Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Numbers: HQC2FMC1801, HQC2FMC 1802, HQC2FMC1805, HQC2FMC1806
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Transonic Systems Inc
Reason for Recall:
Software defect; programming errors in the blood flow probes EPROM. Transonic received several complaints from users indicating that when the flow probe is connected to a flow meter the meter reads Used Probe or Irrelevant Use Number.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Transonic Flow Probe, 2 mm, Catalog Number HQC2FMC Product Usage: to measure flow intra-operatively.

Product Codes/Lot Numbers:

Serial Numbers: HQC2FMC1801, HQC2FMC 1802, HQC2FMC1805, HQC2FMC1806

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2720-2015

Related Recalls

The sterility of the device cannot be assured. There is a potential for an increased risk of infection.

Feb 15, 2018 Diagnostic Equipment Nationwide View Details →