Plum 360 Infusion System
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model No. 30010
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ICU Medical Inc
- Reason for Recall:
- Potential for the Connectivity Engine Module of the affected unit to disengage from the main chassis, causing the unit to shut down without an alarm notification. This would delay initiation of therapy or interrupt an infusion.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Plum 360 Infusion System
Product Codes/Lot Numbers:
Model No. 30010
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-3031-2017
Related Recalls
Class I - Dangerous
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Class I - Dangerous
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
Class I - Dangerous
Port weld of drug transfer device may separate or break during use and potentially result in a leak.