Impella CP Smart Assist Set; Product Code: 1000402;

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Product Code: 1000402; UDI-DI: 00813502012873; Serial Numbers: 385946 385942 385944 385945 385950 385941 385949 385947 385948 385943 392855 392853 392854 392852 392851 392848 392849 392879 392878 427388 427387 427386 430896 430875 430874 430873 430872 430871 430870 430869 430868 430867 431652 430895 430894 486297 486298 486299 486300; Batch Numbers: 2023251705 2023251861 2023251862 2023251869 2023251872 2023251874 2023251876 2023251879 2023251881 2023251902 2023283187 2023283192 2023283194 2023283196 2023283200 2023283208 2023283216 2024305462 2024305463 2024305465 2024305466 2024305467 2024308137 2024308138 2024308139 2024308141 2024308143 2024308144 2024308146 2024308148 2024308149 2024308150 2024308159 2024308160 2024308162 2025501277 2025501292 2025501297 2025501301;
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abiomed, Inc.
Reason for Recall:
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Impella CP Smart Assist Set; Product Code: 1000402;

Product Codes/Lot Numbers:

Product Code: 1000402; UDI-DI: 00813502012873; Serial Numbers: 385946 385942 385944 385945 385950 385941 385949 385947 385948 385943 392855 392853 392854 392852 392851 392848 392849 392879 392878 427388 427387 427386 430896 430875 430874 430873 430872 430871 430870 430869 430868 430867 431652 430895 430894 486297 486298 486299 486300; Batch Numbers: 2023251705 2023251861 2023251862 2023251869 2023251872 2023251874 2023251876 2023251879 2023251881 2023251902 2023283187 2023283192 2023283194 2023283196 2023283200 2023283208 2023283216 2024305462 2024305463 2024305465 2024305466 2024305467 2024308137 2024308138 2024308139 2024308141 2024308143 2024308144 2024308146 2024308148 2024308149 2024308150 2024308159 2024308160 2024308162 2025501277 2025501292 2025501297 2025501301;

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2969-2024

Related Recalls

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

Jan 17, 2011 Other Medical Devices Nationwide View Details →

Potential that the "Purge System Blocked" alarm display on the Automated Impella Controller (AIC) could be delayed when using first generation Impella 5.5 (0550-0008*) pumps.

Feb 16, 2026 Other Medical Devices View Details →