Plum A+3 Infusion Pump with Hospira MedNet Software. List Number 20678. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.
Class I - DangerousWhat Should You Do?
- Check if you have this product: List number 20678. Serial numbers: 15541001, 15541011, 15541021, 15541031, 15541041, 15541051, 15541061, 15541071, 15541081, 15541091, 15541101, 15541111, 15541121, 15541131, 15541141, 15541151, 15541161, 15541171, 15541181, 15541191, 15541201, 15541211, 15541231, 15541241, 15541261, 15541271, 15541281, 15541291, 15541301, 15541311, 15541321, 15541331, 15541341, 15541351, 15541361, 15541371, 15541381, 15541391, 15541401, 15541411, 15541421, 15541431, 15541441, 15541451, 15541461, 15541471, 15541481, 15541491, 15541501, 15541511, 15541521, 15541531, 15541541, 15541551, 15541561, 15541571, 15541581, 15541591, 15541601, 15541611, 15541621, 15541631, 15541641, 15541651, 15541661, 15541671, 15541681, 15541691, 15541701, 15541711, 15541721, 15541731, 15541741, 15541751, 15541761, 15541771, 15541781, 15541791, 15541801, 15541811, 15541821, 15541831, 15541841, 15541851, 15541861, 15541871, 15541881, 15541891, 15541901, 15541911, 15541921, 15541931, 15541941, 15541951, 15541961, 15541971, 15541981, 15541991, 15542001, 15542011, 15542021, 15542031, 15542041, 15542051, 15542061, 15542071, 15542081, 15542091, 15542101, 15542111, 15542121, 15542131, 15542141, 15542151, 15542161, 15542171, 15542181, 15542191, 15542201, 15542211, 15542221, 15542231, 15542241, 15542251, 15542261, 15542271, 15542281, 15542291, 15542301, 15542311, 15542321, 15542331, 15542341, 15542351, 15542361, 15542371, 15542381, 15542391, 15542401, 15542411, 15542431, 15542441, 15542451, 15542461, 15542471, 15542481, 15542491, 15542501, 15542511, 15542521, 15542531, 15542541, 15542551, 15542561, 15542571, 15542581, 15542591, 15542601, 15542611, 15542621, 15542631, 15542641, 15542651, 15542661, 15542671, 15542681, 15542691, 15542701, 15542711, 15542721, 15542731, 15542741, 15542751, 15542761, 15542771, 15542781, 15542791, 15542801, 15542811, 15542821, 15542831, 15542841, 15542851, 15542861, 15542871, 15542881, 15542891, 15542911, 15542921, 15542931, 15542941, 15542951, 15542961, 15542971, 15542981, 15542991, 15543001, 15543011, 15543021, 15543031, 15543041, 15543051, 15543061, 15543071, 15543081, 15543101, 15543111, 15543121, 15543131, 15543141, 15543151, 15543161, 15543181, 15543191, 15543211, 15543221, 15543231, 15543241, 15543261, 15543271, 15543281, 15543291, 15543301, 15543311, 15543321, 15543331, 15543341, 15543351, 15543361, 15543371, 15543391, 15543401, 15543411, 15543421, 15543431, 15543441, 15543451, 15543461, 15543471, 15543481, 15543491, 15543501, 15543511, 15543521, 15543531, 15543541, 15543551, 15543571, 15543581, 15543591, 15543601, 15543611, 15543621, 15543631, 15543641, 15543651, 15543661, 15543671, 15543681, 15543691, 15543711, 15543721, 15543731, 15543741, 15543751, 15543761, 15543771, 15543781, 15543791, 15543801, 15543811, 15543821, 15543831, 15543841, 15543851, 15543861, 15543871, 15543881, 15543891, 15543901, 15543911, 15543921, 15543931, 15543941, 15543951, 15543961, 15543971, 15543981, 15543991, 15544001, 15544011, 15544021, 15544031, 15544041, 15544051, 15544061, 15544071, 15544081, 15544091, 15544101, 15544111, 15544121, 15544131, 15544141, 15544151, 15544161, 15544191, 15544201, 15544211, 15544221, 15544231, 15544241, 15544251, 15544261, 15544271, 15544281, 15544291, 15544301, 15544311, 15544321, 15544331, 15544341, 15544351, 15544371, 15544381, 15544401, 15544411, 15544421, 15544431, 15544441, 15544451, 15544461, 15544471, 15544481, 15544501, 15544511, 15544521, 15544531, 15544541, 15544551, 15544561, 15544571, 15544581, 15544591, 15544601, 15544611, 15544621, 15544641, 15544661, 15544671, 15544681, 15544691, 15544701, 15544711, 15544721, 15544731, 15544741, 15544751, 15544761, 15544771, 15544781, 15544791, 15544801, 15544811, 15544821, 15544831, 15544841, 15544851, 15544871, 15544891, 15544901, 15544911, 15544921, 15544931, 15544941, 15544951, 15544961, 15544971, 15544981, 15544991, 15545001, 15545021, 15545031, 15545041, 15545051, 15545061, 15545071, 15545091, 15545101, 15545111, 15545121, 15545131, 15545141, 15545151, 15545161, 15545171, 15545181, 15545191, 15545201, 15545211, 15545231, 15545241, 15545251, 15545261, 15545271, 15545281, 15545291, 15545301, 15545311, 15545321, 15545331, 15545341, 15545351, 15545361, 15545371, 15545381, 15545391, 15545401, 15545411, 15545421, 15545431, 15545441, 15545451, 15545461, 15545471, 15545481, 15545491, 15545501, 15545511, 15545521, 15545531, 15545541, 15545551, 15545561, 15545571, 15545581, 15545591, 15545601, 15545611, 15545621, 15545631, 15545641, 15545651, 15545661, 15545671, 15545691, 15545701, 15545711, 15545721, 15545741, 15545751, 15545761, 15545771, 15545781, 15545791, 15545801, 15545811, 15545821, 15545831, 15545841, 15545851, 15545861, 15545871, 15545881, 15545891, 15545911, 15545921, 15545931, 15545941, 15545951, 15545961, 15545971, 15545981, 15545991, 15546001, 15546011, 15546021, 15546031, 15546041, 15546051, 15546061, 15546071, 15546081, 15546091, 15546101, 15546121, 15546131, 15546141, 15546151, 15546161, 15546171, 15546181, 15546191, 15546201, 15546211, 15546221, 15546231, 15546241, 15546251, 15546281, 15546291, 15546301, 15546311, 15546321, 15546331, 15546341, 15546351, 15546361, 15546371, 15546381, 15546391, 15546401, 15546411, 15546421, 15546431, 15546441, 15546451, 15546461, 15546481, 15546501, 15546511, 15546521, 15546531, 15546541, 15546551, 15546561, 15546571, 15546581, 15546591, 15546601, 15546611, 15546621, 15546631, 15546641, 15546651, 15546661, 15546671, 15546681, 15546691, 15546701, 15546711, 15546721, 15546731, 15546741, 15546751, 15546761, 15546771, 15546791, 15546801, 15546811, 15546821, 15546831, 15546841, 15546851, 15546861, 15546871, 15546881, 15546891, 15546901, 15546911, 15546921, 15546931, 15546941, 15546951, 15546961, 15546971, 15546981, 15546991, 15547001, 15547011, 15547021, 15547031, 15547041, 15547051, 15547061, 15547071, 15547081, 15547091, 15547111, 15547121, 15547131, 15547141, 15547151, 15547161, 15547171, 15547181, 15547191, 15547201, 15547211, 15547221, 15547231, 15547241, 15547251, 15547261, 15547271, 15547281, 15547291, 15547301, 15547321, 15547331, 15547341, 15547351, 15547361, 15547371, 15547381, 15547391, 15547401, 15547411, 15547421, 15547431, 15547441, 15547451, 15547461, 15547471, 15547481, 15547491, 15547501, 15547511, 15547521, 15547531, 15547541, 15547551, 15547561, 15547571, 15547581, 15547591, 15547601, 15547611, 15547621, 15547631, 15547641, 15547651, 15547661, 15547671, 15547681, 15547691, 15547701, 15547711, 15547721, 15547731, 15547741, 15547751, 15547761, 15547771, 15547781, 15547791, 15547801, 15547811, 15547821, 15547831, 15547851, 15547861, 15547871, 15547881, 15547891, 15547901, 15547911, 15547921, 15547931, 15547941, 15547951, 15547961, 15547971, 15547981, 15548001, 15548011, 15548031, 15548041, 15548051, 15548061, 15548071, 15548081, 15548091, 15548101, 15548111, 15548121, 15548131, 15548141, 15548151, 15548161, 15548171, 15548181, 15548191, 15548201, 15548211, 15548221, 15548231, 15548241, 15548251, 15548261, 15548271, 15548281, 15548291, 15548301, 15548311, 15548321, 15548331, 15548341, 15548351, 15548361, 15548371, 15548381, 15548391, 15548401, 15548411, 15548421, 15548431, 15548441, 15548451, 15548461, 15548471, 15548481, 15548491, 15548501, 15548511, 15548521, 15548531, 15548541, 15548551, 15548561, 15548571, 15548581, 15548591, 15548601, 15548611, 15548621, 15548631, 15548641, 15548651, 15548661, 15548671, 15548681, 15548691, 15548701, 15548711, 15548731, 15548741, 15548751, 15548761, 15548771, 15548781, 15548791, 15548801, 15548811, 15548821, 15548831, 15548841, 15548851, 15548861, 15548871, 15548881, 15548891, 15548901, 15548911, 15548921, 15548931, 15548941, 15548961, 15548971, 15548981, 15548991, 15549001, 15549011, 15549031, 15549041, 15549051, 15549061, 15549071, 15549081, 15549091, 15549101, 15549111, 15549121, 15549131, 15549141, 15549151, 15549161, 15549171, 15549181, 15549191, 15549201, 15549211, 15549221, 15549231, 15549241, 15549251, 15549261, 15549271, 15549281, 15549291, 15549301, 15549321, 15549331, 15549341, 15549351, 15549361, 15549371, 15549381, 15549391, 15549401, 15549411, 15549421, 15549431, 15549441, 15549451, 15549461, 15549471, 15549481, 15549491, 15549501, 15549511, 15549521, 15549531, 15549541, 15549551, 15549561, 15549571, 15549581, 15549591, 15549601, 15549611, 15549621, 15549641, 15549651, 15549661, 15549671, 15549681, 15549691, 15549701, 15549711, 15549721, 15549741, 15549751, 15549761, 15549771, 15549781, 15549791, 15549801, 15549811, 15549821, 15549831, 15549841, 15549851, 15549861, 15549871, 15549881, 15549891, 15549901, 15549911, 15549921, 15549931, 15549941, 15549951, 15549961, 15549971, 15549981, 15550001, 15550011, 15550021, 15550031, 15550041, 15550051, 15550061, 15550071, 15550081, 15550091, 15550101, 15550111, 15550121, 15550131, 15550141, 15550151, 15550161, 15550171, 15550181, 15550191, 15550201, 15550211, 15550221, 15550231, 15550241, 15550251, 15550261, 15550271, 15550281, 15550291, 15550301, 15550311, 15550321, 15550331, 15550341, 15550351, 15550361, 15550371, 15550381, 15550391, 15550401, 15550411, 15550421, 15550431, 15550441, 15550451, 15550461, 15550481, 15550491, 15550511, 15550521, 15550531, 15550541, 15550551, 15550561, 15550571, 15550581, 15550591, 15550601, 15550611, 15550621, 15550631, 15550641, 15550651, 15550661, 15550671, 15550681, 15550691, 15550701, 15550711, 15550721, 15550731, 15550741, 15550751, 15550761, 15550771, 15550781, 15550791, 15550801, 15550811, 15550821, 15550831, 15550841, 15550851, 15550861, 15550871, 15550881, 15550901, 15550911, 15550921, 15550931, 15550941, 15550951, 15550961, 15550971, 15550981, 15550991, 15861001, 15861011, 15861021, 15861031, 15861041, 15861051, 15861061, 15861071, 15861081, 15861091, 15861101, 15861111, 15861121, 15861131, 15861141, 15861151, 15861161, 15861171, 15861181, 15861191, 15861201, 15861211, 15861221, 15861231, 15861241, 15861251, 15861261, 15861271, 15861281, 15861291, 15861301, 15861311, 15861321, 15861331, 15861341, 15861351, 15861361, 15861371, 15861381, 15861391, 15861401, 15861411, 15861421, 15861431, 15861441, 15861451, 15861461, 15861471, 15861481, 15861491, 15861501, 15861511, 15861521, 15861531, 15861541, 15861551, 15861561, 15861571, 15861581, 15861591, 15861601, 15861621, 15861631, 15861641, 15861651, 15861661, 15861681, 15861691, 15861701, 15861711, 15861721, 15861731, 15861741, 15861751, 15861761, 15861771, 15861781, 15861791, 15861801, 15861821, 15861831, 15861841, 15861851, 15861861, 15861871, 15861881, 15861891, 15861901, 15861921, 15861941, 15861951, 15861961, 15861981, 15861991, 15862001, 15862011, 15862021, 15862031, 15862041, 15862051, 15862061, 15862071, 15862081, 15862101, 15862111, 15862121, 15862131, 15862141, 15862151, 15862161, 15862171, 15862181, 15862191, 15862201, 15862211, 15862221, 15862231, 15862251, 15862261, 15862271, 15862281, 15862291, 15862301, 15862311, 15862321, 15862331, 15862341, 15862351, 15862361, 15862371, 15862451, 15862461, 15862471, 15862481, 15862491, 15862501, 15862511, 15862521, 15862531, 15862541, 15862551, 15862561, 15862571, 15862581, 15862591, 15862601, 15862611, 15862621, 15862631, 15862641, 15862651, 15862661, 15862671, 15862681, 15862691, 15862701, 15862711, 15862721, 15862731, 15862741, 15862751, 15862781, 15862791, 15862811, 15862821, 15862831, 15862841, 15862851, 15862861, 15862871, 15862881, 15862891, 15862911, 15862921, 15862931, 15862941, 15862951, 15862961, 15862971, 15862981, 15862991, 15863001, 15863021, 15863031, 15863041, 15863051, 15863061, 15863071, 15863081, 15863091, 15863101, 15863111, 15863121, 15863131, 15863141, 15863151, 15863161, 15863171, 15863181, 15863191, 15863201, 15863211, 15863221, 15863231, 15863241, 15863251, 15863261, 15863271, 15863281, 15863291, 15863301, 15863311, 15863321, 15863331, 15863341, 15863351, 15863361, 15863371, 15863381, 15863391, 15863401, 15863411, 15863421, 15863431, 15863441, 15863451, 15863461, 15863471, 15863481, 15863491, 15863501, 15863511, 15863521, 15863531, 15863541, 15863551, 15863561, 15863571, 15863581, 15863591, 15863601, 15863611, 15863621, 15863631, 15863641, 15863651, 15863661, 15863671, 15863681, 15863691, 15863701, 15863711, 15863721, 15863731, 15863741, 15863751, 15863761, 15863771, 15863781, 15863791, 15863801, 15863811, 15863821, 15863831, 15863841, 15863851, 15863861, 15863871, 15863881, 15863891, 15863901, 15863911, 15863921, 15863931, 15863941, 15863951, 15863961, 15863971, 15863981, 15863991, 15864001, 15864011, 15864021, 15864031, 15864041, 15864051, 15864061, 15864071, 15864081, 15864091, 15864101, 15864111, 15864131, 15864141, 15864151, 15864161, 15864171, 15864181, 15864191, 15864201, 15864211, 15864221, 15864231, 15864241, 15864251, 15864261, 15864271, 15864281, 15864291, 15864311, 15864321, 15864331, 15864351, 15864371, 15864381, 15864391, 15864411, 15864421, 15864431, 15864441, 15864471, 15864481, 15864491, 15864501, 15864511, 15864521, 15864531, 15864541, 15864551, 15864561, 15864571, 15864581, 15864591, 15864601, 15864611, 15864621, 15864631, 15864641, 15864651, 15864661, 15864671, 15864681, 15864691, 15864701, 15864711, 15864721, 15864731, 15864751, 15864761, 15864771, 15864781, 15864791, 15864801, 15864811, 15864821, 15864831, 15864841, 15864851, 15864861, 15864871, 15864881, 15864891, 15864901, 15864911, 15864921, 15864931, 15864941, 15864951, 15864961, 15864971, 15864981, 15864991, 15865001, 15865011, 15865021, 15865031, 15865041, 15865051, 15865061, 15865071, 15865081, 15865101, 15865111, 15865121, 15865131, 15865141, 15865151, 15865161, 15865171, 15865181, 15865191, 15865201, 15865211, 15865221, 15865231, 15865241, 15865251, 15865261, 15865271, 15865281, 15865291, 15865301, 15865311, 15865321, 15865331, 15865341, 15865351, 15865361, 15865371, 15865381, 15865391, 15865401, 15865411, 15865421, 15865431, 15865441, 15865451, 15865461, 15865471, 15865481, 15865491, 15865501, 15865511, 15865521, 15865551, 15865561, 15865571, 15865591, 15865601, 15865611, 15865621, 15865631, 15865641, 15865651, 15865661, 15865671, 15865681, 15865691, 15865701, 15865711, 15865721, 15865731, 15865741, 15865751, 15865761, 15865771, 15865781, 15865791, 15865801, 15865811, 15865821, 15865831, 15865841, 15865851, 15865861, 15865871, 15865881, 15865891, 15865901, 15865911, 15865921, 15865931, 15865941, 15865951, 15865961, 15865971, 15865981, 15865991, 15866001, 15866011, 15866021, 15866031, 15866041, 15866051, 15866061, 15866071, 15866081, 15866091, 15866101, 15866111, 15866121, 15866131, 15866141, 15866151, 15866161, 15866171, 15866181, 15866191, 15866201, 15866211, 15866221, 15866231, 15866241, 15866251, 15866261, 15866271, 15866291, 15866301, 15866311, 15866321, 15866331, 15866341, 15866351, 15866361, 15866371, 15866381, 15866391, 15866401, 15866411, 15866421, 15866431, 15866451, 15866461, 15866471, 15866481, 15866491, 15866501, 15866511, 15866521, 15866531, 15866541, 15866551, 15866561, 15866571, 15866581, 15866591, 15866601, 15866611, 15866621, 15866631, 15866641, 15866651, 15866661, 15866671, 15866681, 15866691, 15866701, 15866711, 15866721, 15866731, 15866741, 15866751, 15866761, 15866771, 15866781, 15866791, 15866801, 15866811, 15866821, 15866831, 15866841, 15866851, 15866861, 15866871, 15866881, 15866891, 15866901, 15866911, 15866921, 15866931, 15866941, 15866951, 15866961, 15866971, 15866981, 15866991, 15867001, 15867011, 15867021, 15867031, 15867051, 15867061, 15867071, 15867081, 15867091, 15867101, 15867111, 15867121, 15867131, 15867141, 15867151, 15867171, 15867181, 15867191, 15867201, 15867211, 15867221, 15867231, 15867241, 15867251, 15867261, 15867271, 15867281, 15867291, 15867301, 15867311, 15867321, 15867331, 15867341, 15867351, 15867361, 15867371, 15867381, 15867391, 15867401, 15867411, 15867421, 15867431, 15867441, 15867451, 15867461, 15867471, 15867481, 15867491, 15867501, 15867511, 15867521, 15867531, 15867541, 15867551, 15867561, 15867571, 15867581, 15867591, 15867601, 15867611, 15867621, 15867631, 15867641, 15867651, 15867661, 15867671, 15867681, 15867691, 15867701, 15867711, 15867721, 15867731, 15867741, 15867751, 15867761, 15867771, 15867781, 15867791, 15867801, 15867811, 15867821, 15867831, 15867841, 15867851, 15867861, 15867871, 15867881, 15867891, 15867901, 15867911, 15867921, 15867931, 15867941, 15867951, 15867961, 15867971, 15867981, 15867991, 15868001, 15868011, 15868021, 15868031, 15868051, 15868061, 15868071, 15868081, 15868091, 15868101, 15868111, 15868121, 15868131, 15868141, 15868151, 15868161, 15868171, 15868181, 15868191, 15868201, 15868211, 15868221, 15868231, 15868241, 15868251, 15868261, 15868271, 15868281, 15868301, 15868311, 15868321, 15868331, 15868341, 15868351, 15868361, 15868371, 15868381, 15868391, 15868401, 15868411, 15868421, 15868431, 15868441, 15868451, 15868461, 15868471, 15868481, 15868491, 15868501, 15868511, 15868531, 15868541, 15868551, 15868561, 15868571, 15868581, 15868591, 15868601, 15868611, 15868621, 15868631, 15868641, 15868651, 15868661, 15868671, 15868681, 15868691, 15868701, 15868711, 15868721, 15868731, 15868741, 15868751, 15868761, 15868771, 15868781, 15868791, 15868801, 15868811, 15868821, 15868831, 15868841, 15868851, 15868861, 15868871, 15868881, 15868891, 15868901, 15868911, 15868921, 15868931, 15868941, 15868951, 15868961, 15868971, 15868981, 15869001, 15869011, 15869021, 15869031, 15869041, 15869051, 15869061, 15869071, 15869081, 15869091, 15869101, 15869111, 15869121, 15869131, 15869141, 15869151, 15869161, 15869181, 15869191, 15869201, 15869211, 15869221, 15869231, 15869241, 15869251, 15869261, 15869271, 15869281, 15869291, 15869301, 15869311, 15869321, 15869331, 15869341, 15869351, 15869361, 15869371, 15869381, 15869401, 15869411, 15869421, 15869441, 15869451, 15869461, 15869471, 15869481, 15869491, 15869501, 15869511, 15869521, 15869531, 15869541, 15869551, 15869561, 15869571, 15869581, 15869591, 15869601, 15869611, 15869621, 15869631, 15869641, 15869651, 15869661, 15869671, 15869681, 15869691, 15869701, 15869711, 15869721, 15869731, 15869741, 15869751, 15869761, 15869771, 15869781, 15869791, 15869811, 15869821, 15869831, 15869841, 15869851, 15869861, 15869871, 15869881, 15869891, 15869901, 15869911, 15869921, 15869931, 15869941, 15869951, 15869961, 15869971, 15869981, 15869991, 15870001, 15870011, 15870021, 15870031, 15870041, 15870051, 15870061, 15870071, 15870081, 15870091, 15870101, 15870111, 15870121, 15870131, 15870141, 15870151, 15870161, 15870171, 15870181, 15870191, 15870201, 15870211, 15870221, 15870231, 15870241, 15870251, 15870261, 15870271, 15870281, 15870291, 15870301, 15870311, 15870321, 15870331, 15870341, 15870351, 15870361, 15870371, 15870381, 15870391, 15870401, 15870411, 15870421, 15870431, 15870441, 15870461, 15870471, 15870481, 15870491, 15870501, 15870511, 15870521, 15870531, 15870541, 15870551, 15870561, 15870571, 15870581, 15870591, 15870601, 15870611, 15870631, 15870651, 15870661, 15870671, 15870681, 15870691, 15870701, 15870711, 15870721, 15870731, 15870741, 15870751, 15870771, 15870781, 15870791, 15870801, 15870811, 15870821, 15870831, 15870841, 15870851, 15870861, 15870881, 15870891, 15870901, 15870911, 15870931, 15870951, 15870961, 15870971, 15870981, 15870991, 15871001, 15871011, 15871021, 15871031, 15871041, 15871061, 15871071, 15871081, 15871091, 15871101, 15871111, 15871121, 15871131, 15871141, 15871151, 15871161, 15871171, 15871181, 15871191, 15871201, 15871211, 15871221, 15871231, 15871241, 15871251, 15871261, 15871271, 15871281, 15871291, 15871301, 15871311, 15871321, 15871331, 15871341, 15871351, 15871361, 15871371, 15871381, 15871391, 15871401, 15871411, 15871431, 15871441, 15871451, 15871461, 15871481, 15871491, 15871501, 15871511, 15871521, 15871531, 15871541, 15871551, 15871561, 15871571, 15871581, 15871591, 15871601, 15871611, 15871621, 15871631, 15871641, 15871651, 15871661, 15871671, 15871681, 15871691, 15871701, 15871711, 15871721, 15871731, 15871741, 15871751, 15871761, 15871771, 15871781, 15871791, 15871801, 15871811, 15871821, 15871831, 15871841, 15871851, 15871861, 15871871, 15871891, 15871901, 15871911, 15871921, 15871931, 15871941, 15871951, 15871961, 15871971, 15871981, 15871991, 15872001, 15872011, 15872021, 15872031, 15872041, 15872051, 15872061, 15872071, 15872081, 15872091, 15872101, 15872111, 15872121, 15872131, 15872141, 15872151, 15872161, 15872171, 15872181, 15872191, 15872201, 15872211, 15872221, 15872231, 15872241, 15872251, 15872261, 15872271, 15872281, 15872291, 15872301, 15872311, 15872321, 15872331, 15872341, 15872351, 15872361, 15872371, 15872381, 15872391, 15872401, 15872411, 15872421, 15872431, 15872441, 15872451, 15872471, 15872481, 15872491, 15872501, 15872511, 15872521, 15872531, 15872541, 15872551, 15872561, 15872571, 15872581, 15872591, 15872601, 15872611, 15872621, 15872631, 15872641, 15872651, 15872661, 15872671, 15872681, 15872691, 15872701, 15872711, 15872721, 15872731, 15872741, 15872751, 15872761, 15872771, 15872781, 15872791, 15872801, 15872811, 15872821, 15872841, 15872851, 15872861, 15872871, 15872881, 15872891, 15872901, 15872911, 15872921, 15872931, 15872941, 15872951, 15872961, 15872971, 15872981, 15872991, 15873001, 15873011, 15873021, 15873031, 15873041, 15873051, 15873061, 15873071, 15873081, 15873091, 15873101, 15873121, 15873131, 15873141, 15873151, 15873161, 15873171, 15873181, 15873191, 15873201, 15873211, 15873221, 15873231, 15873241, 15873261, 15873271, 15873281, 15873291, 15873301, 15873311, 15873321, 15873331, 15873341, 15873351, 15873361, 15873371, 15873381, 15873391, 15873401, 15873411, 15873421, 15873431, 15873441, 15873451, 15873461, 15873471, 15873481, 15873491, 15873501, 15873511, 15873521, 15873531, 15873541, 15873551, 15873571, 15873581, 15873591, 15873601, 15873611, 15873621, 15873631, 15873641, 15873651, 15873661, 15873671, 15873681, 15873691, 15873701, 15873711, 15873721, 15873731, 15873741, 15873751, 15873771, 15873781, 15873791, 15873801, 15873811, 15873821, 15873831, 15873841, 15873851, 15873861, 15873871, 15873881, 15873891, 15873901, 15873921, 15873931, 15873941, 15873951, 15873961, 15873971, 15873981, 15873991, 15874001, 15874011, 15874021, 15874031, 15874041, 15874051, 15874061, 15874071, 15874081, 15874091, 15874101, 15874111, 15874121, 15874131, 15874141, 15874151, 15874161, 15874171, 15874181, 15874191, 15874201, 15874211, 15874221, 15874231, 15874241, 15874251, 15874271, 15874281, 15874291, 15874301, 15874311, 15874321, 15874331, 15874341, 15874351, 15874361, 15874371, 15874381, 15874391, 15874401, 15874421, 15874431, 15874441, 15874451, 15874461, 15874471, 15874481, 15874491, 15874501, 15874511, 15874521, 15874531, 15874541, 15874551, 15874561, 15874571, 15874581, 15874591, 15874601, 15874611, 15874621, 15874631, 15874641, 15874651, 15874661, 15874671, 15874681, 15874691, 15874701, 15874711, 15874721, 15874731, 15874741, 15874751, 15874761, 15874771, 15874781, 15874791, 15874801, 15874811, 15874821, 15874831, 15874841, 15874851, 15874861, 15874871, 15874881, 15874891, 15874901, 15874911, 15874921, 15874931, 15874941, 15874951, 15874961, 15874971, 15874981, 15874991, 15875001, 15875011, 15875021, 15875031, 15875041, 15875051, 15875061, 15875071, 15875081, 15875091, 15875101, 15875111, 15875121, 15875131, 15875141, 15875151, 15875161, 15875171, 15875181, 15875191, 15875211, 15875221, 15875231, 15875241, 15875251, 15875261, 15875271, 15875281, 15875291, 15875301, 15875311, 15875321, 15875331, 15875341, 15875351, 15875361, 15875371, 15875381, 15875391, 15875401, 15875411, 15875421, 15875431, 15875441, 15875451, 15875461, 15875471, 15875481, 15875491, 15875501, 15875511, 15875521, 15875531, 15875541, 15875551, 15875561, 15875581, 15875591, 15875601, 15875611, 15875621, 15875631, 15875641, 15875651, 15875671, 15875681, 15875701, 15875711, 15875721, 15875731, 15875741, 15875751, 15875761, 15875771, 15875781, 15875791, 15875801, 15875811, 15875821, 15875841, 15875851, 15875861, 15875871, 15875881, 15875891, 15876101, 15876111, 15876121, 15876131, 15876141, 15876151, 15876161, 15876171, 15876181, 15876191, 15876201, 15876211, 15876221, 15876231, 15876241, 15876251, 15876261, 15876271, 15876281, 15876291, 15876311, 15876321, 15876331, 15876341, 15876351, 15876361, 15876371, 15876381, 15876391, 15876401, 15876411, 15876421, 15876431, 15876441, 15876451, 15876461, 15876471, 15876481, 15876491, 15876501, 15876511, 15876521, 15876531, 15876541, 15876551, 15876561, 15876571, 15876581, 15876591, 15876601, 15876621, 15876631, 15876641, 15876651, 15876661, 15876671, 15876681, 15876691, 15876701, 15876711, 15876721, 15876731, 15876741, 15876751, 15876761, 15876771, 15876781, 15876791, 15876801, 15876811, 15876821, 15876831, 15876841, 15876851, 15876861, 15876871, 15876881, 15876891, 15876901, 15876911, 15876921, 15876931, 15876941, 15876951, 15876961, 15876971, 15876981, 15876991, 15877001, 15877031, 15877041, 15877061, 15877081, 15877091, 15877111, 15877121, 15877131, 15877151, 15877161, 15877171, 15877181, 15877201, 15877211, 15877221, 15877231, 15877241, 15877251, 15877261, 15877271, 15877281, 15877291, 15877301, 15877311, 15877321, 15877331, 15877341, 15877351, 15877361, 15877371, 15877381, 15877391, 15877401, 15877411, 15877421, 15877441, 15877451, 15877461, 15877471, 15877491, 15877501, 15877511, 15877521, 15877531, 15877541, 15877551, 15877571, 15877581, 15877591, 15877601, 15877611, 15877621, 15877631, 15877641, 15877651, 15877661, 15877671, 15877681, 15877691, 15877701, 15877711, 15877721, 15877731, 15877741, 15877751, 15877761, 15877771, 15877781, 15877791, 15877801, 15877811, 15877821, 15877831, 15877841, 15877851, 15877861, 15877871, 15877881, 15877891, 15877901, 15877911, 15877921, 15877931, 15877941, 15877951, 15877961, 15877971, 15877981, 15877991, 15878001, 15878021, 15878031, 15878051, 15878061, 15878071, 15878081, 15878091, 15878101, 15878111, 15878121, 15878131, 15878141, 15878151, 15878161, 15878171, 15878181, 15878191, 15878201, 15878211, 15878221, 15878231, 15878241, 15878251, 15878261, 15878271, 15878281, 15878291, 15878301, 15878311, 15878321, 15878331, 15878341, 15878351, 15878361, 15878381, 15878391, 15878401, 15878411, 15878421, 15878431, 15878441, 15878451, 15878461, 15878471, 15878481, 15878491, 15878501, 15878521, 15878531, 15878541, 15878551, 15878561, 15878571, 15878581, 15878591, 15878601, 15878611, 15878621, 15878631, 15878641, 15878651, 15878661, 15878671, 15878681, 15878691, 15878701, 15878711, 15878721, 15878731, 15878741, 15878751, 15878761, 15878771, 15878781, 15878791, 15878801, 15878811, 15878831, 15878841, 15878851, 15878861, 15878871, 15878881, 15878901, 15878911, 15878921, 15878931, 15878951, 15878961, 15878971, 15878981, 15879001, 15879011, 15879021, 15879031, 15879041, 15879051, 15879061, 15879071, 15879081, 15879091, 15879101, 15879121, 15879131, 15879141, 15879151, 15879161, 15879171, 15879181, 15879191, 15879211, 15879221, 15879231, 15879241, 15879251, 15879261, 15879271, 15879281, 15879291, 15879301, 15879321, 15879331, 15879341, 15879351, 15879361, 15879371, 15879381, 15879401, 15879411, 15879421, 15879431, 15879441, 15879451, 15879461, 15879471, 15879481, 15879491, 15879501, 15879511, 15879521, 15879531, 15879541, 15879551, 15879561, 15879571, 15879581, 15879591, 15879601, 15879611, 15879621, 15879631, 15879651, 15879661, 15879671, 15879681, 15879691, 15879701, 15879711, 15879721, 15879731, 15879741, 15879751, 15879771, 15879781, 15879791, 15879801, 15879811, 15879821, 15879841, 15879851, 15879861, 15879871, 15879881, 15879891, 15879901, 15879911, 15879921, 15879931, 15879941, 15879951, 15879961, 15879971, 15879981, 15879991, 15880001, 15880011, 15880021, 15880031, 15880041, 15880051, 15880061, 15880071, 15880081, 15880091, 15880101, 15880111, 15880121, 15880131, 15880141, 15880151, 15880161, 15880171, 15880181, 15880201, 15880211, 15880221, 15880231, 15880241, 15880251, 15880261, 15880271, 15880281, 15880291, 15880301, 15880311, 15880321, 15880331, 15880341, 15880351, 15880361, 15880371, 15880381, 15880391, 15880401, 15880411, 15880421, 15880431, 15880441, 15880451, 15880471, 15880481, 15880491, 15880511, 15880521, 15880531, 15880541, 15880551, 15880561, 15880571, 15880581, 15880591, 15880601, 15880611, 15880621, 15880631, 15880651, 15880661, 15881121, 15881131, 15881141, 15881151, 15881161, 15881171, 15881181, 15881191, 15881201, 15881211, 15881221, 15881231, 15881241, 15881251, 15881261, 15881271, 15881281, 15881291, 15881301, 15881311, 15881321, 15881331, 15881341, 15881361, 15881371, 15881381, 15881391, 15881401, 15881411, 15881421, 15881431, 15881441, 15881451, 15881461, 15881471, 15881481, 15881491, 15881501, 15881511, 15881521, 15881531, 15881541, 15881561, 15881571, 15881581, 15881591, 15881601, 15881611, 15881621, 15881631, 15881641, 15881651, 15881661, 15881671, 15881681, 15881691, 15881711, 15881721, 15881731, 15881741, 15881751, 15881771, 15881781, 15881791, 15881801, 15881811, 15881821, 15881831, 15881841, 15881851, 15881861, 15881871, 15881881, 15881891, 15881901, 15881921, 15881931, 15881941, 15881951, 15881961, 15881971, 15881981, 15881991, 15882001, 15882011, 15882021, 15882031, 15882041, 15882051, 15882061, 15882071, 15882091, 15882101, 15882111, 15882121, 15882131, 15882141, 15882151, 15882161, 15882171, 15882181, 15882191, 15882201, 15882211, 15882221, 15882231, 15882241, 15882251, 15882261, 15882271, 15882281, 15882291, 15882301, 15882311, 15882321, 15882351, 15882361, 15882371, 15882381, 15882391, 15882401, 15882411, 15882421, 15882431, 15882451, 15882461, 15882471, 15882481, 15882491, 15882501, 15882511, 15882521, 15882531, 15882541, 15882551, 15882561, 15882571, 15882581, 15882591, 15882601, 15882611, 15882621, 15882631, 15882641, 15882651, 15882661, 15882671, 15882691, 15882701, 15882721, 15882731, 15882741, 15882751, 15882761, 15882771, 15882781, 15882791, 15882801, 15882811, 15882821, 15882831, 15882841, 15882851, 15882861, 15882871, 15882891, 15882901, 15882911, 15882921, 15882941, 15882951, 15882961, 15882971, 15882981, 15882991, 15883001, 15883011, 15883021, 15883031, 15883041, 15883051, 15883061, 15883071, 15883081, 15883091, 15883101, 15883111, 15883121, 15883131, 15883141, 15883151, 15883161, 15883171, 15883181, 15883191, 15883201, 15883211, 15883221, 15883231, 15883241, 15883261, 15883271, 15883281, 15883291, 15883301, 15883311, 15883321, 15883341, 15883351, 15883361, 15883371, 15883381, 15883391, 15883401, 15883411, 15883421, 15883431, 15883441, 15883451, 15883461, 15883481, 15883491, 15883501, 15883511, 15883521, 15883531, 15883541, 15883551, 15883571, 15883581, 15883591, 15883601, 15883611, 15883621, 15883631, 15883651, 15883661, 15883671, 15883711, 15883721, 15883731, 15883741, 15883751, 15883761, 15883771, 15883781, 15883791, 15883801, 15883811, 15883821, 15883831, 15883841, 15883851, 15883861, 15883871, 15883881, 15883891, 15883901, 15883911, 15883921, 15883931, 15883941, 15883951, 15883961, 15883971, 15883981, 15883991, 15884001, 15884011, 15884021, 15884031, 15884041, 15884051, 15884061, 15884071, 15884081, 15884091, 15884101, 15884111, 15884121, 15884131, 15884141, 15884151, 15884161, 15884171, 15884181, 15884191, 15884201, 15884211, 15884221, 15884231, 15884241, 15884251, 15884261, 15884271, 15884281, 15884291, 15884301, 15884311, 15884321, 15884331, 15884341, 15884351, 15884361, 15884371, 15884381, 15884391, 15884401, 15884411, 15884421, 15884431, 15884441, 15884451, 15884461, 15884481, 15884491, 15884501, 15884511, 15884521, 15884531, 15884541, 15884551, 15884561, 15884571, 15884581, 15884591, 15884601, 15884611, 15884621, 15884631, 15884641, 15884651, 15884661, 15884671, 15884681, 15884691, 15884701, 15884711, 15884721, 15884731, 15884741, 15884751, 15884761, 15884771, 15884781, 15884791, 15884801, 15884811, 15884821, 15884831, 15884841, 15884851, 15884861, 15884871, 15884881, 15884891, 15884901, 15884911, 15884921, 15884931, 15884941, 15884951, 15884961, 15884971, 15884981, 15884991, 15885001, 15885011, 15885021, 15885031, 15885041, 15885051, 15885061, 15885071, 15885081, 15885091, 15885101, 15885111, 15885121, 15885131, 15885141, 15885151, 15885161, 15885171, 15885181, 15885191, 15885201, 15885211, 15885221, 15885231, 15885241, 15885251, 15885261, 15885271, 15885281, 15885291, 15885301, 15885311, 15885321, 15885331, 15885341, 15885351, 15885361, 15885371, 15885381, 15885391, 15885401, 15885411, 15885421, 15885431, 15885441, 15885451, 15885461, 15885471, 15885481, 15885491, 15885501, 15885511, 15885521, 15885531, 15885541, 15885551, 15885561, 15885571, 15885581, 15885591, 15885601, 15885611, 15885621, 15885631, 15885641, 15885651, 15885661, 15885671, 15885681, 15885691, 15885701, 15885711, 15885721, 15885731, 15885741, 15885751, 15885761, 15885771, 15885781, 15885791, 15885801, 15885811, 15885821, 15885831, 15885841, 15885851, 15885861, 15885871, 15885881, 15885891, 158
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Hospira Inc.
- Reason for Recall:
- The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Plum A+3 Infusion Pump with Hospira MedNet Software. List Number 20678. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.
Product Codes/Lot Numbers:
List number 20678. Serial numbers: 15541001, 15541011, 15541021, 15541031, 15541041, 15541051, 15541061, 15541071, 15541081, 15541091, 15541101, 15541111, 15541121, 15541131, 15541141, 15541151, 15541161, 15541171, 15541181, 15541191, 15541201, 15541211, 15541231, 15541241, 15541261, 15541271, 15541281, 15541291, 15541301, 15541311, 15541321, 15541331, 15541341, 15541351, 15541361, 15541371, 15541381, 15541391, 15541401, 15541411, 15541421, 15541431, 15541441, 15541451, 15541461, 15541471, 15541481, 15541491, 15541501, 15541511, 15541521, 15541531, 15541541, 15541551, 15541561, 15541571, 15541581, 15541591, 15541601, 15541611, 15541621, 15541631, 15541641, 15541651, 15541661, 15541671, 15541681, 15541691, 15541701, 15541711, 15541721, 15541731, 15541741, 15541751, 15541761, 15541771, 15541781, 15541791, 15541801, 15541811, 15541821, 15541831, 15541841, 15541851, 15541861, 15541871, 15541881, 15541891, 15541901, 15541911, 15541921, 15541931, 15541941, 15541951, 15541961, 15541971, 15541981, 15541991, 15542001, 15542011, 15542021, 15542031, 15542041, 15542051, 15542061, 15542071, 15542081, 15542091, 15542101, 15542111, 15542121, 15542131, 15542141, 15542151, 15542161, 15542171, 15542181, 15542191, 15542201, 15542211, 15542221, 15542231, 15542241, 15542251, 15542261, 15542271, 15542281, 15542291, 15542301, 15542311, 15542321, 15542331, 15542341, 15542351, 15542361, 15542371, 15542381, 15542391, 15542401, 15542411, 15542431, 15542441, 15542451, 15542461, 15542471, 15542481, 15542491, 15542501, 15542511, 15542521, 15542531, 15542541, 15542551, 15542561, 15542571, 15542581, 15542591, 15542601, 15542611, 15542621, 15542631, 15542641, 15542651, 15542661, 15542671, 15542681, 15542691, 15542701, 15542711, 15542721, 15542731, 15542741, 15542751, 15542761, 15542771, 15542781, 15542791, 15542801, 15542811, 15542821, 15542831, 15542841, 15542851, 15542861, 15542871, 15542881, 15542891, 15542911, 15542921, 15542931, 15542941, 15542951, 15542961, 15542971, 15542981, 15542991, 15543001, 15543011, 15543021, 15543031, 15543041, 15543051, 15543061, 15543071, 15543081, 15543101, 15543111, 15543121, 15543131, 15543141, 15543151, 15543161, 15543181, 15543191, 15543211, 15543221, 15543231, 15543241, 15543261, 15543271, 15543281, 15543291, 15543301, 15543311, 15543321, 15543331, 15543341, 15543351, 15543361, 15543371, 15543391, 15543401, 15543411, 15543421, 15543431, 15543441, 15543451, 15543461, 15543471, 15543481, 15543491, 15543501, 15543511, 15543521, 15543531, 15543541, 15543551, 15543571, 15543581, 15543591, 15543601, 15543611, 15543621, 15543631, 15543641, 15543651, 15543661, 15543671, 15543681, 15543691, 15543711, 15543721, 15543731, 15543741, 15543751, 15543761, 15543771, 15543781, 15543791, 15543801, 15543811, 15543821, 15543831, 15543841, 15543851, 15543861, 15543871, 15543881, 15543891, 15543901, 15543911, 15543921, 15543931, 15543941, 15543951, 15543961, 15543971, 15543981, 15543991, 15544001, 15544011, 15544021, 15544031, 15544041, 15544051, 15544061, 15544071, 15544081, 15544091, 15544101, 15544111, 15544121, 15544131, 15544141, 15544151, 15544161, 15544191, 15544201, 15544211, 15544221, 15544231, 15544241, 15544251, 15544261, 15544271, 15544281, 15544291, 15544301, 15544311, 15544321, 15544331, 15544341, 15544351, 15544371, 15544381, 15544401, 15544411, 15544421, 15544431, 15544441, 15544451, 15544461, 15544471, 15544481, 15544501, 15544511, 15544521, 15544531, 15544541, 15544551, 15544561, 15544571, 15544581, 15544591, 15544601, 15544611, 15544621, 15544641, 15544661, 15544671, 15544681, 15544691, 15544701, 15544711, 15544721, 15544731, 15544741, 15544751, 15544761, 15544771, 15544781, 15544791, 15544801, 15544811, 15544821, 15544831, 15544841, 15544851, 15544871, 15544891, 15544901, 15544911, 15544921, 15544931, 15544941, 15544951, 15544961, 15544971, 15544981, 15544991, 15545001, 15545021, 15545031, 15545041, 15545051, 15545061, 15545071, 15545091, 15545101, 15545111, 15545121, 15545131, 15545141, 15545151, 15545161, 15545171, 15545181, 15545191, 15545201, 15545211, 15545231, 15545241, 15545251, 15545261, 15545271, 15545281, 15545291, 15545301, 15545311, 15545321, 15545331, 15545341, 15545351, 15545361, 15545371, 15545381, 15545391, 15545401, 15545411, 15545421, 15545431, 15545441, 15545451, 15545461, 15545471, 15545481, 15545491, 15545501, 15545511, 15545521, 15545531, 15545541, 15545551, 15545561, 15545571, 15545581, 15545591, 15545601, 15545611, 15545621, 15545631, 15545641, 15545651, 15545661, 15545671, 15545691, 15545701, 15545711, 15545721, 15545741, 15545751, 15545761, 15545771, 15545781, 15545791, 15545801, 15545811, 15545821, 15545831, 15545841, 15545851, 15545861, 15545871, 15545881, 15545891, 15545911, 15545921, 15545931, 15545941, 15545951, 15545961, 15545971, 15545981, 15545991, 15546001, 15546011, 15546021, 15546031, 15546041, 15546051, 15546061, 15546071, 15546081, 15546091, 15546101, 15546121, 15546131, 15546141, 15546151, 15546161, 15546171, 15546181, 15546191, 15546201, 15546211, 15546221, 15546231, 15546241, 15546251, 15546281, 15546291, 15546301, 15546311, 15546321, 15546331, 15546341, 15546351, 15546361, 15546371, 15546381, 15546391, 15546401, 15546411, 15546421, 15546431, 15546441, 15546451, 15546461, 15546481, 15546501, 15546511, 15546521, 15546531, 15546541, 15546551, 15546561, 15546571, 15546581, 15546591, 15546601, 15546611, 15546621, 15546631, 15546641, 15546651, 15546661, 15546671, 15546681, 15546691, 15546701, 15546711, 15546721, 15546731, 15546741, 15546751, 15546761, 15546771, 15546791, 15546801, 15546811, 15546821, 15546831, 15546841, 15546851, 15546861, 15546871, 15546881, 15546891, 15546901, 15546911, 15546921, 15546931, 15546941, 15546951, 15546961, 15546971, 15546981, 15546991, 15547001, 15547011, 15547021, 15547031, 15547041, 15547051, 15547061, 15547071, 15547081, 15547091, 15547111, 15547121, 15547131, 15547141, 15547151, 15547161, 15547171, 15547181, 15547191, 15547201, 15547211, 15547221, 15547231, 15547241, 15547251, 15547261, 15547271, 15547281, 15547291, 15547301, 15547321, 15547331, 15547341, 15547351, 15547361, 15547371, 15547381, 15547391, 15547401, 15547411, 15547421, 15547431, 15547441, 15547451, 15547461, 15547471, 15547481, 15547491, 15547501, 15547511, 15547521, 15547531, 15547541, 15547551, 15547561, 15547571, 15547581, 15547591, 15547601, 15547611, 15547621, 15547631, 15547641, 15547651, 15547661, 15547671, 15547681, 15547691, 15547701, 15547711, 15547721, 15547731, 15547741, 15547751, 15547761, 15547771, 15547781, 15547791, 15547801, 15547811, 15547821, 15547831, 15547851, 15547861, 15547871, 15547881, 15547891, 15547901, 15547911, 15547921, 15547931, 15547941, 15547951, 15547961, 15547971, 15547981, 15548001, 15548011, 15548031, 15548041, 15548051, 15548061, 15548071, 15548081, 15548091, 15548101, 15548111, 15548121, 15548131, 15548141, 15548151, 15548161, 15548171, 15548181, 15548191, 15548201, 15548211, 15548221, 15548231, 15548241, 15548251, 15548261, 15548271, 15548281, 15548291, 15548301, 15548311, 15548321, 15548331, 15548341, 15548351, 15548361, 15548371, 15548381, 15548391, 15548401, 15548411, 15548421, 15548431, 15548441, 15548451, 15548461, 15548471, 15548481, 15548491, 15548501, 15548511, 15548521, 15548531, 15548541, 15548551, 15548561, 15548571, 15548581, 15548591, 15548601, 15548611, 15548621, 15548631, 15548641, 15548651, 15548661, 15548671, 15548681, 15548691, 15548701, 15548711, 15548731, 15548741, 15548751, 15548761, 15548771, 15548781, 15548791, 15548801, 15548811, 15548821, 15548831, 15548841, 15548851, 15548861, 15548871, 15548881, 15548891, 15548901, 15548911, 15548921, 15548931, 15548941, 15548961, 15548971, 15548981, 15548991, 15549001, 15549011, 15549031, 15549041, 15549051, 15549061, 15549071, 15549081, 15549091, 15549101, 15549111, 15549121, 15549131, 15549141, 15549151, 15549161, 15549171, 15549181, 15549191, 15549201, 15549211, 15549221, 15549231, 15549241, 15549251, 15549261, 15549271, 15549281, 15549291, 15549301, 15549321, 15549331, 15549341, 15549351, 15549361, 15549371, 15549381, 15549391, 15549401, 15549411, 15549421, 15549431, 15549441, 15549451, 15549461, 15549471, 15549481, 15549491, 15549501, 15549511, 15549521, 15549531, 15549541, 15549551, 15549561, 15549571, 15549581, 15549591, 15549601, 15549611, 15549621, 15549641, 15549651, 15549661, 15549671, 15549681, 15549691, 15549701, 15549711, 15549721, 15549741, 15549751, 15549761, 15549771, 15549781, 15549791, 15549801, 15549811, 15549821, 15549831, 15549841, 15549851, 15549861, 15549871, 15549881, 15549891, 15549901, 15549911, 15549921, 15549931, 15549941, 15549951, 15549961, 15549971, 15549981, 15550001, 15550011, 15550021, 15550031, 15550041, 15550051, 15550061, 15550071, 15550081, 15550091, 15550101, 15550111, 15550121, 15550131, 15550141, 15550151, 15550161, 15550171, 15550181, 15550191, 15550201, 15550211, 15550221, 15550231, 15550241, 15550251, 15550261, 15550271, 15550281, 15550291, 15550301, 15550311, 15550321, 15550331, 15550341, 15550351, 15550361, 15550371, 15550381, 15550391, 15550401, 15550411, 15550421, 15550431, 15550441, 15550451, 15550461, 15550481, 15550491, 15550511, 15550521, 15550531, 15550541, 15550551, 15550561, 15550571, 15550581, 15550591, 15550601, 15550611, 15550621, 15550631, 15550641, 15550651, 15550661, 15550671, 15550681, 15550691, 15550701, 15550711, 15550721, 15550731, 15550741, 15550751, 15550761, 15550771, 15550781, 15550791, 15550801, 15550811, 15550821, 15550831, 15550841, 15550851, 15550861, 15550871, 15550881, 15550901, 15550911, 15550921, 15550931, 15550941, 15550951, 15550961, 15550971, 15550981, 15550991, 15861001, 15861011, 15861021, 15861031, 15861041, 15861051, 15861061, 15861071, 15861081, 15861091, 15861101, 15861111, 15861121, 15861131, 15861141, 15861151, 15861161, 15861171, 15861181, 15861191, 15861201, 15861211, 15861221, 15861231, 15861241, 15861251, 15861261, 15861271, 15861281, 15861291, 15861301, 15861311, 15861321, 15861331, 15861341, 15861351, 15861361, 15861371, 15861381, 15861391, 15861401, 15861411, 15861421, 15861431, 15861441, 15861451, 15861461, 15861471, 15861481, 15861491, 15861501, 15861511, 15861521, 15861531, 15861541, 15861551, 15861561, 15861571, 15861581, 15861591, 15861601, 15861621, 15861631, 15861641, 15861651, 15861661, 15861681, 15861691, 15861701, 15861711, 15861721, 15861731, 15861741, 15861751, 15861761, 15861771, 15861781, 15861791, 15861801, 15861821, 15861831, 15861841, 15861851, 15861861, 15861871, 15861881, 15861891, 15861901, 15861921, 15861941, 15861951, 15861961, 15861981, 15861991, 15862001, 15862011, 15862021, 15862031, 15862041, 15862051, 15862061, 15862071, 15862081, 15862101, 15862111, 15862121, 15862131, 15862141, 15862151, 15862161, 15862171, 15862181, 15862191, 15862201, 15862211, 15862221, 15862231, 15862251, 15862261, 15862271, 15862281, 15862291, 15862301, 15862311, 15862321, 15862331, 15862341, 15862351, 15862361, 15862371, 15862451, 15862461, 15862471, 15862481, 15862491, 15862501, 15862511, 15862521, 15862531, 15862541, 15862551, 15862561, 15862571, 15862581, 15862591, 15862601, 15862611, 15862621, 15862631, 15862641, 15862651, 15862661, 15862671, 15862681, 15862691, 15862701, 15862711, 15862721, 15862731, 15862741, 15862751, 15862781, 15862791, 15862811, 15862821, 15862831, 15862841, 15862851, 15862861, 15862871, 15862881, 15862891, 15862911, 15862921, 15862931, 15862941, 15862951, 15862961, 15862971, 15862981, 15862991, 15863001, 15863021, 15863031, 15863041, 15863051, 15863061, 15863071, 15863081, 15863091, 15863101, 15863111, 15863121, 15863131, 15863141, 15863151, 15863161, 15863171, 15863181, 15863191, 15863201, 15863211, 15863221, 15863231, 15863241, 15863251, 15863261, 15863271, 15863281, 15863291, 15863301, 15863311, 15863321, 15863331, 15863341, 15863351, 15863361, 15863371, 15863381, 15863391, 15863401, 15863411, 15863421, 15863431, 15863441, 15863451, 15863461, 15863471, 15863481, 15863491, 15863501, 15863511, 15863521, 15863531, 15863541, 15863551, 15863561, 15863571, 15863581, 15863591, 15863601, 15863611, 15863621, 15863631, 15863641, 15863651, 15863661, 15863671, 15863681, 15863691, 15863701, 15863711, 15863721, 15863731, 15863741, 15863751, 15863761, 15863771, 15863781, 15863791, 15863801, 15863811, 15863821, 15863831, 15863841, 15863851, 15863861, 15863871, 15863881, 15863891, 15863901, 15863911, 15863921, 15863931, 15863941, 15863951, 15863961, 15863971, 15863981, 15863991, 15864001, 15864011, 15864021, 15864031, 15864041, 15864051, 15864061, 15864071, 15864081, 15864091, 15864101, 15864111, 15864131, 15864141, 15864151, 15864161, 15864171, 15864181, 15864191, 15864201, 15864211, 15864221, 15864231, 15864241, 15864251, 15864261, 15864271, 15864281, 15864291, 15864311, 15864321, 15864331, 15864351, 15864371, 15864381, 15864391, 15864411, 15864421, 15864431, 15864441, 15864471, 15864481, 15864491, 15864501, 15864511, 15864521, 15864531, 15864541, 15864551, 15864561, 15864571, 15864581, 15864591, 15864601, 15864611, 15864621, 15864631, 15864641, 15864651, 15864661, 15864671, 15864681, 15864691, 15864701, 15864711, 15864721, 15864731, 15864751, 15864761, 15864771, 15864781, 15864791, 15864801, 15864811, 15864821, 15864831, 15864841, 15864851, 15864861, 15864871, 15864881, 15864891, 15864901, 15864911, 15864921, 15864931, 15864941, 15864951, 15864961, 15864971, 15864981, 15864991, 15865001, 15865011, 15865021, 15865031, 15865041, 15865051, 15865061, 15865071, 15865081, 15865101, 15865111, 15865121, 15865131, 15865141, 15865151, 15865161, 15865171, 15865181, 15865191, 15865201, 15865211, 15865221, 15865231, 15865241, 15865251, 15865261, 15865271, 15865281, 15865291, 15865301, 15865311, 15865321, 15865331, 15865341, 15865351, 15865361, 15865371, 15865381, 15865391, 15865401, 15865411, 15865421, 15865431, 15865441, 15865451, 15865461, 15865471, 15865481, 15865491, 15865501, 15865511, 15865521, 15865551, 15865561, 15865571, 15865591, 15865601, 15865611, 15865621, 15865631, 15865641, 15865651, 15865661, 15865671, 15865681, 15865691, 15865701, 15865711, 15865721, 15865731, 15865741, 15865751, 15865761, 15865771, 15865781, 15865791, 15865801, 15865811, 15865821, 15865831, 15865841, 15865851, 15865861, 15865871, 15865881, 15865891, 15865901, 15865911, 15865921, 15865931, 15865941, 15865951, 15865961, 15865971, 15865981, 15865991, 15866001, 15866011, 15866021, 15866031, 15866041, 15866051, 15866061, 15866071, 15866081, 15866091, 15866101, 15866111, 15866121, 15866131, 15866141, 15866151, 15866161, 15866171, 15866181, 15866191, 15866201, 15866211, 15866221, 15866231, 15866241, 15866251, 15866261, 15866271, 15866291, 15866301, 15866311, 15866321, 15866331, 15866341, 15866351, 15866361, 15866371, 15866381, 15866391, 15866401, 15866411, 15866421, 15866431, 15866451, 15866461, 15866471, 15866481, 15866491, 15866501, 15866511, 15866521, 15866531, 15866541, 15866551, 15866561, 15866571, 15866581, 15866591, 15866601, 15866611, 15866621, 15866631, 15866641, 15866651, 15866661, 15866671, 15866681, 15866691, 15866701, 15866711, 15866721, 15866731, 15866741, 15866751, 15866761, 15866771, 15866781, 15866791, 15866801, 15866811, 15866821, 15866831, 15866841, 15866851, 15866861, 15866871, 15866881, 15866891, 15866901, 15866911, 15866921, 15866931, 15866941, 15866951, 15866961, 15866971, 15866981, 15866991, 15867001, 15867011, 15867021, 15867031, 15867051, 15867061, 15867071, 15867081, 15867091, 15867101, 15867111, 15867121, 15867131, 15867141, 15867151, 15867171, 15867181, 15867191, 15867201, 15867211, 15867221, 15867231, 15867241, 15867251, 15867261, 15867271, 15867281, 15867291, 15867301, 15867311, 15867321, 15867331, 15867341, 15867351, 15867361, 15867371, 15867381, 15867391, 15867401, 15867411, 15867421, 15867431, 15867441, 15867451, 15867461, 15867471, 15867481, 15867491, 15867501, 15867511, 15867521, 15867531, 15867541, 15867551, 15867561, 15867571, 15867581, 15867591, 15867601, 15867611, 15867621, 15867631, 15867641, 15867651, 15867661, 15867671, 15867681, 15867691, 15867701, 15867711, 15867721, 15867731, 15867741, 15867751, 15867761, 15867771, 15867781, 15867791, 15867801, 15867811, 15867821, 15867831, 15867841, 15867851, 15867861, 15867871, 15867881, 15867891, 15867901, 15867911, 15867921, 15867931, 15867941, 15867951, 15867961, 15867971, 15867981, 15867991, 15868001, 15868011, 15868021, 15868031, 15868051, 15868061, 15868071, 15868081, 15868091, 15868101, 15868111, 15868121, 15868131, 15868141, 15868151, 15868161, 15868171, 15868181, 15868191, 15868201, 15868211, 15868221, 15868231, 15868241, 15868251, 15868261, 15868271, 15868281, 15868301, 15868311, 15868321, 15868331, 15868341, 15868351, 15868361, 15868371, 15868381, 15868391, 15868401, 15868411, 15868421, 15868431, 15868441, 15868451, 15868461, 15868471, 15868481, 15868491, 15868501, 15868511, 15868531, 15868541, 15868551, 15868561, 15868571, 15868581, 15868591, 15868601, 15868611, 15868621, 15868631, 15868641, 15868651, 15868661, 15868671, 15868681, 15868691, 15868701, 15868711, 15868721, 15868731, 15868741, 15868751, 15868761, 15868771, 15868781, 15868791, 15868801, 15868811, 15868821, 15868831, 15868841, 15868851, 15868861, 15868871, 15868881, 15868891, 15868901, 15868911, 15868921, 15868931, 15868941, 15868951, 15868961, 15868971, 15868981, 15869001, 15869011, 15869021, 15869031, 15869041, 15869051, 15869061, 15869071, 15869081, 15869091, 15869101, 15869111, 15869121, 15869131, 15869141, 15869151, 15869161, 15869181, 15869191, 15869201, 15869211, 15869221, 15869231, 15869241, 15869251, 15869261, 15869271, 15869281, 15869291, 15869301, 15869311, 15869321, 15869331, 15869341, 15869351, 15869361, 15869371, 15869381, 15869401, 15869411, 15869421, 15869441, 15869451, 15869461, 15869471, 15869481, 15869491, 15869501, 15869511, 15869521, 15869531, 15869541, 15869551, 15869561, 15869571, 15869581, 15869591, 15869601, 15869611, 15869621, 15869631, 15869641, 15869651, 15869661, 15869671, 15869681, 15869691, 15869701, 15869711, 15869721, 15869731, 15869741, 15869751, 15869761, 15869771, 15869781, 15869791, 15869811, 15869821, 15869831, 15869841, 15869851, 15869861, 15869871, 15869881, 15869891, 15869901, 15869911, 15869921, 15869931, 15869941, 15869951, 15869961, 15869971, 15869981, 15869991, 15870001, 15870011, 15870021, 15870031, 15870041, 15870051, 15870061, 15870071, 15870081, 15870091, 15870101, 15870111, 15870121, 15870131, 15870141, 15870151, 15870161, 15870171, 15870181, 15870191, 15870201, 15870211, 15870221, 15870231, 15870241, 15870251, 15870261, 15870271, 15870281, 15870291, 15870301, 15870311, 15870321, 15870331, 15870341, 15870351, 15870361, 15870371, 15870381, 15870391, 15870401, 15870411, 15870421, 15870431, 15870441, 15870461, 15870471, 15870481, 15870491, 15870501, 15870511, 15870521, 15870531, 15870541, 15870551, 15870561, 15870571, 15870581, 15870591, 15870601, 15870611, 15870631, 15870651, 15870661, 15870671, 15870681, 15870691, 15870701, 15870711, 15870721, 15870731, 15870741, 15870751, 15870771, 15870781, 15870791, 15870801, 15870811, 15870821, 15870831, 15870841, 15870851, 15870861, 15870881, 15870891, 15870901, 15870911, 15870931, 15870951, 15870961, 15870971, 15870981, 15870991, 15871001, 15871011, 15871021, 15871031, 15871041, 15871061, 15871071, 15871081, 15871091, 15871101, 15871111, 15871121, 15871131, 15871141, 15871151, 15871161, 15871171, 15871181, 15871191, 15871201, 15871211, 15871221, 15871231, 15871241, 15871251, 15871261, 15871271, 15871281, 15871291, 15871301, 15871311, 15871321, 15871331, 15871341, 15871351, 15871361, 15871371, 15871381, 15871391, 15871401, 15871411, 15871431, 15871441, 15871451, 15871461, 15871481, 15871491, 15871501, 15871511, 15871521, 15871531, 15871541, 15871551, 15871561, 15871571, 15871581, 15871591, 15871601, 15871611, 15871621, 15871631, 15871641, 15871651, 15871661, 15871671, 15871681, 15871691, 15871701, 15871711, 15871721, 15871731, 15871741, 15871751, 15871761, 15871771, 15871781, 15871791, 15871801, 15871811, 15871821, 15871831, 15871841, 15871851, 15871861, 15871871, 15871891, 15871901, 15871911, 15871921, 15871931, 15871941, 15871951, 15871961, 15871971, 15871981, 15871991, 15872001, 15872011, 15872021, 15872031, 15872041, 15872051, 15872061, 15872071, 15872081, 15872091, 15872101, 15872111, 15872121, 15872131, 15872141, 15872151, 15872161, 15872171, 15872181, 15872191, 15872201, 15872211, 15872221, 15872231, 15872241, 15872251, 15872261, 15872271, 15872281, 15872291, 15872301, 15872311, 15872321, 15872331, 15872341, 15872351, 15872361, 15872371, 15872381, 15872391, 15872401, 15872411, 15872421, 15872431, 15872441, 15872451, 15872471, 15872481, 15872491, 15872501, 15872511, 15872521, 15872531, 15872541, 15872551, 15872561, 15872571, 15872581, 15872591, 15872601, 15872611, 15872621, 15872631, 15872641, 15872651, 15872661, 15872671, 15872681, 15872691, 15872701, 15872711, 15872721, 15872731, 15872741, 15872751, 15872761, 15872771, 15872781, 15872791, 15872801, 15872811, 15872821, 15872841, 15872851, 15872861, 15872871, 15872881, 15872891, 15872901, 15872911, 15872921, 15872931, 15872941, 15872951, 15872961, 15872971, 15872981, 15872991, 15873001, 15873011, 15873021, 15873031, 15873041, 15873051, 15873061, 15873071, 15873081, 15873091, 15873101, 15873121, 15873131, 15873141, 15873151, 15873161, 15873171, 15873181, 15873191, 15873201, 15873211, 15873221, 15873231, 15873241, 15873261, 15873271, 15873281, 15873291, 15873301, 15873311, 15873321, 15873331, 15873341, 15873351, 15873361, 15873371, 15873381, 15873391, 15873401, 15873411, 15873421, 15873431, 15873441, 15873451, 15873461, 15873471, 15873481, 15873491, 15873501, 15873511, 15873521, 15873531, 15873541, 15873551, 15873571, 15873581, 15873591, 15873601, 15873611, 15873621, 15873631, 15873641, 15873651, 15873661, 15873671, 15873681, 15873691, 15873701, 15873711, 15873721, 15873731, 15873741, 15873751, 15873771, 15873781, 15873791, 15873801, 15873811, 15873821, 15873831, 15873841, 15873851, 15873861, 15873871, 15873881, 15873891, 15873901, 15873921, 15873931, 15873941, 15873951, 15873961, 15873971, 15873981, 15873991, 15874001, 15874011, 15874021, 15874031, 15874041, 15874051, 15874061, 15874071, 15874081, 15874091, 15874101, 15874111, 15874121, 15874131, 15874141, 15874151, 15874161, 15874171, 15874181, 15874191, 15874201, 15874211, 15874221, 15874231, 15874241, 15874251, 15874271, 15874281, 15874291, 15874301, 15874311, 15874321, 15874331, 15874341, 15874351, 15874361, 15874371, 15874381, 15874391, 15874401, 15874421, 15874431, 15874441, 15874451, 15874461, 15874471, 15874481, 15874491, 15874501, 15874511, 15874521, 15874531, 15874541, 15874551, 15874561, 15874571, 15874581, 15874591, 15874601, 15874611, 15874621, 15874631, 15874641, 15874651, 15874661, 15874671, 15874681, 15874691, 15874701, 15874711, 15874721, 15874731, 15874741, 15874751, 15874761, 15874771, 15874781, 15874791, 15874801, 15874811, 15874821, 15874831, 15874841, 15874851, 15874861, 15874871, 15874881, 15874891, 15874901, 15874911, 15874921, 15874931, 15874941, 15874951, 15874961, 15874971, 15874981, 15874991, 15875001, 15875011, 15875021, 15875031, 15875041, 15875051, 15875061, 15875071, 15875081, 15875091, 15875101, 15875111, 15875121, 15875131, 15875141, 15875151, 15875161, 15875171, 15875181, 15875191, 15875211, 15875221, 15875231, 15875241, 15875251, 15875261, 15875271, 15875281, 15875291, 15875301, 15875311, 15875321, 15875331, 15875341, 15875351, 15875361, 15875371, 15875381, 15875391, 15875401, 15875411, 15875421, 15875431, 15875441, 15875451, 15875461, 15875471, 15875481, 15875491, 15875501, 15875511, 15875521, 15875531, 15875541, 15875551, 15875561, 15875581, 15875591, 15875601, 15875611, 15875621, 15875631, 15875641, 15875651, 15875671, 15875681, 15875701, 15875711, 15875721, 15875731, 15875741, 15875751, 15875761, 15875771, 15875781, 15875791, 15875801, 15875811, 15875821, 15875841, 15875851, 15875861, 15875871, 15875881, 15875891, 15876101, 15876111, 15876121, 15876131, 15876141, 15876151, 15876161, 15876171, 15876181, 15876191, 15876201, 15876211, 15876221, 15876231, 15876241, 15876251, 15876261, 15876271, 15876281, 15876291, 15876311, 15876321, 15876331, 15876341, 15876351, 15876361, 15876371, 15876381, 15876391, 15876401, 15876411, 15876421, 15876431, 15876441, 15876451, 15876461, 15876471, 15876481, 15876491, 15876501, 15876511, 15876521, 15876531, 15876541, 15876551, 15876561, 15876571, 15876581, 15876591, 15876601, 15876621, 15876631, 15876641, 15876651, 15876661, 15876671, 15876681, 15876691, 15876701, 15876711, 15876721, 15876731, 15876741, 15876751, 15876761, 15876771, 15876781, 15876791, 15876801, 15876811, 15876821, 15876831, 15876841, 15876851, 15876861, 15876871, 15876881, 15876891, 15876901, 15876911, 15876921, 15876931, 15876941, 15876951, 15876961, 15876971, 15876981, 15876991, 15877001, 15877031, 15877041, 15877061, 15877081, 15877091, 15877111, 15877121, 15877131, 15877151, 15877161, 15877171, 15877181, 15877201, 15877211, 15877221, 15877231, 15877241, 15877251, 15877261, 15877271, 15877281, 15877291, 15877301, 15877311, 15877321, 15877331, 15877341, 15877351, 15877361, 15877371, 15877381, 15877391, 15877401, 15877411, 15877421, 15877441, 15877451, 15877461, 15877471, 15877491, 15877501, 15877511, 15877521, 15877531, 15877541, 15877551, 15877571, 15877581, 15877591, 15877601, 15877611, 15877621, 15877631, 15877641, 15877651, 15877661, 15877671, 15877681, 15877691, 15877701, 15877711, 15877721, 15877731, 15877741, 15877751, 15877761, 15877771, 15877781, 15877791, 15877801, 15877811, 15877821, 15877831, 15877841, 15877851, 15877861, 15877871, 15877881, 15877891, 15877901, 15877911, 15877921, 15877931, 15877941, 15877951, 15877961, 15877971, 15877981, 15877991, 15878001, 15878021, 15878031, 15878051, 15878061, 15878071, 15878081, 15878091, 15878101, 15878111, 15878121, 15878131, 15878141, 15878151, 15878161, 15878171, 15878181, 15878191, 15878201, 15878211, 15878221, 15878231, 15878241, 15878251, 15878261, 15878271, 15878281, 15878291, 15878301, 15878311, 15878321, 15878331, 15878341, 15878351, 15878361, 15878381, 15878391, 15878401, 15878411, 15878421, 15878431, 15878441, 15878451, 15878461, 15878471, 15878481, 15878491, 15878501, 15878521, 15878531, 15878541, 15878551, 15878561, 15878571, 15878581, 15878591, 15878601, 15878611, 15878621, 15878631, 15878641, 15878651, 15878661, 15878671, 15878681, 15878691, 15878701, 15878711, 15878721, 15878731, 15878741, 15878751, 15878761, 15878771, 15878781, 15878791, 15878801, 15878811, 15878831, 15878841, 15878851, 15878861, 15878871, 15878881, 15878901, 15878911, 15878921, 15878931, 15878951, 15878961, 15878971, 15878981, 15879001, 15879011, 15879021, 15879031, 15879041, 15879051, 15879061, 15879071, 15879081, 15879091, 15879101, 15879121, 15879131, 15879141, 15879151, 15879161, 15879171, 15879181, 15879191, 15879211, 15879221, 15879231, 15879241, 15879251, 15879261, 15879271, 15879281, 15879291, 15879301, 15879321, 15879331, 15879341, 15879351, 15879361, 15879371, 15879381, 15879401, 15879411, 15879421, 15879431, 15879441, 15879451, 15879461, 15879471, 15879481, 15879491, 15879501, 15879511, 15879521, 15879531, 15879541, 15879551, 15879561, 15879571, 15879581, 15879591, 15879601, 15879611, 15879621, 15879631, 15879651, 15879661, 15879671, 15879681, 15879691, 15879701, 15879711, 15879721, 15879731, 15879741, 15879751, 15879771, 15879781, 15879791, 15879801, 15879811, 15879821, 15879841, 15879851, 15879861, 15879871, 15879881, 15879891, 15879901, 15879911, 15879921, 15879931, 15879941, 15879951, 15879961, 15879971, 15879981, 15879991, 15880001, 15880011, 15880021, 15880031, 15880041, 15880051, 15880061, 15880071, 15880081, 15880091, 15880101, 15880111, 15880121, 15880131, 15880141, 15880151, 15880161, 15880171, 15880181, 15880201, 15880211, 15880221, 15880231, 15880241, 15880251, 15880261, 15880271, 15880281, 15880291, 15880301, 15880311, 15880321, 15880331, 15880341, 15880351, 15880361, 15880371, 15880381, 15880391, 15880401, 15880411, 15880421, 15880431, 15880441, 15880451, 15880471, 15880481, 15880491, 15880511, 15880521, 15880531, 15880541, 15880551, 15880561, 15880571, 15880581, 15880591, 15880601, 15880611, 15880621, 15880631, 15880651, 15880661, 15881121, 15881131, 15881141, 15881151, 15881161, 15881171, 15881181, 15881191, 15881201, 15881211, 15881221, 15881231, 15881241, 15881251, 15881261, 15881271, 15881281, 15881291, 15881301, 15881311, 15881321, 15881331, 15881341, 15881361, 15881371, 15881381, 15881391, 15881401, 15881411, 15881421, 15881431, 15881441, 15881451, 15881461, 15881471, 15881481, 15881491, 15881501, 15881511, 15881521, 15881531, 15881541, 15881561, 15881571, 15881581, 15881591, 15881601, 15881611, 15881621, 15881631, 15881641, 15881651, 15881661, 15881671, 15881681, 15881691, 15881711, 15881721, 15881731, 15881741, 15881751, 15881771, 15881781, 15881791, 15881801, 15881811, 15881821, 15881831, 15881841, 15881851, 15881861, 15881871, 15881881, 15881891, 15881901, 15881921, 15881931, 15881941, 15881951, 15881961, 15881971, 15881981, 15881991, 15882001, 15882011, 15882021, 15882031, 15882041, 15882051, 15882061, 15882071, 15882091, 15882101, 15882111, 15882121, 15882131, 15882141, 15882151, 15882161, 15882171, 15882181, 15882191, 15882201, 15882211, 15882221, 15882231, 15882241, 15882251, 15882261, 15882271, 15882281, 15882291, 15882301, 15882311, 15882321, 15882351, 15882361, 15882371, 15882381, 15882391, 15882401, 15882411, 15882421, 15882431, 15882451, 15882461, 15882471, 15882481, 15882491, 15882501, 15882511, 15882521, 15882531, 15882541, 15882551, 15882561, 15882571, 15882581, 15882591, 15882601, 15882611, 15882621, 15882631, 15882641, 15882651, 15882661, 15882671, 15882691, 15882701, 15882721, 15882731, 15882741, 15882751, 15882761, 15882771, 15882781, 15882791, 15882801, 15882811, 15882821, 15882831, 15882841, 15882851, 15882861, 15882871, 15882891, 15882901, 15882911, 15882921, 15882941, 15882951, 15882961, 15882971, 15882981, 15882991, 15883001, 15883011, 15883021, 15883031, 15883041, 15883051, 15883061, 15883071, 15883081, 15883091, 15883101, 15883111, 15883121, 15883131, 15883141, 15883151, 15883161, 15883171, 15883181, 15883191, 15883201, 15883211, 15883221, 15883231, 15883241, 15883261, 15883271, 15883281, 15883291, 15883301, 15883311, 15883321, 15883341, 15883351, 15883361, 15883371, 15883381, 15883391, 15883401, 15883411, 15883421, 15883431, 15883441, 15883451, 15883461, 15883481, 15883491, 15883501, 15883511, 15883521, 15883531, 15883541, 15883551, 15883571, 15883581, 15883591, 15883601, 15883611, 15883621, 15883631, 15883651, 15883661, 15883671, 15883711, 15883721, 15883731, 15883741, 15883751, 15883761, 15883771, 15883781, 15883791, 15883801, 15883811, 15883821, 15883831, 15883841, 15883851, 15883861, 15883871, 15883881, 15883891, 15883901, 15883911, 15883921, 15883931, 15883941, 15883951, 15883961, 15883971, 15883981, 15883991, 15884001, 15884011, 15884021, 15884031, 15884041, 15884051, 15884061, 15884071, 15884081, 15884091, 15884101, 15884111, 15884121, 15884131, 15884141, 15884151, 15884161, 15884171, 15884181, 15884191, 15884201, 15884211, 15884221, 15884231, 15884241, 15884251, 15884261, 15884271, 15884281, 15884291, 15884301, 15884311, 15884321, 15884331, 15884341, 15884351, 15884361, 15884371, 15884381, 15884391, 15884401, 15884411, 15884421, 15884431, 15884441, 15884451, 15884461, 15884481, 15884491, 15884501, 15884511, 15884521, 15884531, 15884541, 15884551, 15884561, 15884571, 15884581, 15884591, 15884601, 15884611, 15884621, 15884631, 15884641, 15884651, 15884661, 15884671, 15884681, 15884691, 15884701, 15884711, 15884721, 15884731, 15884741, 15884751, 15884761, 15884771, 15884781, 15884791, 15884801, 15884811, 15884821, 15884831, 15884841, 15884851, 15884861, 15884871, 15884881, 15884891, 15884901, 15884911, 15884921, 15884931, 15884941, 15884951, 15884961, 15884971, 15884981, 15884991, 15885001, 15885011, 15885021, 15885031, 15885041, 15885051, 15885061, 15885071, 15885081, 15885091, 15885101, 15885111, 15885121, 15885131, 15885141, 15885151, 15885161, 15885171, 15885181, 15885191, 15885201, 15885211, 15885221, 15885231, 15885241, 15885251, 15885261, 15885271, 15885281, 15885291, 15885301, 15885311, 15885321, 15885331, 15885341, 15885351, 15885361, 15885371, 15885381, 15885391, 15885401, 15885411, 15885421, 15885431, 15885441, 15885451, 15885461, 15885471, 15885481, 15885491, 15885501, 15885511, 15885521, 15885531, 15885541, 15885551, 15885561, 15885571, 15885581, 15885591, 15885601, 15885611, 15885621, 15885631, 15885641, 15885651, 15885661, 15885671, 15885681, 15885691, 15885701, 15885711, 15885721, 15885731, 15885741, 15885751, 15885761, 15885771, 15885781, 15885791, 15885801, 15885811, 15885821, 15885831, 15885841, 15885851, 15885861, 15885871, 15885881, 15885891, 158
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-2730-2016
Related Recalls
Presence of Particulate Matter: human hair found within an internal sample syringe.
Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.
Failed Stability Specifications: The recalled lots did not meet the specification for color and pH throughout shelf life.