DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software versions 1.17.0 or below, UDI-DI 15099590732103.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter, Inc.
- Reason for Recall:
- When a DxI 9000 Access Immunoassay Analyzer is connected to a Laboratory Automation System (LAS) and is running with the system software versions 1.17.0 and below, the analyzer does not send a message to the LAS when the sample wheel is full, and the LAS continues sending samples despite the limited capacity to process new samples. Therefore, sample tests may be aliquoted but not processed when the number of tests associated with onboard patient samples exceeds the throughput limit. If this issue occurs test results are flagged with a sample wheel timeout (SWT) error flag, and there may be a delay in reporting test results which could lead to a delay in patient treatment.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
DxI 9000 Access Immunoassay Analyzer, Catalog #C11137.
Product Codes/Lot Numbers:
Software versions 1.17.0 or below, UDI-DI 15099590732103.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2735-2024
Related Recalls
Access Total T4 Calibrator, Catalog No. 33805
Beckman Coulter
Beckman Coulter has identified that Access Total T4 Calibrator lots (PN 33805) exhibit a negative bias on DxI 600/800 instruments, with slopes outside product specifications. A false low patient result could cause a patient to undergo unnecessary additional diagnostic testing. A patient may be treated for a disorder not present or receive delayed treatment for hyperthyroidism. For a pregnant patient being monitored for thyroid hormone supplementation, incorrect treatment or dose may be used. A delayed patient results could cause repeat blood draw with associated complications.
MicroScan Neg MIC 3J REF C54814
Beckman Coulter
Due to a manufacturing issue - missing a drug in well 4/4 but contain double the amount of antibiotics
Due to specific lots of diluents contributing to elevated platelet Daily Checks background counts cycle with use with analyzers.