Adaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in traumatology and orthopedic surgery of the skeleton, i.e. drilling, reaming or cutting bone. The Adaptors (05.001.024 and 05.001.108) can be used with one of the electrical consoles belonging to the Electric Pen Drive (EPD) to operate the Small Battery Drive and Small Battery Drive II.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model/Catalog: Adapter for Small Battery Drive and Small Battery Drive II for Electric Pen Drive 05.001.024; Lot/Set/Serial: All Manufactured Lots/Serial Numbers.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
The Anspach Effort, Inc.
Reason for Recall:
There was a potential for the Adaptor and Light Adaptor for Small Battery Drive and Small Battery Drive II to generate excessive internal pressure that may cause the products to burst.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Adaptor for Small Battery Drive and Small Battery Drive II. Product Usage: The Light Adaptor for Small Battery Drive (SBD) /Small Battery Drive II (SBD II) is designed for use in traumatology and orthopedic surgery of the skeleton, i.e. drilling, reaming or cutting bone. The Adaptors (05.001.024 and 05.001.108) can be used with one of the electrical consoles belonging to the Electric Pen Drive (EPD) to operate the Small Battery Drive and Small Battery Drive II.

Product Codes/Lot Numbers:

Model/Catalog: Adapter for Small Battery Drive and Small Battery Drive II for Electric Pen Drive 05.001.024; Lot/Set/Serial: All Manufactured Lots/Serial Numbers.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2784-2016

Related Recalls

Class I - Dangerous

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Dec 7, 2022 Other Medical Devices Nationwide View Details →

Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury

Dec 7, 2022 Other Medical Devices Nationwide View Details →