Multiphor II Electrophoresis System GE HealthCare Product Number: 18-1018-06 PerkinElmer Part Number: 2117-004 Product Usage: Multiphor II Electrophoresis System is a versatile modular electrophoresis system for horizontal electrophoresis, isoelectric focusing, 2-D electrophoresis and electrophoretic transfer Note: The Multiphor II Electrophoresis System (Class I) is the platform used with the hemoglobin electrophoresis gel kit ( Class II) sold by PerkinElmer as the Resolve Hemoglobin Kit (K050709: Product code JJN). The RESOLVE Systems Hemoglobin kit (K050709) is designed to separate whole blood, cord blood or dried blood spot specimen for detection of normal and variant hemoglobins by isoelectric focusing.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial Number: 1171221 1171231 1171232 1171234 1181923 1181925 1181928 1187499 1189475 1189484 1195060 1195074 1197741 1197743 1197752 1197753 1200106 1202993 1202994 1251991 1252000 1253298 1253301 1253307 1254797 1254798 1255972 1264957 1264960 1302820 1302823 1313213 1320921 1320925 1320927 1331757 1383503 1383506 1406432 1413361 1434395 1434396 1434399 1435683 1436876 1442735 1463140 1465792 1465793 1465794 1480331 1480829 1480834 1486179 1486180 1487157 1487158 1487159 1487160 1487161 1487162 1487182 1497464 1497490 1509220 1525734 1525737 1525739 1525741 1525744 1525745 1525748 1540014 1540016 1540018 1540019 1540020 1550838 1550839 1561681 1564304 1604476 1604481 1607341 1607345 1607346 1624871 1624873
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Perkinelmer Life Sciences, Inc.
Reason for Recall:
Potential for electric shock in the case of a failure to install the grommets, supplied with the system.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Multiphor II Electrophoresis System GE HealthCare Product Number: 18-1018-06 PerkinElmer Part Number: 2117-004 Product Usage: Multiphor II Electrophoresis System is a versatile modular electrophoresis system for horizontal electrophoresis, isoelectric focusing, 2-D electrophoresis and electrophoretic transfer Note: The Multiphor II Electrophoresis System (Class I) is the platform used with the hemoglobin electrophoresis gel kit ( Class II) sold by PerkinElmer as the Resolve Hemoglobin Kit (K050709: Product code JJN). The RESOLVE Systems Hemoglobin kit (K050709) is designed to separate whole blood, cord blood or dried blood spot specimen for detection of normal and variant hemoglobins by isoelectric focusing.

Product Codes/Lot Numbers:

Serial Number: 1171221 1171231 1171232 1171234 1181923 1181925 1181928 1187499 1189475 1189484 1195060 1195074 1197741 1197743 1197752 1197753 1200106 1202993 1202994 1251991 1252000 1253298 1253301 1253307 1254797 1254798 1255972 1264957 1264960 1302820 1302823 1313213 1320921 1320925 1320927 1331757 1383503 1383506 1406432 1413361 1434395 1434396 1434399 1435683 1436876 1442735 1463140 1465792 1465793 1465794 1480331 1480829 1480834 1486179 1486180 1487157 1487158 1487159 1487160 1487161 1487162 1487182 1497464 1497490 1509220 1525734 1525737 1525739 1525741 1525744 1525745 1525748 1540014 1540016 1540018 1540019 1540020 1550838 1550839 1561681 1564304 1604476 1604481 1607341 1607345 1607346 1624871 1624873

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-2685-2017

Related Recalls

We have become aware through internal testing that JANUS and Janus G3 Automated Workstations were failing production final acceptance testing due to a Horizontal Motion Error message displayed on the system. Further investigation concluded that two printed circuit boards (PCB) used in MDT models of Janus and Janus G3 were manufactured with the incorrect resistor values. This anomaly affects component part number 7101640 (PCB-Quad Motor Controller MDT) and part number 7101647 (PCB-Quad Motor Controller Gripper). This anomaly could result in the Horizontal Motion Error being displayed and result in instrument down time. Until PerkinElmer s corrective measures are completed, we request you; ¿ Be aware that there is potential for the Horizontal Motion Error message to display which could result in instrument down time. ¿ Follow guidance provided in the Janus Automated Workstation User Manual when the Horizontal Motion Error message displays. ¿ If you have additional questions, contact PerkinElmer Technical Support for further assistance (800) 762-4000.

Aug 2, 2017 Surgical Instruments Nationwide View Details →