DigitalDiagnost Classic, Dual-Detector (Stitching Patient Support) 712052
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 422102/SN0902020 435210/SN0902204 460471/SN1002019 433651/SN0902114 427610/SN0902108 470160/SN1102003 447013/SN1002008 445778/SN1002006 420208/SN0902019
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Medical Systems Gmbh, DMC
- Reason for Recall:
- The hook does not securely hold the footplate in the vertical position. Therefore, the footplate can fall or swing down and injury operator or patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
DigitalDiagnost Classic, Dual-Detector (Stitching Patient Support) 712052
Product Codes/Lot Numbers:
Serial Numbers: 422102/SN0902020 435210/SN0902204 460471/SN1002019 433651/SN0902114 427610/SN0902108 470160/SN1102003 447013/SN1002008 445778/SN1002006 420208/SN0902019
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2857-2018
Related Recalls
CombiDiagnost R90 Software Version R1.0 and R1.1
Philips Medical Systems Gmbh, DMC
Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
CombiDiagnost GCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Philips Medical Systems Gmbh, DMC
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.
CombiDiagnost PCF, Software v. 1.0.5, Product Code 709030 - Product Usage: Xray system, diagnostic, fluoroscopic, general purpose.
Philips Medical Systems Gmbh, DMC
When using the Table Up/Down button, the system may experience Error 80, which locks the geometry in that specific state, requiring the intervention of a service Engineer. Additionally, the thermos switch, which handles power down of the unit in case of transformer overheating, was installed incorrectly at production.