3D V, model. no. 7042042 7042059 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 7042042 : S/N 1002 to 1787 7042059: S/N 1002 to 1921
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc
- Reason for Recall:
- In rare cases of insufficient maintenance or high clinical workload, the first rope of the ceiling stand, which is designed to take the load, could break without triggering the safety lock and this could lead to overloading the safety rope which is not designed for continuous load and ongoing movement under load. This may lead to mechanical fatigue and cause the arm to drop down during patient positioning. Patients and users may be seriously injured.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
3D V, model. no. 7042042 7042059 3070021 Product Usage: The ceiling stand is used to hold the X-ray tube and / or the detector and is moveable on a ceiling rail system in three axis. The tube and / or detector can be rotated above a horizontal and a vertical axle. The ceiling stand can be used within a stationary diagnostic X-ray system as well.
Product Codes/Lot Numbers:
7042042 : S/N 1002 to 1787 7042059: S/N 1002 to 1921
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2838-2018
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.