Sealapex Xpress, Part No. 33639 The brand name of the device is Sealapex Xpress, a root canal filling resin. Sealapex Xpress is intended to be used during endodontic therapy to fill the root canal of a tooth.
Class I - DangerousWhat Should You Do?
- Check if you have this product: lot # 5-1310
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Ormco/Sybronendo
- Reason for Recall:
- SybronEndo is recalling the Sealapex Xpress because it was labeled with an incorrect expiration date of 2017-11 while the syringe was labeled with the correct expiration date of 2016-11.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Sealapex Xpress, Part No. 33639 The brand name of the device is Sealapex Xpress, a root canal filling resin. Sealapex Xpress is intended to be used during endodontic therapy to fill the root canal of a tooth.
Product Codes/Lot Numbers:
lot # 5-1310
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2836-2016
Related Recalls
The bracket torque was mislabeled on the packaging.
SnapLink, Ref Part No. 438-2190, L6R SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only
Ormco/Sybronendo
The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months.
SnapLink, Ref Part No. 438-2191, L6L SnapLink, Slot .022, TQ--28, Ang +2, Rot +2, Qty 10, Made in Mexico, CE 0086, Rx Only
Ormco/Sybronendo
The action is being taken due to an increase in complaints relating to the lower buccal tube slide doors not staying close. In the event of functional failure, broken/damage buccal tubes it may cause the tooth to rotate, possibly delay treatment and inconvenience the patient and clinician. If not noticed it may extend treatment by approximately 1 -3 months.