Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.
Class I - DangerousWhat Should You Do?
- Check if you have this product: In USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Tecan US, Inc.
- Reason for Recall:
- Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.
Product Codes/Lot Numbers:
In USA RE62119 Lot # LCL129 exp. date: 2017-11-30, RE62119 Lot # LCL133 exp. date: 2019-03-31, and RE62111 Lot # LCL130 exp. date: 2018-03-31. OUS Lot 131, Lot 132
Distribution:
Distributed in: US, GA, IA, LA, MA, MN, OR, PA, SC, UT
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2923-2018
Related Recalls
Gas springs not always replaced within the interval specified.
Freedom EV0-2 with any version of EVOware and EVOware Sample Tracking Add-On V1.0 or higher
Tecan US
There is a software issue that can potentially impact sample identification and traceability under specific conditions when using EVOware Add On Software: Sample Tracking -Multi Instrument V1.0 or higher, Part Number 30019971.