SHERPA NX ACTIVE GUIDING CATHETER, 6F JR6.0, .070", REF SA6JR60. for cardiovascular use

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN: 00613994821782 All Lot/Serial Numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Medtronic Vascular
Reason for Recall:
There is a potential for extensive loss of primary segment material exposing underlying stainless-steel braid wires on a subset of 6F Sherpa catheters.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

SHERPA NX ACTIVE GUIDING CATHETER, 6F JR6.0, .070", REF SA6JR60. for cardiovascular use

Product Codes/Lot Numbers:

GTIN: 00613994821782 All Lot/Serial Numbers

Distribution:

Distributed in: US

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2715-2019

Related Recalls

Packaging for Balloon Catheters may be damaged resulting in loss of sterility.

Mar 23, 2022 Surgical Instruments Nationwide View Details β†’

Endurant, Endurant II and Endurant IIs Stent Graft System: Medtronic Endurant Stent Graft System Model/REF ENBF2313C120EE; ENBF2313C120EE; ENBF2313C145EE; ENBF2313C170EE; ENBF2316C120EE; ENBF2316C145EE; ENBF2316C170EE; ENBF2513C120EE; ENBF2513C145EE; ENBF2513C170EE; ENBF2516C120EE; ENBF2516C145EE; ENBF2516C170EE; ENBF2813C145EE; ENBF2813C170EE; ENBF2816C120EE; ENBF2816C145EE; ENBF2816C170EE; ENBF3216C120EE; ENBF3216C145EE; ENBF3216C170EE; ENBF3616C145EE; ENBF3616C170EE; ENCF2323C45EE; ENCF2525C45EE; ENCF2828C45EE; ENCF3232C45EE; ENCF3636C45EE; ENTF2323C70EE; ENTF2525C70EE; ENTF2828C70EE; Medtronic Endurant II Stent Graft System Model/REF ETBF2313C124E; ETBF2313C124EE; ETBF2313C145EE; ETBF2313C145EJ; ETBF2313C166EE; ETBF2313C166EJ; ETBF2316C124E; ETBF2316C124EE; ETBF2316C124EJ; ETBF2316C145E; ETBF2316C145EE; ETBF2316C145EJ; ETBF2316C166E; ETBF2316C166EE; ETBF2316C166EJ; ETBF2513C124E; ETBF2513C124EE; ETBF2513C124EJ; ETBF2513C145EE; ETBF2513C145EJ; ETBF2513C166E; ETBF2513C166EE; ETBF2513C166EJ; ETBF2516C124EE; ETBF2516C145E; ETBF2516C145EE; ETBF2516C145EJ; ETBF2516C166E; ETBF2516C166EE; ETBF2516C166EJ; ETBF2813C124EE; ETBF2813C124EJ; ETBF2813C145E; ETBF2813C145EE; ETBF2813C145EJ; ETBF2813C166E; ETBF2813C166EE; ETBF2813C166EJ; ETBF2816C124EE; ETBF2816C124EJ; ETBF2816C145E; ETBF2816C145EE; ETBF2816C145EJ; ETBF2816C166E; ETBF2816C166EE; ETBF2816C166EJ; ETBF2820C124EE; ETBF2820C145EE; ETBF2820C145EJ; ETBF2820C166E; ETBF2820C166EE; ETBF2820C166EJ; ETBF3216C124E; ETBF3216C124EE; ETBF3216C124EJ; ETBF3216C145E; ETBF3216C145EE; ETBF3216C145EJ; ETBF3216C166EE; ETBF3216C166EJ; ETBF3220C124EE; ETBF3220C145EE; ETBF3220C166E; ETBF3220C166EE; ETBF3220C166EJ; ETBF3616C145E; ETBF3616C145EE; ETBF3616C145EJ; ETBF3616C166E; ETBF3616C166EE; ETBF3616C166EJ; ETBF3620C145EE; ETBF3620C166EE; ETBF3620C166EJ; ETCF2323C49E; ETCF2323C49EE; ETCF2525C49E; ETCF2525C49EE; ETCF2525C49EJ; ETCF2828C49E; ETCF2828C49EE; ETCF2828C49EJ; ETCF3232C49E; ETCF3232C49EE; ETCF3232C49EJ; ETCF3636C49E; ETCF3636C49EE; ETCF3636C49EJ; ETTF2323C70E; ETTF2323C70EE; ETTF2323C70EJ; ETTF2525C70E; ETTF2525C70EE; ETTF2525C70EJ; ETTF2828C70E; ETTF2828C70EE; ETTF2828C70EJ; ETTF3232C70E; ETTF3232C70EE; ETTF3232C70EJ; ETTF3636C70E; ETTF3636C70EE; ETTF3636C70EJ; ETUF2314C102E; ETUF2314C102EE; ETUF2314C102EJ; ETUF2514C102E; ETUF2514C102EE; ETUF2514C102EJ; ETUF2814C102E; ETUF2814C102EE; ETUF2814C102EJ; ETUF3214C102E; ETUF3214C102EE; ETUF3614C102E; ETUF3614C102EE; Medtronic Endurant IIs Stent Graft System Model/REF: ESBF2314C103E; ESBF2314C103EE; ESBF2314C103EJ; ESBF2514C103E; ESBF2514C103EE; ESBF2514C103EJ; ESBF2814C103E; ESBF2814C103EE; ESBF2814C103EJ; ESBF3214C103E; ESBF3214C103EE; ESBF3214C103EJ; ESBF3614C103E; ESBF3614C103EE; ESBF3614C103EJ; Used to treat infrarenal abdominal aortic or aortoiliac aneurysms using an endovascular approach.

Medtronic Vascular

Class I - Dangerous

Stent Graft Systems built with specific batches of taper tip assemblies have the potential for the taper tip to detach from the delivery system. Detachment of the taper tip during the implant procedure may lead to a secondary intervention to attempt to remove the taper tip, either by endovascular retrieval or surgical conversion. There is also potential that the taper tip is intentionally left behind in the patient, at the physician's discretion, if not easily removable.

Dec 21, 2021 Implants & Prosthetics Nationwide View Details β†’