Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial numbers between xxx20000000 and xxx23284925, or less than xxx06335297 Updated serial numbers on 10/12/2019: Aviva II: Between XXX20000000 and XXX23284925 Updated 10/18/2019: Aviva II from XXX05000001 to XXX06335297 OR from XXX20000000 to XXX23284925
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Roche Diabetes Care, Inc.
- Reason for Recall:
- Meters may contain a damaged C29 capacitor resulting in issues with power allegation (short battery life or meter not turning on).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Accu-Chek Aviva II Meter - Product Usage: a. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. b. The Accu-Chek Aviva Plus Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The Accu-Chek Aviva Plus Blood Glucose Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly). c. The Accu-Chek Aviva Plus Test Strips are for use with the Accu-Chek Aviva Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, forearm, upper arm, or palm.
Product Codes/Lot Numbers:
Serial numbers between xxx20000000 and xxx23284925, or less than xxx06335297 Updated serial numbers on 10/12/2019: Aviva II: Between XXX20000000 and XXX23284925 Updated 10/18/2019: Aviva II from XXX05000001 to XXX06335297 OR from XXX20000000 to XXX23284925
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2765-2019
Related Recalls
The reason for recall is the meters may show an incorrect measurement unit (mmol/L rather than mg/dL) and may result in the wrong unit of measure. Which may subsequently cause an error in interpreting a result as approximately 18 times lower that the actual blood glucose. Potential immediate health consequences of the glucose meter providing a reading in the wrong units of measurement (UOM) may result in inappropriate rescue therapy for presumed hypoglycemia including urgent and repeat administration of carbohydrates. This in turn could lead to frequent episodes of hyperglycemia which is unrecognized and undertreated.Long range consequences of continued use of the affected meters could include mismanagement of the patient s diabetes for longer periods of time potentially resulting in a major health event (in general persistent and/or severe hyperglycemia) that may lead them to seek medical attention, and depending on the duration of hyperglycemia, could contribute to the development of microvascular (i.e., retinopathy, neuropathy, nephropathy) and/or macrovascular (i.e., myocardial infarction, ischemic stroke) complications of diabetes and could lead to fetal damage (abnormal fetal growth and altered organ development and maturation, fetal hypoxemia, spontaneous abortion, and congenital anomalies) in pregnant women. This issue was identified by three initial consumer complaints, with reports that the meters were presenting the incorrect unit of measure.
RocheDiabetes Care Platform Software version 2.5.2 material number 08888353001. This is a web application software hosted on a server. There is no physical product.
Roche Diabetes Care
Potential for patient data mismatch when using browser "back" button to navigate between patients when using the diabetes management software.
Accu-Chek Performa Test Strip, Blood Glucose Monitoring System, Model number 07299702001
Roche Diabetes Care
Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.