Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL. Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Reagent lot numbers: A7125, exp. date 01/9/2019; A7126, exp. date 01/11/2019; A7127, exp. date 01/09/2019, A7128, exp. 01/15/2019; A7129, exp. 01/15/2019; A7130, exp. 01/17/2019; A7131, exp. 01/17/2019, A8003, exp. 01/31/2019; A8004, exp. 01/31/2019; A8005, exp. 02/04/2019; and A8006, exp. 02/04/2019.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Sysmex America, Inc.
- Reason for Recall:
- The fluorescent platelet counts (PLT-F) and immature platelet fraction (IPF) values can be falsely low when using affected lots of Fluorocell PLT.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Sysmex XN-Series Fluorocell PLT, Catalog #CD994563. Distributed in the U.S.A. by Sysmex America, Inc., Lincolnshire, IL. Fluorocell PLT is used to stain platelets in a diluted blood sample for the counting of platelets with the Sysmex XN-Series automated hematology analyzers.
Product Codes/Lot Numbers:
Reagent lot numbers: A7125, exp. date 01/9/2019; A7126, exp. date 01/11/2019; A7127, exp. date 01/09/2019, A7128, exp. 01/15/2019; A7129, exp. 01/15/2019; A7130, exp. 01/17/2019; A7131, exp. 01/17/2019, A8003, exp. 01/31/2019; A8004, exp. 01/31/2019; A8005, exp. 02/04/2019; and A8006, exp. 02/04/2019.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2953-2018
Related Recalls
Sysmex was made aware of a report of false results due to carry-over caused by the sporadic omission of the probe wash step during antibody pipetting on the PS-10 Sample Preparation System.
Insufficient amount of antibody without an error message or alarm
Sysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10 analyzers.
Sysmex America
When excessive amounts of urine spills on the rack transport track and reaches the electrical components of the CV-11 conveyor, there is a potential for smoke and/or sparks which may present a combustion hazard.