BVI CustomEyes Procedure Packs containing Sol-M Luer Lock and Luer Slip 1ml syringes, Part Number: 58000757;
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part Number: 58000757; UDI-DI: 30886158020265; Lot/Batch Number: 6069002, 6071184
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beaver Visitec International, Inc.
- Reason for Recall:
- Due to a voluntary recall of low dead space syringes, and luer slip tip syringes manufactured by Sol-Millennium Medical (Sol-M), Inc. These syringes manufactured by Sol-M are present in some of BVI CustomEyes Procedure Packs.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
BVI CustomEyes Procedure Packs containing Sol-M Luer Lock and Luer Slip 1ml syringes, Part Number: 58000757;
Product Codes/Lot Numbers:
Part Number: 58000757; UDI-DI: 30886158020265; Lot/Batch Number: 6069002, 6071184
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2821-2024
Related Recalls
BVI Leos Laser and Endoscopy System. Model Number: OME6000U.
Beaver-Visitec International
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
BVI CustomEyes Procedure Packs containing Sol-M Luer Lock and Luer Slip 1ml syringes, Part Number: 58001113;
Beaver Visitec International
Due to a voluntary recall of low dead space syringes, and luer slip tip syringes manufactured by Sol-Millennium Medical (Sol-M), Inc. These syringes manufactured by Sol-M are present in some of BVI CustomEyes Procedure Packs.
BVI CustomEyes Procedure Packs containing Sol-M Luer Lock and Luer Slip 1ml syringes, Part Number: 58001737;
Beaver Visitec International
Due to a voluntary recall of low dead space syringes, and luer slip tip syringes manufactured by Sol-Millennium Medical (Sol-M), Inc. These syringes manufactured by Sol-M are present in some of BVI CustomEyes Procedure Packs.