Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog numbers: ART 700 PEDL, ART 700 TABL, ART 700 OCS, which utilize Software Version SW 005.006_SH. Affected serial numbers: 20329, 20570, 101148, 20392, 20425, 20427, 30046, 100945, 100186, 21255, 30086, 20025, 101153, 101159, 20714, 100873, 20437, 100033, 100249, 20333, 100511, 101206, 100002, 100890, 101208, 20323, 20914, 20342, 10172, 10074, 20195, 20196, 20850, 20929, 101106, 30020, 30024, 30033, 20252, 10087, 20851, 100009, 100864, 100335, 10095, 20302, 20306, 100013, 20918, 100141, 20158, 100784, 20735, 20422, 101149, 20842, 20018, 101114, 100056, 100075, 20434, 21024, 20311, 20763, 10055, 20786, 20701, 20729, 20040, 100092, 20501
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Bayer Healthcare
- Reason for Recall:
- Software Version SW 005.006_SH, resulted in the removal of purge enforcement from traditional New-Case, Power Up and Syringe Change use cases while the injector head is in the upright position.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Angiographic Injector. Medrad Mark 7 Arterion Injection System, which utilizes Software Version SW 005.006_SH; to be used specifically for the purposes of injecting contrast medium and common flushing solutions into humans for angiographic studies.
Product Codes/Lot Numbers:
Catalog numbers: ART 700 PEDL, ART 700 TABL, ART 700 OCS, which utilize Software Version SW 005.006_SH. Affected serial numbers: 20329, 20570, 101148, 20392, 20425, 20427, 30046, 100945, 100186, 21255, 30086, 20025, 101153, 101159, 20714, 100873, 20437, 100033, 100249, 20333, 100511, 101206, 100002, 100890, 101208, 20323, 20914, 20342, 10172, 10074, 20195, 20196, 20850, 20929, 101106, 30020, 30024, 30033, 20252, 10087, 20851, 100009, 100864, 100335, 10095, 20302, 20306, 100013, 20918, 100141, 20158, 100784, 20735, 20422, 101149, 20842, 20018, 101114, 100056, 100075, 20434, 21024, 20311, 20763, 10055, 20786, 20701, 20729, 20040, 100092, 20501
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2775-2015
Related Recalls
Bayer has determined that all current Source Administration Sets (SAS) used with the Medrad Intego Pet Infusion System may produce particulates in radiopharmaceutical (RP) vials when the needle is inserted.
PET Infusion Pump
Bayer Healthcare
The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.
CertegraWorkstation when used in conjunction with the Medrad Stellant CT Injection System may exhibit thermal events, such as: smoke and sparks of fire.