, 0206960658, 0206960666, 0206960667, 0206960668, 0206960669, 0206960670, 0206960671, 0206960676, 0206960677, 0206960678, 0206960679, 0206960680, 0206960681, 0206960694, 0206960695, 0206960696, 0206969559, 0206969560, 0206969561, 0206969562, 0206969563, 0206969564, 0206969565, 0206969566, 0206969567, 0206969568, 0206969569, 0206969570, 0206969571, 0206969572, 0206969573, 0206969574, 0206969575, 0206969576, 0206969577, 0206969578, 0206969691, 0206969692, 0206970927, 0206973200, 0206973201, 0206973202, 0206973203, 0206973204, 0206973205, 0206973206, 0206973800, 0206973801, 0206978789, 0206980541, 0206980542, 0206980543, 0206980970, 0206980971, 0206981705, 0206981706, 0206981707, 0206981714, 0206981715, 0206981754, 0206981771, 0206981772, 0206984645, 0206984646, 0206984650, 0206984675, 0206991945, 0206991946, 0206991947, 0206991948, 0206991949, 0206991950, 0206991951, 0206991952, 0206991953, 0206991954, 0206991955, 0206991956, 0206991957, 0206991958, 0206991959, 0206991960, 0206991961, 0206991962, 0206991963, 0206991964, 0206991965, 0206991966, 0206991967, 0206991968, 0206991969, 0206991970, 0206991971, 0206991972, 0206991973, 0206991974, 0206991975, 0206991976, 0206991977, 0206991978, 0206991979, 0206991984, 0206991985, 0206991986, 0206991987, 0206991988, 0206991989, 0206991990, 0206991991, 0206991992, 0206991993, 0206991994, 0206991995, 0206991996, 0206991997, 0206991998, 0206991999, 0206992000, 0206992001, 0206992002, 0206992003, 0206992004, 0206992005, 0206992006, 0206992007, 0206992008, 0206992009, 0206992010, 0206992011, 0206992012, 0206992013, 0206992014, 0206992015, 0206992016, 0206992017, 0206992018, 0206992019, 0206992020, 0206992021, 0206992022, 0206992023, 0206992024, 0206992025, 0206992026, 0206992027, 0206992028, 0206992029, 0206992030, 0206992031, 0206992032, 0206992033, 0206992034, 0206992035, 0206992036, 0206992037, 0206992038, 0206992039, 0206992040, 0206996540, 0206996541, 0206996542, 0206996543, 0206996544, 0206996545, 0206996546, 0206996547, 0206996548, 0206996549, 0206996550, 0206996551, 0206996552, 0206996553, 0206996554, 0206996555, 0206996556, 0206996557, 0206996558, 0206996559, 0206996560, 0206996561, 0206996562, 0206996563, 0207000225, 0207006520, 0207006777, 0207006778, 0207006779, 0207006780, 0207007919, 0207008220, 0207008221, 0207008222, 0207011212, 0207011213, 0207015958, 0207016168, 0207016796, 0207018882, 0207018883, 0207018884, 0207021792, 0207021793, 0207021794, 0207021795, 0207021796, 0207021797, 0207021798, 0207021799, 0207021800, 0207021801, 0207021802, 0207021803, 0207022120, 0207022121, 0207022122, 0207027753, 0207030645, 0207030646, 0207030647, 0207030648, 0207030649, 0207030650, 0207030651, 0207030652, 0207030653, 0207030654, 0207030655, 0207030656, 0207030657, 0207030658, 0207030659, 0207030780, 0207030781, 0207030980, 0207030981, 0207030982, 0207030983, 0207030985, 0207038190, 0207038192, 0207038195, 0207038423, 0207038424, 0207038425, 0207038426, 0207038427, 0207038434, 0207038435, 0207038436, 0207038437, 0207038438, 0207038439, 0207038440, 0207038441, 0207038442, 0207038443, 0207038457, 0207038458, 0207038459, 0207038460, 0207038461, 0207038462, 0207040184, 0207040185, 0207040187, 0207040188, 0207040189, 0207040190, 0207040191, 0207040239, 0207040240, 0207040241, 0207040242, 0207040243, 0207040244, 0207040573, 0207040574, 0207040575, 0207040576, 0207040712, 0207040713, 0207040714, 0207040715, 0207040718, 0207040719, 0207041016, 0207041811, 0207043628, 0207043629, 0207043630, 0207043631, 0207043632, 0207043658, 0207043659, 0207043660, 0207043871, 0207043872, 0207043874, 0207043875, 0207047746, 0207052711, 0207054947, 0207056625, 0207056626, 0207056628, 0207056629, 0207056630, 0207056631, 0207056807, 0207057256, 0207057257, 0207057427, 0207057428, 0207057429, 0207057430, 0207057431, 0207057432, 0207057433, 0207057434, 0207057435, 0207057436, 0207057437, 0207057438, 0207057439, 0207057440, 0207057441, 0207057442, 0207059495, 0207059565, 0207059566, 0207059567, 0207059956, 0207059959, 0207059963, 0207059985, 0207059986, 0207059987, 0207062240, 0207063038, 0207063039, 0207063040, 0207063133, 0207063134, 0207063135, 0207063136, 0207063137, 0207063138, 0207063139, 0207063140, 0207063141, 0207063142, 0207067134, 0207067135, 0207067136, 0207067137, 0207067138, 0207067139, 0207067140, 0207067141, 0207067142, 0207067143, 0207067146, 0207067147, 0207067148, 0207067847, 0207067848, 0207068325, 0207068326, 0207068327, 0207068449, 0207070300, 0207070321, 0207071704, 0207071705, 0207072960, 0207072961, 0207072962, 0207072963, 0207072964, 0207072965, 0207072966, 0207072967, 0207072968, 0207072969, 0207072970, 0207072971, 0207072972, 0207072973, 0207072974, 0207079003, 0207079004, 0207079339, 0207079340, 0207079341, 0207079342, 0207079343, 0207079414, 0207079415, 0207079416, 0207079417, 0207079418, 0207079419, 0207079435, 0207079436, 0207079437, 0207079438, 0207079439, 0207079440, 0207079441, 0207079467, 0207079468, 0207079469, 0207079470, 0207079471, 0207079472, 0207079473, 0207079474, 0207079475, 0207079476, 0207079477, 0207079478, 0207079479, 0207079480, 0207079481, 0207079482, 0207079483, 0207079484, 0207079485, 0207079486, 0207079487, 0207079488, 0207079489, 0207079490, 0207079491, 0207079492, 0207079493, 0207079494, 0207079495, 0207079496, 0207079497, 0207079498, 0207079641, 0207079642, 0207079643, 0207079645, 0207079680, 0207079681, 0207079701, 0207083910, 0207083911, 0207083912, 0207084903, 0207084904, 0207087901, 0207087902, 0207087903, 0207087904, 0207087905, 0207087906, 0207087907, 0207087908, 0207087909, 0207087910, 0207087951, 0207087952, 0207088003, 0207088004, 0207088005, 0207088431, 0207094140, 0207094142, 0207097722, 0207097723, 0207098161, 0207098162, 0207098163, 0207098164, 0207100512, 0207100514, 0207100515, 0207100516, 0207100517, 0207100810, 0207100848, 0207101008, 0207101009, 0207101010, 0207101023, 0207101031, 0207101053, 0207101120, 0207105162, 0207105163, 0207105164, 0207105165, 0207105166, 0207105167, 0207105168, 0207105169, 0207105170, 0207105171, 0207105172, 0207105176, 0207105321, 0207105322, 0207105323, 0207105324, 0207105325, 0207105326, 0207105327, 0207105328, 0207105329, 0207105330, 0207105331, 0207105332, 0207105333, 0207105334, 0207105335, 0207105336, 0207105352, 0207105356, 0207108624, 0207108625, 0207108635, 0207108636, 0207109103, 0207109704, 0207109705, 0207109706, 0207109708, 0207109709, 0207109710, 0207109711, 0207109712, 0207109713, 0207109714, 0207109715, 0207109716, 0207109717, 0207109718, 0207109719, 0207109720, 0207109725, 0207109726, 0207109728, 0207109729, 0207109733, 0207109734, 0207109737, 0207109815, 0207110541, 0207110542, 0207110543, 0207112954, 0207113814, 0207113815, 0207113816, 0207113817, 0207113818, 0207114006, 0207114007, 0207114746, 0207114852, 0207114853, 0207114854, 0207114855, 0207117588, 0207117589, 0207117590, 0207117741, 0207118073, 0207119315, 0207119316, 0207119317, 0207119318, 0207119319, 0207120110, 0207121868, 0207122061, 0207122062, 0207122063, 0207122064, 0207122065, 0207122066, 0207122067, 0207122068, 0207122069, 0207122070, 0207122071, 0207122072, 0207122099, 0207122100, 0207122101, 0207122127, 0207122128, 0207122129, 0207122183, 0207125639, 0207125641, 0207125642, 0207125643, 0207125754, 0207125851, 0207125876, 0207129909, 0207129910, 0207129912, 0207129913, 0207129917, 0207129920, 0207129921, 0207129922, 0207129923, 0207130018, 0207130019, 0207130020, 0207130021, 0207130022, 0207130023, 0207130024, 0207130025, 0207130026, 0207130027, 0207130028, 0207130029, 0207130081, 0207130299, 0207130300, 0207130301, 0207130302, 0207130303, 0207130304, 0207130305, 0207130306, 0207130307, 0207130308, 0207130724, 0207130786, 0207130787, 0207130788, 0207130789, 0207130790, 0207130845, 0207130846, 0207130847, 0207130848, 0207130849, 0207130850, 0207130851, 0207131732, 0207131734, 0207131757, 0207131758, 0207131759, 0207131780, 0207132032, 0207132033, 0207132098, 0207132677, 0207135004, 0207136742, 0207136743, 0207139918, 0207139919, 0207139920, 0207139921, 0207139922, 0207139923, 0207139924, 0207139925, 0207139926, 0207139927, 0207140714, 0207140715, 0207140716, 0207140717, 0207140718, 0207140719, 0207140720, 0207140724, 0207140725, 0207141032, 0207141033, 0207141034, 0207141035, 0207141036, 0207143663, 0207143678, 0207143679, 0207143680, 0207143681, 0207143682, 0207143683, 0207143697, 0207143698, 0207143700, 0207143701, 0207143707, 0207143708, 0207143712, 0207143713, 0207143714, 0207143715, 0207143716, 0207143717, 0207143718, 0207143719, 0207144809, 0207144817, 0207144818, 0207144819, 0207144840, 0207144841, 0207144842, 0207144843, 0207144844, 0207144846, 0207144847, 0207144848, 0207144883, 0207144884, 0207144885, 0207144886, 0207144887, 0207144888, 0207144889, 0207144890, 0207145866, 0207145867, 0207148304, 0207148305, 0207148306, 0207159258, 0207159259, 0207159260, 0207160611, 0207160612, 0207160614, 0207160615, 0207160678, 0207160679, 0207160680, 0207160708, 0207160709, 0207160710, 0207161115, 0207163879, 0207163881, 0207163882, 0207163883, 0207163884, 0207163887, 0207163889, 0207164673, 0207164674, 0207164675, 0207164676, 0207164677, 0207164678, 0207164870, 0207164871, 0207164872, 0207168561, 0207168562, 0207168563, 0207168564, 0207168565, 0207168566, 0207168567, 0207168568, 0207168569, 0207168570, 0207168571, 0207168572, 0207168573, 0207168574, 0207168575, 0207168576, 0207168577, 0207168595, 0207168596, 0207168597, 0207168598, 0207168599, 0207168600, 0207168929, 0207168930, 0207168931, 0207168932, 0207168954, 0207168955, 0207168956, 0207168957, 0207168959, 0207168961, 0207168962, 0207168963, 0207168965, 0207168966, 0207168967, 0207168968, 0207168969, 0207168970, 0207168972, 0207168979, 0207168980, 0207168985, 0207168986, 0207168987, 0207169010, 0207174625, 0207174627, 0207174628, 0207174629, 0207174630, 0207174654, 0207175291, 0207175292, 0207175293, 0207175294, 0207175295, 0207175296, 0207176745, 0207177151, 0207177152, 0207177153, 0207177154, 0207180633, 0207180634, 0207180635, 0207180636, 0207180637, 0207180638, 0207180639, 0207180640, 0207180641, 0207180642, 0207180643, 0207180644, 0207180645, 0207180646, 0207180647, 0207180648, 0207180649, 0207180650, 0207180651, 0207181930, 0207181931, 0207181932, 0207181933, 0207181934, 0207181987, 0207181988, 0207181989, 0207181990, 0207181991, 0207181992, 0207181993, 0207181994, 0207181995, 0207181996, 0207181997, 0207182077, 0207182782, 0207182783, 0207182785, 0207182786, 0207182787, 0207182788, 0207182790, 0207182791, 0207182793, 0207182794, 0207182795, 0207182796, 0207182797, 0207182798, 0207182799, 0207182840, 0207182841, 0207182842, 0207182843, 0207182844, 0207182846, 0207185517, 0207185518, 0207185519, 0207185520, 0207185521, 0207185522, 0207185523, 0207185524, 0207185525, 0207185636, 0207185651, 0207192753, 0207192754, 0207192790, 0207192791, 0207192815, 0207192816, 0207192817, 0207192818, 0207192819, 0207192820, 0207192821, 0207192822, 0207192823, 0207192824, 0207192825, 0207192826, 0207192827, 0207192828, 0207192829, 0207192910, 0207192911, 0207202710, 0207202711, 0207202712, 0207202717, 0207202718, 0207202719, 0207202740, 0207202741, 0207202742, 0207202743, 0207202744, 0207202745, 0207202746, 0207202747, 0207202748, 0207202750, 0207202754, 0207202755, 0207202756, 0207202757, 0207202758, 0207202764, 0207202765, 0207202766, 0207202801, 0207202805, 0207202932, 0207202936, 0207202943, 0207202944, 0207202961, 0207203570, 0207203571, 0207208667, 0207208668, 0207208714, 0207208715, 0207208716, 0207208723, 0207220085, 0207220740, 0207220741, 0207220742, 0207220743, 0207220745, 0207220746, 0207225509, 0207225523, 0207225564, 0207225565, 0207225566, 0207225567, 0207225568, 0207225569, 0207225570, 0207225571, 0207226531, 0207226532, 0207226533, 0207226534, 0207226535, 0207226626, 0207226627, 0207226628, 0207226629, 0207226630, 0207226631, 0207226632, 0207226633, 0207226634, 0207226635, 0207226636, 0207226637, 0207230154, 0207232091, 0207232126, 0207232266, 0207234900, 0207234901, 0207234902, 0207240828, 0207240832, 0207240833, 0207240834, 0207240841, 0207240842, 0207240843, 0207240844, 0207240845, 0207240856, 0207240857, 0207240858, 0207241071, 0207241072, 0207241073, 0207241444, 0207241445, 0207241455, 0207241456, 0207241457, 0207241458, 0207241459, 0207241460, 0207241461, 0207241463, 0207246705, 0207247848, 0207247849, 0207247850, 0207250567, 0207250707, 0207250708, 0207250710, 0207250711, 0207250712, 0207250713, 0207250714, 0207250715, 0207250716, 0207250717, 0207250718, 0207250719, 0207250720, 0207250721, 0207250722, 0207250723, 0207250724, 0207250725, 0207250726, 0207250727, 0207250728, 0207250729, 0207250730, 0207250731, 0207250809, 0207250810, 0207250812, 0207250813, 0207250814, 0207250815, 0207250816, 0207250817, 0207250818, 0207250829, 0207254564, 0207254565, 0207254742, 0207255595, 0207257987, 0207257988, 0207257989, 0207257990, 0207257991, 0207258020, 0207258045, 0207258259, 0207258280, 0207258638, 0207258639, 0207258640, 0207258641, 0207259128, 0207259174, 0207259175, 0207260433, 0207260434, 0207261780, 0207261788, 0207261789, 0207261790, 0207266491, 0207266492, 0207266530, 0207266531, 0207266532, 0207268177, 0207268178, 0207268179, 0207268231, 0207268232, 0207270938, 0207270939, 0207270972, 0207270973, 0207275052, 0207275053, 0207275054, 0207275058, 0207275218, 0207275219, 0207278567, 0207278568, 0207278569, 0207278570, 0207278572, 0207278573, 0207278574, 0207278575, 0207278576, 0207278577, 0207278578, 0207278579, 0207278639, 0207279750, 0207280151, 0207280180, 0207280181, 0207280200, 0207280201, 0207280202, 0207280203, 0207280204, 0207280205, 0207280206, 0207280207, 0207280208, 0207280209, 0207280210, 0207280211, 0207280217, 0207280218, 0207280219, 0207280220, 0207280221, 0207280222, 0207280223, 0207280224, 0207280225, 0207280226, 0207280227, 0207280228, 0207280229, 0207280230, 0207280231, 0207280232, 0207280233, 0207280238, 0207280239, 0207283309, 0207283310, 0207283614, 0207283627, 0207283628, 0207283629, 0207283630, 0207283631, 0207283632, 0207283638, 0207283639, 0207283841, 0207283883, 0207283884, 0207283885, 0207283886, 0207283903, 0207283944, 0207283945, 0207283946, 0207283947, 0207283948, 0207283949, 0207283950, 0207283951, 0207283952, 0207283953, 0207283954, 0207283955, 0207283956, 0207283957, 0207288214, 0207288219, 0207289211, 0207289212, 0207289214, 0207289215, 0207289216, 0207291547, 0207291548, 0207291549, 0207291686, 0207291687, 0207291688, 0207294519, 0207298853, 0207298854, 0207298855, 0207298856, 0207298857, 0207298858, 0207298859, 0207298996, 0207299404, 0207299920, 0207299921, 0207299922, 0207299923, 0207299924, 0207299925, 0207299926, 0207299927, 0207299928, 0207300520, 0207300521, 0207300522, 0207300523, 0207300699, 0207301220, 0207301221, 0207303119, 0207303499, 0207303641, 0207303656, 0207303885, 0207303886, 0207304245, 0207304800, 0207304801, 0207305118, 0207305119, 0207306460, 0207308323, 0207308324, 0207308325, 0207308326, 0207308327, 0207308328, 0207309474, 0207309475, 0207309476, 0207309477, 0207309478, 0207309479, 0207309480, 0207309481, 0207309482, 0207309483, 0207309484, 0207309485, 0207309486, 0207309487, 0207309488, 0207309489, 0207309490, 0207309491, 0207309492, 0207309493, 0207309494, 0207309495, 0207309496, 0207309497, 0207309498, 0207309499, 0207309500, 0207309501, 0207309502, 0207309505, 0207309506, 0207309507, 0207309508, 0207309516, 0207309517, 0207309518, 0207309519, 0207309520, 0207309521, 0207309522, 0207309523, 0207309524, 0207309525, 0207309526, 0207309527, 0207309530, 0207309531, 0207309532, 0207309533, 0207309534, 0207309535, 0207309536, 0207319594, 0207324654, 0207324655, 0207324656, 0207324657, 0207324658, 0207324659, 0207324680, 0207324681, 0207324682, 0207324683, 0207324684, 0207324685, 0207324686, 0207324687, 0207324688, 0207324689, 0207324690, 0207324691, 0207324692, 0207324693, 0207324694, 0207324695, 0207324696, 0207324697, 0207324698, 0207324699, 0207324700, 0207326089, 0207326090, 0207326091, 0207326131, 0207326132, 0207327178, 0207337977, 0207342360, 0207348166, 0207348167, 0207348168, 0207348211, 0207348212, 0207348214, 0207348215, 0207348216, 0207348217, 0207348218, 0207348583, 0207350341, 0207353632, 0207353633, 0207353634, 0207353751, 0207353898, 0207359405, 0207359406, 0207359407, 0207359408, 0207359409, 0207359410, 0207359411, 0207359412, 0207359413, 0207359414, 0207359415, 0207359416, 0207359417, 0207359418, 0207359419, 0207360423, 0207360424, 0207360425, 0207360426, 0207360427, 0207360428, 0207360429, 0207360430, 0207360431, 0207360432, 0207360433, 0207360434, 0207360435, 0207360436, 0207360437, 0207360438, 0207360439, 0207361120, 0207361121, 0207361122, 0207361123, 0207361280, 0207361281, 0207361282, 0207361283, 0207361284, 0207361285, 0207361286, 0207361287, 0207361348, 0207368928, 0207368929, 0207368930, 0207368931, 0207368932, 0207368933, 0207368934, 0207368935, 0207372779, 0207372793, 0207372892, 0207372893, 0207372895, 0207372896, 0207379102, 0207379355, 0207379356, 0207379405, 0207379406, 0207380120, 0207380121, 0207380122, 0207380123, 0207385226, 0207396642, 0207396707, 0207396708, 0207397023, 0207397864, 0207397880, 0207397887, 0207397888, 0207397890, 0207397892, 0207397896, 0207397897, 0207397898, 0207397899, 0207397900, 0207397901, 0207397902, 0207397904, 0207397905, 0207397906, 0207397907, 0207397956, 0207401852, 0207401853, 0207404964, 0207407581, 0207409006, 0207409007, 0207409008, 0207409009, 0207413918, 0207413919, 0207414060, 0207414061, 0207414062, 0207414063, 0207414064, 0207414065, 0207414066, 0207414067, 0207419438, 0207422718, 0207424300, 0207424302, 0207424303, 0207424304, 0207424305, 0207424756, 0207428968, 0207428969, 0207428970, 0207428971, 0207428972, 0207428973, 0207430334, 0207430399, 0207430403, 0207430404, 0207430405, 0207432103, 0207432104, 0207432105, 0207432666, 0207437724, 0207441876, 0207444058, 0207444059, 0207444300, 0207446984, 0207448988, 0207448989, 0207448990, 0207448991, 0207448992, 0207448996, 0207449981, 0207449982, 0207449993, 0207449994, 0207449995, 0207449996, 0207449997, 0207449998, 0207454223, 0207454224, 0207454225, 0207454226, 0207454227, 0207454228, 0207454229, 0207454401, 0207454402, 0207454403, 0207454404, 0207454405, 0207454406, 0207454407, 0207454408, 0207454409, 0207456279, 0207457340, 0207457341, 0207457342, 0207457343, 0207457344, 0207457345, 0207457346, 0207457347, 0207457348, 0207457349, 0207457350, 0207457351, 0207457352, 0207457353, 0207457354, 0207457355, 0207457356, 0207457357, 0207457358, 0207457359, 0207457360, 0207457361, 0207457362, 0207457363, 0207457364, 0207457365, 0207457366, 0207457367, 0207457368, 0207457369, 0207457370, 0207457371, 0207457372, 0207457373, 0207457374, 0207457375, 0207457376, 0207457377, 0207457378, 0207457379, 0207457380, 0207457381, 0207457382, 0207457383, 0207457384, 0207457385, 0207457386, 0207457387, 0207457388, 0207457389, 0207457390, 0207457391, 0207457392, 0207457393, 0207457394, 0207457395, 0207457396, 0207457397, 0207457398, 0207457399, 0207457400, 0207457401, 0207457402, 0207457403, 0207457404, 0207457405, 0207457406, 0207457407, 0207457408, 0207457409, 0207457410, 0207457411, 0207457412, 0207457413, 0207457414, 0207457415, 0207457416, 0207457417, 0207457418, 0207457419, 0207457420, 0207457421, 0207457422, 0207460357, 0207460358, 0207460359, 0207460360, 0207460361, 0207460871, 0207460872, 0207466531, 0207466532, 0207466591, 0207466592, 0207466593, 0207466594, 0207466595, 0207466596, 0207466597, 0207466598, 0207466599, 0207466600, 0207466601, 0207466602, 0207466603, 0207466604, 0207466605, 0207466606, 0207466607, 0207466608, 0207466674, 0207469251, 0207469252, 0207471095, 0207471096, 0207474219, 0207498243, 0207498244, 0207498245, 0207498246, 0207498247, 0207498250, 0207498251, 0207507151, 0207507152, 0207507153, 0207507154, 0207507447, 0207507448, 0207507449, 0207520659, 0207523074, 0207526687, 0207531718, 0207531719, 0207531720, 0207534092, 0207534093, 0207534573, 0207534575, 0207536498, 0207536499, 0207536500, 0207536501, 0207536503, 0207536504, 0207536505, 0207536506, 0207538329, 0207538744, 0207538745, 0207541018, 0207546649, 0207546650, 0207546651, 0207546652, 0207546653, 0207546654, 0207546655, 0207546656, 0207546657, 0207546658, 0207547088, 0207547089, 0207547090, 0207547091, 0207547092, 0207547093, 0207547094, 0207547095, 0207547096, 0207547097, 0207547098, 0207547099, 0207547100, 0207547101, 0207547102, 0207547111, 0207547112, 0207547113, 0207547114, 0207547115, 0207547116, 0207547117, 0207547118, 0207547119, 0207547120, 0207547121, 0207547122, 0207547123, 0207547311, 0207548014, 0207548015, 0207548639, 0207548883, 0207548884, 0207548885, 0207548886, 0207549280, 0207550658, 0207554148, 0207555823, 0207561059, 0207561060, 0207567863, 0207574365, 0207585359, 0207588282, 0207588640, 0207588643, 0207588644, 0207588645, 0207588672, 0207588674, 0207588808, 0207588809, 0207588810, 0207588811, 0207588860, 0207588861, 0207588865, 0207588866, 0207592741, 0207595000, 0207595001, 0207595002, 0207595003, 0207595017, 0207598167, 0207598168, 0207598169, 0207600533, 0207600534, 0207600535, 0207601075, 0207602702, 0207604544, 0207605819, 0207606004, 0207609972, 0207609973, 0207609974, 0207609975, 0207609976, 0207609977, 0207609978, 0207611974, 0207611975, 0207612377, 0207615806, 0207615807, 0207615808, 0207615809, 0207615810, 0207615811, 0207615812, 0207615813, 0207615814, 0207615815, 0207615816, 0207615817, 0207615818, 0207615819, 0207615835, 0207615836, 0207620203, 0207620204, 0207620205, 0207620206, 0207620207, 0207620364, 0207620790, 0207620791, 0207620793, 0207620794, 0207620795, 0207620796, 0207621611, 0207621612, 0207621613, 0207621614, 0207621615, 0207621616, 0207621617, 0207621618, 0207621619, 0207621620, 0207621621, 0207621622, 0207621623, 0207621624, 0207621625, 0207621626, 0207621627, 0207621628, 0207621629, 0207621630, 0207621631, 0207621632, 0207621633, 0207621634, 0207621635, 0207621636, 0207621637, 0207621638, 0207621639, 0207621640, 0207621641, 0207621642, 0207621643, 0207621644, 0207621645, 0207621646, 0207621647, 0207621648, 0207621649, 0207621650, 0207621651, 0207621652, 0207621653, 0207621654, 0207621830, 0207621831, 0207622044, 0207622045, 0207622169, 0207622176, 0207622177, 0207622178, 0207622179, 0207623055, 0207625459, 0207627667, 0207627668, 0207627669, 0207627690, 0207627691, 0207627692, 0207627693, 0207627694, 0207627695, 0207627696, 0207627949, 0207628172, 0207631094, 0207632116, 0207634225, 0207634233, 0207637637, 0207642804, 0207642807, 0207642813, 0207647090, 0207647249, 0207651194, 0207651195, 0207651196, 0207651197, 0207651352, 0207652932, 0207652933, 0207653543, 0207660129, 0207660130, 0207660131, 0207660132, 0207660133, 0207660134, 0207660135, 0207660136, 0207660171, 0207660878, 0207664645, 0207664646, 0207664647, 0207670273, 0207670274, 0207670278, 0207670279, 0207674827, 0207674828, 0207674829, 0207681428, 0207682857, 0207682886, 0207682888, 0207682890, 0207682892, 0207682903, 0207682904, 0207682906, 0207682908, 0207682909, 0207682913, 0207682917, 0207682918, 0207682940, 0207682942, 0207682944, 0207682946, 0207682948, 0207682950, 0207682951, 0207682953, 0207682955, 0207682957, 0207682959, 0207682960, 0207682962, 0207682964, 0207683075, 0207683076, 0207683077, 0207683078, 0207683079, 0207683100, 0207683101, 0207683103, 0207683104, 0207683105, 0207683106, 0207683108, 0207683109, 0207683110, 0207683111, 0207683112, 0207683113, 0207683114, 0207683115, 0207683116, 0207683117, 0207684419, 0207684662, 0207684663, 0207685023, 0207685138, 0207685139, 0207685454, 0207687157, 0207687571, 0207687572, 0207692231, 0207692232, 0207694357, 0207699502, 0207699503, 0207699504, 0207699507, 0207700963, 0207704751, 0207704753, 0207704959, 0207704960, 0207705029, 0207705030, 0207705031, 0207705032, 0207705033, 0207705034, 0207705035, 0207705036, 0207705041, 0207705042, 0207705075, 0207708402, 0207710979, 0207711007, 0207720013, 0207720602, 0207721774, 0207722086, 0207729498, 0207729499, 0207730540, 0207730541, 0207730542, 0207730543, 0207730544, 0207730545, 0207730546, 0207730547, 0207730548, 0207730549, 0207730550, 0207731594, 0207731595, 0207733668, 0207733678, 0207733695, 0207733696, 0207733697, 0207733698, 0207733699, 0207733819, 0207734053, 0207734799, 0207735690, 0207737625, 0207737626, 0207737627, 0207737628, 0207737629, 0207737630, 0207737838, 0207737839, 0207742414, 0207742415, 0207742416, 0207742417, 0207742418, 0207742419, 0207742420, 0207742421, 0207742422, 0207742423, 0207742452, 0207742453, 0207742454, 0207742455, 0207742456, 0207742457, 0207742458, 0207742459, 0207743128, 0207747304, 0207747305, 0207747306, 0207747307, 0207747308, 0207747309, 0207747417, 0207747418, 0207747419, 0207750080, 0207752765, 0207752766, 0207752767, 0207752768, 0207752769, 0207752770, 0207752771, 0207752772, 0207756643, 0207756644, 0207756645, 0207756646, 0207756648, 0207756716, 0207756717, 0207756718, 0207756719, 0207756720, 0207756721, 0207756722, 0207756757, 0207757244, 0207757245, 0207757246, 0207757247, 0207757352, 0207757353, 0207757354, 0207757355, 0207757356, 0207757357, 0207757358, 0207757359, 0207757361, 0207757362, 0207757451, 0207757452, 0207757453, 0207761819, 0207761900, 0207762185, 0207762186, 0207762187, 0207762188, 0207762189, 0207762190, 0207762191, 0207762192, 0207762193, 0207762194, 0207762195, 0207762196, 0207762197, 0207762198, 0207762199, 0207762200, 0207762201, 0207762202, 0207762203, 0207762204, 0207762205, 0207762206, 0207762207, 0207762208, 0207762209, 0207762210, 0207762211, 0207762212, 0207762213, 0207762214, 0207762215, 0207762216, 0207762217, 0207762218, 0207762681, 0207762682, 0207762683, 0207762684, 0207762722, 0207762723, 0207764452, 0207767622, 0207767623, 0207767624, 0207767625, 0207767626, 0207767627, 0207767628, 0207767629, 0207767928, 0207767929, 0207768006, 0207768007, 0207768008, 0207768009, 0207768010, 0207768011, 0207768012, 0207768013, 0207769247, 0207769248, 0207769249, 0207769250, 0207769251, 0207769252, 0207769253, 0207770721, 0207770722, 0207770723, 0207773147, 0207773148, 0207773149, 0207773150, 0207773151, 0207777434, 0207777435, 0207777436, 0207777437, 0207777438, 0207777439, 0207777440, 0207777441, 0207777442, 0207777443, 0207777444, 0207777445, 0207777446, 0207777447, 0207777878, 0207778540, 0207778541, 0207780669, 0207780670, 0207780671, 0207780672, 0207780673, 0207780761, 0207780762, 0207785216, 0207787723, 0207787724, 0207787725, 0207787726, 0207787727, 0207787728, 0207787729, 0207789155, 0207795322, 0207795323, 0207795324, 0207800432, 0207800433, 0207800481, 0207800482, 0207800483, 0207800484, 0207800485, 0207800486, 0207800487, 0207800494, 0207800495, 0207800496, 0207800497, 0207800498, 0207800500, 0207800509, 0207800510, 0207800511, 0207800512, 0207800655, 0207806145, 0207810062, 0207815195, 0207815196, 0207815197, 0207815198, 0207815211, 0207815225, 0207817620, 0207820647, 0207820648, 0207820726, 0207828400, 0207828401, 0207836043, 0207836044, 0207836119, 0207836120, 0207836121, 0207836122, 0207836123, 0207836201, 0207836212, 0207836213, 0207836214, 0207836215, 0207836247, 0207836248, 0207836249, 0207836250, 0207836251, 0207840219, 0207845496, 0207845535, 0207845537, 0207845549, 0207845550, 0207845689, 0207845691, 0207845692, 0207845706, 0207845707, 0207845708, 0207845709, 0207845710, 0207845711, 0207845712, 0207845713, 0207845726, 0207845748, 0207845749, 0207845750, 0207845882, 0207845883, 0207852426, 0207852427, 0207852428, 0207852429, 0207852430, 0207852431, 0207852432, 0207852433, 0207852434, 0207852435, 0207852436, 0207852437, 0207852868, 0207856675, 0207856676, 0207856677, 0207856678, 0207858516, 0207858517, 0207858538, 0207861043, 0207861044, 0207861705, 0207864175, 0207864176, 0207864177, 0207864361, 0207864362, 0207864363, 0207864364, 0207864365, 0207864366, 0207864367, 0207864368, 0207864369, 0207864370, 0207864371, 0207864372, 0207864383, 0207864384, 0207864385, 0207864398, 0207864828, 0207864829, 0207866171, 0207866172, 0207866173, 0207866174, 0207866326, 0207868463, 0207868464, 0207868465, 0207868466, 0207868467, 0207868468, 0207868469, 0207868470, 0207868471, 0207868472, 0207868473, 0207868474, 0207869975, 0207869976, 0207869977, 0207874443, 0207874444, 0207874863, 0207874864, 0207874865, 0207874866, 0207874867, 0207874868, 0207874869, 0207874870, 0207874871, 0207874872, 0207874873, 0207874874, 0207874875, 0207874876, 0207875002, 0207877664, 0207893315, 0207893316, 0207893619, 0207893620, 0207893621, 0207893622, 0207893623, 0207894148, 0207898073, 0207898074, 0207908591, 0207908592, 0207908593, 0207908669, 0207908670, 0207908671, 0207911576, 0207911942, 0207911943, 0207914091, 0207914191, 0207914192, 0207914193, 0207915278, 0207915279, 0207915400, 0207915401, 0207915402, 0207915403, 0207915404, 0207915405, 0207915406, 0207915407, 0207915408, 0207915409, 0207915410, 0207915411, 0207915412, 0207915413, 0207915414, 0207915415, 0207915416, 0207915417, 0207915418, 0207915419, 0207915420, 0207915421, 0207915422, 0207915423, 0207915424, 0207915425, 0207915426, 0207915427, 0207915428, 0207915429, 0207915430, 0207915431, 0207915432, 0207915433, 0207915434, 0207915718, 0207915719, 0207915751, 0207915752, 0207915753, 0207915754, 0207917203, 0207917204, 0207920752, 0207920849, 0207928098, 0207928099, 0207930499, 0207931724, 0207931725, 0207931726, 0207931727, 0207931728, 0207931729, 0207931730, 0207931731, 0207931732, 0207932178, 0207932179, 0207932420, 0207932422, 0207932423, 0207932425, 0207932922, 0207932931, 0207932932, 0207932933, 0207932936, 0207933359, 0207933620, 0207933621, 0207933622, 0207933623, 0207936872, 0207936873, 0207939381, 0207939382, 0207946682, 0207946683, 0207946685, 0207946687, 0207953913, 0207954846, 0207954854, 0207954855, 0207954856, 0207954857, 0207954858, 0207954859, 02079

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-3069-2017

Related Recalls

Class I - Dangerous

Evaluation handsets may not be able to communicate with the neurostimulator due to handsets having a faster processing speed than prior handsets. An inability to connect may cause the therapy to stop being delivered and a system error will be displayed. Patient may experience return of underlying disease symptoms (e.g. overactive bladder, urinary retention, or fecal incontinence) based on their response to stimulation during the evaluation period.

Apr 3, 2025 Other Medical Devices Nationwide View Details →